Abstract
OBJECTIVE: The purpose of this study was to determine the accuracy and reliability of Frankfort horizontal plane identification using displays of multi-planar reconstructed MRI images, and propose it as a sufficiently stable and standardized reference plane for craniofacial structures.Entities:
Mesh:
Year: 2012 PMID: 23118970 PMCID: PMC3485237 DOI: 10.1371/journal.pone.0048281
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1Using coronal view as the orientation source and plotting landmarks on the sagittal view.
a) Left Porion. b) Right Porion. c) Left Orbitale.
Figure 2Plotted landmarks on 3D rendered images with MPR view.
Figure 3Images were viewed using multi planar reconstruction (MPR) with coronal view as the center of orientation.
a) Horizontal and vertical lines were adjusted until the standard trans-porionic axis was observed. b) Balance between axial, coronal and sagittal axis adjusted to connect trans-porionic axis (purple) with Orbitale through a coronal axis (blue). c) Axial view through FH after connecting the 3 axes.
Intra-Examiner Reproducibility – Mean Difference, Coefficients of Variability, and Intra-class Correlation statistics.
| First Reading | Second Reading | Second – First Reading | Coefficient of Variability | Intra-class Correlation | |||
| Coordinate | Mean (SD) | Mean (SD) | Mean (SD) | [%] | Coefficient (95% CI) | ||
| Left porion | |||||||
| x | 59.82 (4.85) | 59.29 (4.78) | −0.52 (1.26) | 1.27 | 0.961 (0.920–0.980) | ||
| y | 18.61 (6.72) | 18.41 (6.84) | −0.20 (1.50) | 4.46 | 0.976 (0.956–0.987) | ||
| z | 132.83 (12.21) | 132.22 (12.08) | −0.61 (1.31) | 0.56 | 0.993 (0.984–0.997) | ||
| Right porion | |||||||
| x | 64.10 (4.91) | 63.25 (5.02) | −0.86 (1.43) | 1.40 | 0.945 (0.853–0.975) | ||
| y | 17.39 (7.46) | 16.70 (7.59) | −0.69 (1.13) | 5.86 | 0.985 (0.955–0.993) | ||
| z | 134.57 (11.87) | 133.67 (11.87) | −0.90 (1.64) | 0.75 | 0.988 (0.968–0.994) | ||
| Left Orbitale | |||||||
| x | 34.08 (6.46) | 33.99 (6.66) | −0.09 (0.97) | 1.17 | 0.989 (0.989–0.994) | ||
| y | 94.08 (8.53) | 93.78 (8.62) | −0.30 (0.75) | 0.50 | 0.996 (0.991–0.998) | ||
| z | 134.43 (15.38) | 134.17 (15.15) | −0.26 (0.91) | 0.35 | 0.998 (0.996–0.999) | ||
Intra- and Inter-reader Reliability – Reader Differences in Distances for Points Defining FH Plane.
| Intra-Reader Reliability(Distance d to First Reading) | Inter-Reader Reliability (Distance d to First Reading of Reader 1 as the Gold Standard) | ||||
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | |
| Left Porion | |||||
| Mean (SD), mm | 2.0 (1.4) | 2.3 (1.9) | 2.2 (2.0) | 2.1 (1.6) | 2.2 (2.0) |
| ≤1 mm, % | 18.6 | 18.6 | 23.3 | 20.9 | 25.6 |
| ≤2 mm, % | 62.8 | 55.8 | 69.8 | 67.4 | 62.8 |
| ≤3 mm, % | 79.1 | 83.7 | 81.4 | 81.4 | 79.1 |
| ≤4 mm, % | 86.0 | 90.7 | 81.4 | 88.4 | 83.7 |
| ≤5 mm, % | 93.0 | 90.7 | 88.4 | 90.7 | 90.7 |
| Right Porion | |||||
| Mean (SD), mm | 2.4 (1.5) | 2.9 (2.1) | 2.3 (1.7) | 2.0 (1.2) | 2.2 (1.4) |
| ≤1 mm, % | 14.0 | 9.3 | 9.3 | 18.6 | 16.3 |
| ≤2 mm, % | 46.5 | 41.9 | 53.5 | 58.1 | 53.5 |
| ≤3 mm, % | 67.4 | 65.1 | 76.7 | 81.4 | 72.1 |
| ≤4 mm, % | 83.7 | 76.7 | 88.4 | 90.7 | 90.7 |
| ≤5 mm, % | 95.4 | 88.4 | 93.0 | 97.7 | 95.4 |
| Left Orbitale | |||||
| Mean (SD), mm | 1.3 (0.9) | 1.9 (1.0) | 1.7 (1.0) | 1.6 (0.8) | 1.3 (0.8) |
| ≤1 mm, % | 37.2 | 16.3 | 25.6 | 20.9 | 30.2 |
| ≤2 mm, % | 79.1 | 62.8 | 69.8 | 69.8 | 79.1 |
| ≤3 mm, % | 97.7 | 83.7 | 90.7 | 95.4 | 97.7 |
| ≤4 mm, % | 97.7 | 95.4 | 95.4 | 97.7 | 100.0 |
| ≤5 mm, % | 100.0 | 97.7 | 100.0 | 100.0 | 100.0 |
Intra- Examiner Reproducibility and Inter-Examiner Reliability – Reader Differences in Dihedral Angle of FH Plane.
