Impact of source position on high-dose-rate skin surface applicator dosimetry.

Jeho Jeong1, Christopher A Barker2, Marco Zaider1, Gil'ad N Cohen3.   

Abstract

PURPOSE: Skin surface dosimetric discrepancies between measured and treatment planning system predicted values were traced to source position sag inside the applicator and to source transit time. We quantified their dosimetric impact and propose corrections for clinical use. METHODS AND MATERIALS: We measured the dose profiles from the Varian Leipzig-style high-dose-rate (HDR) skin applicator, using EBT3 film, photon diode, and optically stimulated luminescence dosimeter for three different GammaMedplus HDR afterloaders. The measured dose profiles at several depths were compared with BrachyVision Acuros calculated profiles. To assess the impact of the source sag, two different applicator orientations were considered. The dose contribution during source transit was assessed by comparing diode measurements using an HDR timer and an electrometer timer.
RESULTS: Depth doses measured using the three dosimeters were in good agreement, but were consistently higher than the Acuros dose calculations. Measurements with the applicator face up were significantly (exceeding 10%) lower than those in the face down position, due to source sag inside the applicator. Based on the inverse square law, the effective source sag was evaluated to be about 0.5 mm from the planned position. The additional dose during source transit was evaluated to be about 2.8% for 30 seconds of treatment with a 40700 U (10 Ci) source.
CONCLUSION: With a very short source-to-surface distance, the small source sag inside the applicator has a significant dosimetric impact. This effect is unaccounted for in the vendor's treatment planning template and should be considered before the clinical use of the applicator. Further investigation of other applicators with large source lumen diameter may be warranted.
Copyright © 2016 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Dosimetry; Dwell position uncertainty; HDR brachytherapy; Leipzig skin-applicator; Skin cancer; Source transit dose

Mesh:

Year:  2016        PMID: 27263059      PMCID: PMC5499667          DOI: 10.1016/j.brachy.2016.04.389

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  32 in total

1.  Ir-192 HDR transit dose and radial dose function determination using alanine/EPR dosimetry.

Authors:  Carmen S Guzmán Calcina; Adelaide de Almeida; José R Oliveira Rocha; Felipe Chen Abrego; Oswaldo Baffa
Journal:  Phys Med Biol       Date:  2005-02-23       Impact factor: 3.609

2.  High dose per fraction dosimetry of small fields with gafchromic EBT2 film.

Authors:  Nicholas Hardcastle; Amar Basavatia; Adam Bayliss; Wolfgang A Tomé
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

3.  What degree of accuracy is required and can be achieved in photon and neutron therapy?

Authors:  B J Mijnheer; J J Battermann; A Wambersie
Journal:  Radiother Oncol       Date:  1987-03       Impact factor: 6.280

4.  Superficial x-ray in the treatment of basal and squamous cell carcinomas: a viable option in select patients.

Authors:  Armand B Cognetta; Brett M Howard; Henry P Heaton; Earl R Stoddard; Hyokyoung Grace Hong; W Harris Green
Journal:  J Am Acad Dermatol       Date:  2012-07-19       Impact factor: 11.527

5.  Surface dose characterisation of the Varian Ir-192 HDR conical surface applicator set with a vertically orientated source.

Authors:  Konrad Buchauer; Guido Henke; Hans Schiefer; Ludwig Plasswilm
Journal:  Strahlenther Onkol       Date:  2014-08-01       Impact factor: 3.621

6.  [The indications for and results of HDR afterloading therapy in diseases of the skin and mucosa with standardized surface applicators (the Leipzig applicator)].

Authors:  A Köhler-Brock; W Prager; S Pohlmann; S Kunze
Journal:  Strahlenther Onkol       Date:  1999-04       Impact factor: 3.621

Review 7.  Nonmelanoma skin cancer.

Authors:  Tri H Nguyen; Diana Quynh-Dao Ho
Journal:  Curr Treat Options Oncol       Date:  2002-06

8.  Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements.

Authors:  Paul A Jursinic
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

9.  Dwell position inaccuracy in the Varian GammaMed HDR ring applicator.

Authors:  Robin L Stern; Tianxiao Liu
Journal:  J Appl Clin Med Phys       Date:  2010-09-07       Impact factor: 2.102

10.  Non-melanoma skin cancer treated with HDR Valencia applicator: clinical outcomes.

Authors:  Alejandro Tormo; Francisco Celada; Silvia Rodriguez; Rafael Botella; Antonio Ballesta; Michael Kasper; Zoubir Ouhib; Manuel Santos; Jose Perez-Calatayud
Journal:  J Contemp Brachytherapy       Date:  2014-06-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.