Dosimetry of a novel focused monoenergetic beam for radiotherapy.

Jamshid Moradi-Kurdestany1, Dirk A Bartkoski1, Ramesh Tailor1, Dragan Mirkovic1, Ze'ev Harel2, Aharon Bar-David2, Michael Kleckner2, Shirly Borukhin2, Mohammad Salehpour1.   

Abstract

Objective. A novel treatment modality is currently being developed that produces converging monoenergetic x-rays. Conventional application of dosimetric calibration as presented in protocol TG61 is not applicable. Furthermore, the dosimetry of the focal point of the converging beam is on the order of a few millimeters, requiring a high-resolution dosimeter. Here we present a procedure to calibrate radiochromic film for narrow-beam monoenergetic 60 keV photons as well as absolute dosimetry of monoenergetic focused x-rays. A study of the focal spot dose rate after passing through a bone-equivalent material was also done to quantify the effects of heterogeneous materials.Approach.This was accomplished by configuring a polyenergetic beam of equivalent energy using a clinical orthovoltage machine. Calibrated films were then used to perform absolute dosimetry of the converging beam by measuring the beam profile at various depths in water. Main Results.A method for calibrating radiochromic film has been developed and detailed that allows absolute dosimetry of a monoenergetic photon beam. Absolute dosimetry of a focused, mono-energetic beam resulted in a focal spot dose rate of ∼30 cGy min-1at a depth of 5 cm in water.Significance.This work serves to establish a dosimetry protocol for mono-energetic beam absolute dosimetry as well as the use of such a method for measurement of a novel teletherapy modality. Creative Commons Attribution license.

Entities:  

Keywords:  energy response; gafchromic film dosimetry; orthovoltage dosimetry

Mesh:

Substances:

Year:  2022        PMID: 35272282      PMCID: PMC9020448          DOI: 10.1088/1361-6560/ac5c8f

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   4.174


  9 in total

1.  AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology.

Authors:  C M Ma; C W Coffey; L A DeWerd; C Liu; R Nath; S M Seltzer; J P Seuntjens
Journal:  Med Phys       Date:  2001-06       Impact factor: 4.071

2.  Radiochromic film: the new x-ray dosimetry and imaging tool.

Authors:  Martin J Butson; Tsang Cheung; Peter K N Yu
Journal:  Australas Phys Eng Sci Med       Date:  2004-12       Impact factor: 1.430

3.  Dosimetric properties of improved GafChromic films for seven different digitizers.

Authors:  Slobodan Devic; Jan Seuntjens; Gyorgy Hegyi; Ervin B Podgorsak; Christopher G Soares; Assen S Kirov; Imad Ali; Jeffrey F Williamson; Angel Elizondo
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

4.  Film dosimetry calibration method for pulsed-dose-rate brachytherapy with an 192Ir source.

Authors:  Nathan Schwob; Itzhak Orion
Journal:  Med Phys       Date:  2007-05       Impact factor: 4.071

5.  Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams.

Authors:  Thomas A D Brown; Kenneth R Hogstrom; Diane Alvarez; Kenneth L Matthews; Kyungmin Ham; Joseph P Dugas
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

6.  Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams.

Authors:  Suphalak Khachonkham; Ralf Dreindl; Gerd Heilemann; Wolfgang Lechner; Hermann Fuchs; Hugo Palmans; Dietmar Georg; Peter Kuess
Journal:  Phys Med Biol       Date:  2018-03-15       Impact factor: 3.609

7.  Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG-55.

Authors:  Azam Niroomand-Rad; Sou-Tung Chiu-Tsao; Michael P Grams; David F Lewis; Christopher G Soares; Leo J Van Battum; Indra J Das; Samuel Trichter; Michael W Kissick; Guerda Massillon-Jl; Paola E Alvarez; Maria F Chan
Journal:  Med Phys       Date:  2020-10-30       Impact factor: 4.506

8.  Devices for dosimetric measurements and quality assurance of the Xstrahl 300 orthovoltage unit.

Authors:  Tze Yee Lim; Dragan Mirkovic; Xin Wang; Ramesh Tailor
Journal:  J Appl Clin Med Phys       Date:  2021-03-18       Impact factor: 2.102

9.  Analysis of a novel X-ray lens for converging beam radiotherapy.

Authors:  Dirk A Bartkoski; Aharon Bar-David; Michael Kleckner; Dragan Mirkovic; Ramesh Tailor; Jamshid Moradi-Kurdestany; Shirly Borukhin; Ze'ev Harel; Zeev Burshtein; Asaf Zuck; Mohammad Salehpour
Journal:  Sci Rep       Date:  2021-09-28       Impact factor: 4.379

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.