| Intra-Examiner Reproducibility(Angle to First Reading) | Inter-Examiner Reliability (Angle to First Reading of Examiner) | ||||
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | |
| Mean (SD), ° | 1.0 (0.7) | 1.5 (0.8) | 1.3 (1.0) | 1.1 (1.0) | 1.1 (0.8) |
| ≤1°, % | 53.5 | 25.6 | 53.5 | 58.1 | 55.8 |
| ≤2°, % | 88.4 | 74.4 | 86.0 | 86.0 | 86.0 |
| ≤3°, % | 97.7 | 93.0 | 90.7 | 88.4 | 97.7 |
| ≤4°, % | 100.0 | 100.0 | 97.7 | 100.0 | 100.0 |
| ≤5°, % | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 |
Inter-Examiner Reliability – Mean differences, Coefficients of Variability (CV) and Intra-class Correlation statistics.
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Intra-class Correlation for 5 Examinerss | |||||
| Coordinate | Mean (SD) | Difference to First Examiner Mean (SD) | CV [%] | Difference to First Examiner Mean (SD) | CV [%] | Difference to First Examiner Mean (SD) | CV [%] | Difference to First Examiner Mean(SD) | CV [%] | Coefficient (95% CI) |
| Left porion | ||||||||||
| x | 59.82 (4.85) | 0.43 (1.64) | 1.40 | 0.14 (2.02) | 1.45 | 0.31 (1.39) | 1.25 | 0.08 (1.83) | 1.50 | 0.932 (0.897–0.959) |
| y | 18.61 (6.72) | 0.20 (1.82) | 5.68 | 0.28 (1.66) | 4.29 | 0.42 (1.61) | 4.43 | 0.03 (1.60) | 5.77 | 0.971 (0.955–0.982) |
| z | 132.83 (12.21) | 0.83 (1.35) | 0.66 | 0.20 (1.51) | 0.64 | 0.20 (1.54) | 0.66 | 0.67 (1.63) | 0.65 | 0.992 (0.987–0.995) |
| Right porion | ||||||||||
| x | 64.10 (4.91) | 0.97 (2.34) | 1.76 | 0.72 (1.95) | 1.60 | 0.27 (1.33) | 0.99 | 0.32 (1.70) | 1.40 | 0.914 (0.871–0.948) |
| y | 17.39 (7.46) | 0.51 (1.16) | 5.59 | 0.18 (1.31) | 3.78 | 0.13 (1.01) | 3.07 | 0.18 (1.15) | 3.78 | 0.988 (0.981–0.993) |
| z | 134.57 (11.87) | 1.52 (1.56) | 0.92 | 0.78 (1.29) | 0.65 | 0.73 (1.44) | 0.66 | 0.77 (1.41) | 0.65 | 0.991 (0.984–0.995) |
| Left Orbitale | ||||||||||
| x | 34.08 (6.46) | 0.07 (1.24) | 1.95 | 0.26 (0.95) | 1.32 | 0.13 (0.81) | 1.08 | 0.23 (0.72) | 0.96 | 0.987 (0.980–0.992) |
| y | 94.08 (8.53) | 0.66 (1.19) | 0.83 | 0.32 (1.21) | 0.72 | 0.03 (0.91) | 0.51 | 0.03 (0.75) | 0.46 | 0.990 (0.984–0.994) |
| z | 134.43 (15.38) | 0.18 (1.20) | 0.47 | 0.67 (0.98) | 0.47 | 0.81 (1.06) | 0.57 | 0.52 (1.02) | 0.46 | 0.997 (0.994–0.998) |
Figure 4Bland-Altman Plot for Examiner 1 vs. Examiner 5.
a) In x coordinate of right Porion. The horizontal lines represent the mean within-pair difference (−0.32 for examiner 5– examiner 1), the mean ±1.96 standard deviations (1 SD = 1.70), and the mean ±1.96 SD ± square root (3*SD2/n) for 95% CIs of limits of mean ±1.96 standard deviations. b) In y coordinate of right Porion.
Figure 5Bland-Altman Plot for Examiner 1 vs. Examiner 5.
a) in x coordinate of left Orbitale (the shown digit preference reflects the slice thickness of 1 mm) b) In y coordinate of left Orbitale.
Figure 6Axial head view, A. FH plane estimated prior scan on subjects head B. FH plane identified by landmarks on Images.
(Note the differences in the muscles cross sectional area between both views).