Energy and dose dependence of GafChromic EBT3-V3 film across a wide energy range.

Sarah Bassi1, Donal Cummins1, Patrick McCavana1.   

Abstract

AIM: To determine the energy and dose dependence of GafChromic EBT3-V3 film over an energy range 0.2 mm Al HVL to 6 MV.
BACKGROUND: The decay scheme of a brachytherapy source may be complex and the spectrum of energy can be wide. LiF TLDs are the golden standard recommended for dosimetric measures in brachytherapy, for their energy independence, but TLDs could be not available in some centres. An alternative way to perform dose measurements is to use GafChromic films, but they show energy dependence. METHODS AND MATERIALS: Films have been irradiated at increasing dose with three different beams: 6 MV beam, TPR20, 10 = (0.684 ± 0.01), HVL = (2.00 ± 0.01)mmAl and HVL = (0.20 ± 0.01)mmAl. Calibration curves were generated using the same dose range (0cGy to 850cGy) for the three energies. Using the 6 MV calibration curve as reference, the film response in terms of net optical density (OD) was evaluated.
RESULTS: The difference in the calibration curve obtained by irradiating the film with 6 MV and 2 mm Al HVL energy beams is less than 3 %, within the calibration uncertainty, in the dose range 500-850cGy. The OD of EBT3-V3 film is significantly lower at 0.2 mmAl HVL compared to 6 MV, showing differences up to 25 %.
CONCLUSION: Within the range 6 MV-2 mm Al HVL and dose higher than 500cGy, GafChromic EBT3-V3 films are energy independent. In this dose range, films can be calibrated in a simple geometry, using a 6 MV Linac beam, and can be used for brachytherapy sources dose measures. The use of EBT3 films can be extended to reference dosimetry in Ir-192 clinical brachytherapy.
© 2019 Greater Poland Cancer Centre. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  DIN6809-4 standard; Dose dependence; Dosimetry; Energy dependence; GafChromic EBT3-V3; IPEM standard

Year:  2019        PMID: 31889923      PMCID: PMC6931201          DOI: 10.1016/j.rpor.2019.12.007

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  4 in total

1.  Addendum to the IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL).

Authors:  R J Aukett; J E Burns; A G Greener; R M Harrison; C Moretti; A E Nahum; K E Rosser
Journal:  Phys Med Biol       Date:  2005-05-25       Impact factor: 3.609

2.  Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams.

Authors:  Thomas A D Brown; Kenneth R Hogstrom; Diane Alvarez; Kenneth L Matthews; Kyungmin Ham; Joseph P Dugas
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

3.  Experimental investigation of GafChromic® EBT3 intrinsic energy dependence with kilovoltage x rays, 137 Cs, and 60 Co.

Authors:  Cliff G Hammer; Benjamin Saul Rosen; Jessica M Fagerstrom; Wesley S Culberson; Larry A DeWerd
Journal:  Med Phys       Date:  2017-12-22       Impact factor: 4.071

4.  Dose measurements nearby low energy electronic brachytherapy sources using radiochromic film.

Authors:  Slobodan Devic; LiHeng Liang; Nada Tomic; Hamed Bekerat; Marc Morcos; Marija Popovic; Peter Watson; Saad Aldelaijan; Jan Seuntjens
Journal:  Phys Med       Date:  2019-06-22       Impact factor: 2.685

  4 in total
  1 in total

1.  Evaluation of dose distribution of an intraoperative radiotherapy device using thermoluminescent dosimeters and radiographic films.

Authors:  Omid Baziar; Hamid Gholamhosseinian; Seyed Amir Aledavood; Mahboubeh Sadeghi
Journal:  Rep Pract Oncol Radiother       Date:  2022-07-29
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.