Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams.

Thomas A D Brown1, Kenneth R Hogstrom, Diane Alvarez, Kenneth L Matthews, Kyungmin Ham, Joseph P Dugas.   

Abstract

PURPOSE: This work investigates the dose-response curves of GAFCHROMIC(®) EBT, EBT2, and EBT3 radiochromic films using synchrotron-produced monochromatic x-ray beams. EBT2 film is being utilized for dose verification in photoactivated Auger electron therapy at the Louisiana State University Center for Advanced Microstructures and Devices (CAMD) synchrotron facility.
METHODS: Monochromatic beams of 25, 30, and 35 keV were generated on the tomography beamline at CAMD. Ion chamber depth-dose measurements were used to determine the dose delivered to films irradiated at depths from 0.7 to 8.5 cm in a 10 × 10 × 10-cm(3) polymethylmethacrylate phantom. AAPM TG-61 protocol was applied to convert measured ionization into dose. Films were digitized using an Epson 1680 Professional flatbed scanner and analyzed using the net optical density (NOD) derived from the red channel. A dose-response curve was obtained at 35 keV for EBT film, and at 25, 30, and 35 keV for EBT2 and EBT3 films. Calibrations of films for 4 MV x-rays were obtained for comparison using a radiotherapy accelerator at Mary Bird Perkins Cancer Center.
RESULTS: The sensitivity (NOD per unit dose) of EBT film at 35 keV relative to that for 4-MV x-rays was 0.73 and 0.76 for doses 50 and 100 cGy, respectively. The sensitivity of EBT2 film at 25, 30, and 35 keV relative to that for 4-MV x-rays varied from 1.09-1.07, 1.23-1.17, and 1.27-1.19 for doses 50-200 cGy, respectively. For EBT3 film the relative sensitivity was within 3% of unity for all three monochromatic x-ray beams.
CONCLUSIONS: EBT and EBT2 film sensitivity showed strong energy dependence over an energy range of 25 keV-4 MV, although this dependence becomes weaker for larger doses. EBT3 film shows weak energy dependence, indicating that it would be a better dosimeter for kV x-ray beams where beam hardening effects can result in large changes in the effective energy.

Entities:  

Mesh:

Year:  2012        PMID: 23231291     DOI: 10.1118/1.4767770

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  25 in total

1.  Acceptance, commissioning and clinical use of the WOmed T-200 kilovoltage X-ray therapy unit.

Authors:  Maria M Aspradakis; Paolo Zucchetti
Journal:  Br J Radiol       Date:  2015-07-30       Impact factor: 3.039

2.  Impact of source position on high-dose-rate skin surface applicator dosimetry.

Authors:  Jeho Jeong; Christopher A Barker; Marco Zaider; Gil'ad N Cohen
Journal:  Brachytherapy       Date:  2016-06-01       Impact factor: 2.362

3.  Performance of KCl:Eu2+ storage phosphor dosimeters for low-dose measurements.

Authors:  H Harold Li; Zhiyan Xiao; Rachael Hansel; Nels Knutson; Deshan Yang
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

4.  High throughput film dosimetry in homogeneous and heterogeneous media for a small animal irradiator.

Authors:  L Wack; W Ngwa; E Tryggestad; P Tsiamas; R Berbeco; S K Ng; J Hesser; P Zygmanski
Journal:  Phys Med       Date:  2013-03-17       Impact factor: 2.685

5.  Radiation dose verification of an X-ray based blood irradiator using EBT3 radiochromic films calibrated using Gamma Knife machine.

Authors:  Khaled Soliman; Marouf Adili; Abdullah Alrushoud
Journal:  Rep Pract Oncol Radiother       Date:  2019-06-07

6.  Energy and dose dependence of GafChromic EBT3-V3 film across a wide energy range.

Authors:  Sarah Bassi; Donal Cummins; Patrick McCavana
Journal:  Rep Pract Oncol Radiother       Date:  2019-12-09

7.  Dosimetry of a novel focused monoenergetic beam for radiotherapy.

Authors:  Jamshid Moradi-Kurdestany; Dirk A Bartkoski; Ramesh Tailor; Dragan Mirkovic; Ze'ev Harel; Aharon Bar-David; Michael Kleckner; Shirly Borukhin; Mohammad Salehpour
Journal:  Phys Med Biol       Date:  2022-03-29       Impact factor: 4.174

8.  Film based verification of calculation algorithms used for brachytherapy planning-getting ready for upcoming challenges of MBDCA.

Authors:  Grzegorz Zwierzchowski; Grzegorz Bielęda; Janusz Skowronek; Magdalena Mazur
Journal:  J Contemp Brachytherapy       Date:  2016-08-16

9.  Build-up material requirements in clinical dosimetry during total body irradiation treatments.

Authors:  Martin Butson; Dane Pope; Mamoon Haque; Tom Chen; Guangli Song; May Whitaker
Journal:  J Med Phys       Date:  2016 Apr-Jun

10.  Comparison of EBT and EBT3 RadioChromic Film Usage in Parotid Cancer Radiotherapy.

Authors:  M T Bahreyni Toossi; F Khorshidi; M Ghorbani; N Mohamadian; D Davenport
Journal:  J Biomed Phys Eng       Date:  2016-03-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.