Analytical characterization of electrochemical biosensor test strips for measurement of glucose in low-volume interstitial fluid samples.

M E Collison1, P J Stout, T S Glushko, K N Pokela, D J Mullins-Hirte, J R Racchini, M A Walter, S P Mecca, J Rundquist, J J Allen, M E Hilgers, T B Hoegh.   

Abstract

BACKGROUND: Minimally invasive interstitial fluid (ISF) sampling and glucose measurement technologies were integrated into a hand-held device for diabetic glucose monitoring investigations.
METHODS: Conventional electrochemical test strip technology (Bayer Glucometer Elite) was adapted to measure glucose in small (0.5-2.0 microL) samples of ISF. Test strip glucose measurements were performed on a commercial potentiostat and were compared to various reference glucose methodologies (YSI 2300 analyzer, microhexokinase procedure, Bayer Glucometer Elite). Characterizations of the integrated ISF sampling-glucose test strip design included accuracy and precision in various sample media (saline, ISF surrogates, diabetic ISF samples), sample volume dependence, test strip sterilization studies (electron beam, gamma irradiation), and diabetic ISF sampling and glucose measurements.
RESULTS: Glucose measurements were free from significant media effects. Sample volume variations (0.6-3.2 microL) revealed only modest dependence of glucose measurement bias on sample volume (-1.5% per microliter). Sterilization treatments had only a minor impact on glucose response and test strip aging and no significant impact on interferent responses of the glucose test strips. Diabetic subject testing under minimum fasting conditions of at least 2 h with integrated ISF sampling and glucose measurement gave low ISF glucose measurement imprecision (CV, 4%) and mean glucose results that were indistinguishable from reference (microhexokinase) ISF glucose measurements and from capillary blood glucose measurements (Glucometer Elite).
CONCLUSIONS: Conventional single-use, electrochemical glucose test strip and ISF collection technologies can be readily integrated to provide real-time ISF sampling and glucose measurements for diabetic monitoring applications.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10471681

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  6 in total

Review 1.  A tale of two compartments: interstitial versus blood glucose monitoring.

Authors:  Eda Cengiz; William V Tamborlane
Journal:  Diabetes Technol Ther       Date:  2009-06       Impact factor: 6.118

2.  Periodic extraction of interstitial fluid from the site of subcutaneous insulin infusion for the measurement of glucose: a novel single-port technique for the treatment of type 1 diabetes patients.

Authors:  Werner Regittnig; Stefan Lindpointner; Stefan Korsatko; Dina Tutkur; Manfred Bodenlenz; Thomas R Pieber
Journal:  Diabetes Technol Ther       Date:  2012-11-05       Impact factor: 6.118

3.  Impaired cardiac energy metabolism in embryos lacking adrenergic stimulation.

Authors:  Candice N Baker; Sarah A Gidus; George F Price; Jessica N R Peoples; Steven N Ebert
Journal:  Am J Physiol Endocrinol Metab       Date:  2014-12-16       Impact factor: 4.310

4.  Microneedle-based automated therapy for diabetes mellitus.

Authors:  Puneet Khanna; Joel A Strom; John I Malone; Shekhar Bhansali
Journal:  J Diabetes Sci Technol       Date:  2008-11

5.  System Integration - A Major Step toward Lab on a Chip.

Authors:  Mandy Ly Sin; Jian Gao; Joseph C Liao; Pak Kin Wong
Journal:  J Biol Eng       Date:  2011-05-25       Impact factor: 4.355

6.  Extraction and biomolecular analysis of dermal interstitial fluid collected with hollow microneedles.

Authors:  Philip R Miller; Robert M Taylor; Bao Quoc Tran; Gabrielle Boyd; Trevor Glaros; Victor H Chavez; Raga Krishnakumar; Anupama Sinha; Kunal Poorey; Kelly P Williams; Steven S Branda; Justin T Baca; Ronen Polsky
Journal:  Commun Biol       Date:  2018-10-22
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.