Literature DB >> 11804762

Effect of RF coil excitation on field inhomogeneity at ultra high fields: a field optimized TEM resonator.

T S Ibrahim1, R Lee, B A Baertlein, A M Abduljalil, H Zhu, P M Robitaille.   

Abstract

In this work, computational methods were utilized to optimize the field produced by the transverse electromagnetic (TEM) resonator in the presence of the human head at 8 Tesla. Optimization was achieved through the use of the classical finite difference time domain (FDTD) method and a TEM resonator loaded with an anatomically detailed human head model with a resolution of 2 mm x 2 mm x 2 mm. The head model was developed from 3D MR images. To account for the electromagnetic interactions between the coil and the tissue, the coil and the head were treated as a single system at all the steps of the model including, numerical tuning and excitation. In addition to 2, 3, 4, 6, and 10-port excitations, an antenna array concept was utilized by driving all the possible ports (24) of a 24-strut TEM resonator. The results show that significant improvement in the circularly polarized component of the transverse magnetic field could be obtained when using multiple ports and variable phase and fixed magnitude, or variable phase and variable magnitude excitations.

Entities:  

Mesh:

Year:  2001        PMID: 11804762     DOI: 10.1016/s0730-725x(01)00404-0

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  75 in total

1.  Design of an Electrically Automated RF Transceiver Head Coil in MRI.

Authors:  Sung-Min Sohn; Lance DelaBarre; Anand Gopinath; John Thomas Vaughan
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2014-10-28       Impact factor: 3.833

Review 2.  In vivo structural imaging of the cerebellum, the contribution of ultra-high fields.

Authors:  José P Marques; Rolf Gruetter; Wietske van der Zwaag
Journal:  Cerebellum       Date:  2012-06       Impact factor: 3.847

3.  Experience with magnetic resonance imaging of human subjects with passive implants and tattoos at 7 T: a retrospective study.

Authors:  Yacine Noureddine; Andreas K Bitz; Mark E Ladd; Markus Thürling; Susanne C Ladd; Gregor Schaefers; Oliver Kraff
Journal:  MAGMA       Date:  2015-09-26       Impact factor: 2.310

4.  Practical considerations for the design of sparse-spokes pulses.

Authors:  Marcin Jankiewicz; Huairen Zeng; Jason E Moore; Adam W Anderson; Malcolm J Avison; E Brian Welch; John C Gore
Journal:  J Magn Reson       Date:  2010-02-04       Impact factor: 2.229

5.  Small tip angle three-dimensional tailored radiofrequency slab-select pulse for reduced B1 inhomogeneity at 3 T.

Authors:  Suwit Saekho; Fernando E Boada; Douglas C Noll; V Andrew Stenger
Journal:  Magn Reson Med       Date:  2005-02       Impact factor: 4.668

6.  9.4T human MRI: preliminary results.

Authors:  Thomas Vaughan; Lance DelaBarre; Carl Snyder; Jinfeng Tian; Can Akgun; Devashish Shrivastava; Wanzahn Liu; Chris Olson; Gregor Adriany; John Strupp; Peter Andersen; Anand Gopinath; Pierre-Francois van de Moortele; Michael Garwood; Kamil Ugurbil
Journal:  Magn Reson Med       Date:  2006-12       Impact factor: 4.668

7.  Array-optimized composite pulse for excellent whole-brain homogeneity in high-field MRI.

Authors:  Christopher M Collins; Zhangwei Wang; Weihua Mao; Jieming Fang; Wanzhan Liu; Michael B Smith
Journal:  Magn Reson Med       Date:  2007-03       Impact factor: 4.668

8.  Mitigation of B1+ inhomogeneity using spatially selective excitation with jointly designed quadratic spatial encoding magnetic fields and RF shimming.

Authors:  Yi-Cheng Hsu; Riccardo Lattanzi; Ying-Hua Chu; Martijn A Cloos; Daniel K Sodickson; Fa-Hsuan Lin
Journal:  Magn Reson Med       Date:  2016-10-02       Impact factor: 4.668

Review 9.  Numerical field calculations considering the human subject for engineering and safety assurance in MRI.

Authors:  Christopher M Collins
Journal:  NMR Biomed       Date:  2009-11       Impact factor: 4.044

10.  Precompensation for mutual coupling between array elements in parallel excitation.

Authors:  Yong Pang; Xiaoliang Zhang
Journal:  Quant Imaging Med Surg       Date:  2011-12
View more