An in vivo comparison of the DREAM sequence with current RF shim technology.

Kay Nehrke1, Alois M Sprinkart, Peter Börnert.   

Abstract

OBJECT: In the present study the performance of the dual refocusing echo acquisition mode (DREAM) B1 (+) mapping sequence is evaluated for RF shimming in the abdomen at 3 T and validated against existing RF shim technology.
MATERIALS AND METHODS: In vivo experiments were performed on 19 normal volunteers using a clinical 3 T dual channel MRI system. For each volunteer three different B1 (+) mapping techniques [DREAM, actual flip angle imaging (AFI) and saturated double angle method (SDAM)] were employed for RF shimming of the liver and to subsequently assess the quality of the obtained RF shim settings in terms of the achieved B1 (+) homogeneity and accuracy of the mean B1 (+).
RESULTS: DREAM-based B1 (+) calibration led to an average homogeneity improvement of 39.1 % (AFI = 38.7 %, SDAM = 38.1 %) and a mean B1 (+) of 90.9 % of the prescribed B1 (+) (AFI = 88.9 %, SDAM = 92.0 %). The duration of the B1 (+) calibration scan was reduced from 30 s (AFI) and 15 s (SDAM) to 2.5 s (DREAM).
CONCLUSION: DREAM accelerates RF shimming of the liver by an order of magnitude without compromising RF shimming performance.

Entities:  

Mesh:

Year:  2014        PMID: 25108702     DOI: 10.1007/s10334-014-0454-3

Source DB:  PubMed          Journal:  MAGMA        ISSN: 0968-5243            Impact factor:   2.310


  28 in total

1.  Correction for B(1) and B(0) variations in quantitative T(2) measurements using MRI.

Authors:  J G Sled; G B Pike
Journal:  Magn Reson Med       Date:  2000-04       Impact factor: 4.668

2.  Transmit SENSE.

Authors:  Ulrich Katscher; Peter Börnert; Christoph Leussler; Johan S van den Brink
Journal:  Magn Reson Med       Date:  2003-01       Impact factor: 4.668

3.  Mapping of the radio frequency magnetic field with a MR snapshot FLASH technique.

Authors:  U Klose
Journal:  Med Phys       Date:  1992 Jul-Aug       Impact factor: 4.071

4.  Factors influencing flip angle mapping in MRI: RF pulse shape, slice-select gradients, off-resonance excitation, and B0 inhomogeneities.

Authors:  Jinghua Wang; Weihua Mao; Maolin Qiu; Michael B Smith; R Todd Constable
Journal:  Magn Reson Med       Date:  2006-08       Impact factor: 4.668

5.  Fast 3D radiofrequency field mapping using echo-planar imaging.

Authors:  F Jiru; U Klose
Journal:  Magn Reson Med       Date:  2006-12       Impact factor: 4.668

6.  Eigenmode analysis of transmit coil array for tailored B1 mapping.

Authors:  Kay Nehrke; Peter Börnert
Journal:  Magn Reson Med       Date:  2010-03       Impact factor: 4.668

7.  Volumetric B1 (+) mapping of the brain at 7T using DREAM.

Authors:  Kay Nehrke; Maarten J Versluis; Andrew Webb; Peter Börnert
Journal:  Magn Reson Med       Date:  2013-02-14       Impact factor: 4.668

8.  Trabecular bone volume fraction mapping by low-resolution MRI.

Authors:  M A Fernández-Seara; H K Song; F W Wehrli
Journal:  Magn Reson Med       Date:  2001-07       Impact factor: 4.668

9.  Rapid radiofrequency field mapping in vivo using single-shot STEAM MRI.

Authors:  Gunther Helms; Jürgen Finsterbusch; Nikolaus Weiskopf; Peter Dechent
Journal:  Magn Reson Med       Date:  2008-09       Impact factor: 4.668

10.  Impact of number of channels on RF shimming at 3T.

Authors:  Alexander S Childs; Shaihan J Malik; Declan P O'Regan; Joseph V Hajnal
Journal:  MAGMA       Date:  2013-01-13       Impact factor: 2.310

View more
  5 in total

1.  B1 and T1 mapping of the breast with a reference tissue method.

Authors:  Federico D Pineda; Milica Medved; Xiaobing Fan; Gregory S Karczmar
Journal:  Magn Reson Med       Date:  2015-05-27       Impact factor: 4.668

2.  A novel spectrally selective fat saturation pulse design with robustness to B0 and B1 inhomogeneities: A demonstration on 3D T1-weighted breast MRI at 3 T.

Authors:  Feng Xu; Wenbo Li; Dapeng Liu; Dan Zhu; Michael Schär; Kelly Myers; Qin Qin
Journal:  Magn Reson Imaging       Date:  2020-10-29       Impact factor: 2.546

3.  Estimating B1+ in the breast at 7 T using a generic template.

Authors:  Michael J van Rijssel; Josien P W Pluim; Peter R Luijten; Kenneth G A Gilhuijs; Alexander J E Raaijmakers; Dennis W J Klomp
Journal:  NMR Biomed       Date:  2018-03-23       Impact factor: 4.044

4.  Improved 7 Tesla transmit field homogeneity with reduced electromagnetic power deposition using coupled Tic Tac Toe antennas.

Authors:  Tales Santini; Sossena Wood; Narayanan Krishnamurthy; Tiago Martins; Howard J Aizenstein; Tamer S Ibrahim
Journal:  Sci Rep       Date:  2021-02-09       Impact factor: 4.379

5.  Ultrafast B1 mapping with RF-prepared 3D FLASH acquisition: Correcting the bias due to T1 -induced k-space filtering effect.

Authors:  Dan Zhu; Michael Schär; Qin Qin
Journal:  Magn Reson Med       Date:  2022-04-05       Impact factor: 3.737

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.