Acceptance, commissioning and clinical use of the WOmed T-200 kilovoltage X-ray therapy unit.

Maria M Aspradakis1, Paolo Zucchetti1.   

Abstract

OBJECTIVE: The objective of this work was to characterize the performance of the WOmed T-200-kilovoltage (kV) therapy machine.
METHODS: Mechanical functionality, radiation leakage, alignment and interlocks were investigated. Half-value layers (HVLs) (first and second HVLs) from X-ray beams generated from tube potentials between 30 and 200 kV were measured. Reference dose was determined in water. Beam start-up characteristics, dose linearity and reproducibility, beam flatness, and uniformity as well as deviations from inverse square law were assessed. Relative depth doses (RDDs) were determined in water and water-equivalent plastic. The quality assurance program included a dosimetry audit with thermoluminescent dosemeters.
RESULTS: All checks on machine performance were satisfactory. HVLs ranged between 0.45-4.52 mmAl and 0.69-1.78 mmCu. Dose rates varied between 0.2 and 3 Gy min(-1) with negligible time-end errors. There were differences in measured RDDs from published data. Beam outputs were confirmed with the dosimetry audit. The use of published backscatter factors was implemented to account for changes in phantom scatter for treatments with irregularly shaped fields.
CONCLUSION: Guidance on the determination of HVL and RDD in kV beams can be contradictory. RDDs were determined through measurement and curve fitting. These differed from published RDD data, and the differences observed were larger in the low-kV energy range. ADVANCES IN KNOWLEDGE: This article reports on the comprehensive and novel approach to the acceptance, commissioning and clinical use of a modern kV therapy machine. The challenges in the dosimetry of kV beams faced by the medical physicist in the clinic are highlighted.

Entities:  

Mesh:

Year:  2015        PMID: 26224430      PMCID: PMC4743444          DOI: 10.1259/bjr.20150001

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  25 in total

Review 1.  Review of photon interaction cross section data in the medical and biological context.

Authors:  J H Hubbell
Journal:  Phys Med Biol       Date:  1999-01       Impact factor: 3.609

2.  Addendum to the IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL).

Authors:  R J Aukett; J E Burns; A G Greener; R M Harrison; C Moretti; A E Nahum; K E Rosser
Journal:  Phys Med Biol       Date:  2005-05-25       Impact factor: 3.609

3.  Measurement of percentage depth dose and lateral beam profile for kilovoltage x-ray therapy beams.

Authors:  X A Li; C M Ma; D Salhani
Journal:  Phys Med Biol       Date:  1997-12       Impact factor: 3.609

4.  An alternative beam quality index for medium-energy x-ray dosimetry.

Authors:  K E Rosser
Journal:  Phys Med Biol       Date:  1998-03       Impact factor: 3.609

5.  Performance characteristics of the Pantak DXT-300 kilovoltage X-ray treatment machine.

Authors:  R J Aukett; D W Thomas; A W Seaby; J T Gittins
Journal:  Br J Radiol       Date:  1996-08       Impact factor: 3.039

6.  The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL; 10-300 kV generating potential). Institution of Physics and Engineering in Medicine and Biology.

Authors: 
Journal:  Phys Med Biol       Date:  1996-12       Impact factor: 3.609

7.  Characteristics of orthovoltage x-ray therapy beams at extended SSD for applicators with end plates.

Authors:  X A Li; D Salhani; C M Ma
Journal:  Phys Med Biol       Date:  1997-02       Impact factor: 3.609

8.  Microionization chamber air-kerma calibration coefficients as a function of photon energy for x-ray spectra in the range of 20-250 kVp relative to 60Co.

Authors:  J R Snow; J A Micka; L A DeWerd
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

9.  Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range.

Authors:  J Eduardo Villarreal-Barajas; Rao F H Khan
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

10.  Assessing the deviation from the inverse square law for orthovoltage beams with closed-ended applicators.

Authors:  James Gräfe; Yannick Poirier; Ferenc Jacso; Rao Khan; Hong-Wei Liu; J Eduardo Villarreal-Barajas
Journal:  J Appl Clin Med Phys       Date:  2014-07-08       Impact factor: 2.102

View more
  3 in total

1.  Excessive applicator radiation leakage for a common therapeutic kilovoltage system.

Authors:  Brad Beeksma; Joerg Lehmann
Journal:  Br J Radiol       Date:  2018-11-15       Impact factor: 3.039

2.  Current status of kilovoltage (kV) radiotherapy in the UK: installed equipment, clinical workload, physics quality control and radiation dosimetry.

Authors:  Antony L Palmer; Michael Pearson; Paul Whittard; Katie E McHugh; David J Eaton
Journal:  Br J Radiol       Date:  2016-10-12       Impact factor: 3.039

3.  Off-axis dose distribution with stand-in and stand-off configurations for superficial radiotherapy treatments.

Authors:  Keith Davey; Margaret Moore; Sinéad Cleary; Christoph Kleefeld; Mark J Foley
Journal:  J Appl Clin Med Phys       Date:  2019-10-12       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.