Literature DB >> 19577142

Accuracy and reliability of linear cephalometric measurements from cone-beam computed tomography scans of a dry human skull.

Mauricio Berco1, Paul H Rigali, R Matthew Miner, Stephelynn DeLuca, Nina K Anderson, Leslie A Will.   

Abstract

INTRODUCTION: The purpose of this study was to determine the accuracy and reliability of 3-dimensional craniofacial measurements obtained from cone-beam computed tomography (CBCT) scans of a dry human skull.
METHODS: Seventeen landmarks were identified on the skull. CBCT scans were then obtained, with 2 skull orientations during scanning. Twenty-nine interlandmark linear measurements were made directly on the skull and compared with the same measurements made on the CBCT scans. All measurements were made by 2 operators on 4 separate occasions.
RESULTS: The method errors were 0.19, 0.21, and 0.19 mm in the x-, y- and z-axes, respectively. Repeated measures analysis of variance (ANOVA) showed no significant intraoperator or interoperator differences. The mean measurement error was -0.01 mm (SD, 0.129 mm). Five measurement errors were found to be statistically significantly different; however, all measurement errors were below the known voxel size and clinically insignificant. No differences were found in the measurements from the 2 CBCT scan orientations of the skull.
CONCLUSIONS: CBCT allows for clinically accurate and reliable 3-dimensional linear measurements of the craniofacial complex. Moreover, skull orientation during CBCT scanning does not affect the accuracy or the reliability of these measurements.

Entities:  

Mesh:

Year:  2009        PMID: 19577142     DOI: 10.1016/j.ajodo.2008.08.021

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  48 in total

1.  3D cephalometric analysis obtained from computed tomography. Review of the literature.

Authors:  Giulia Rossini; Costanza Cavallini; Michele Cassetta; Ersilia Barbato
Journal:  Ann Stomatol (Roma)       Date:  2012-01-27

2.  Accuracy of CBCT measurements of a human skull.

Authors:  Kivanç Kamburoğlu; Eray Kolsuz; Hakan Kurt; Cenk Kiliç; Tuncer Özen; Candan Semra Paksoy
Journal:  J Digit Imaging       Date:  2011-10       Impact factor: 4.056

3.  Dry skull positioning device for extra-oral radiology and cone-beam CT.

Authors:  Thiago Leite Beaini; Paulo Eduardo Miamoto Dias; Rodolfo Francisco Haltenhoff Melani
Journal:  Int J Legal Med       Date:  2013-08-29       Impact factor: 2.686

4.  Application of virtual surgical planning with computer assisted design and manufacturing technology to cranio-maxillofacial surgery.

Authors:  Linping Zhao; Pravin K Patel; Mimis Cohen
Journal:  Arch Plast Surg       Date:  2012-07-13

5.  Analysis of linear measurements on 3D surface models using CBCT data segmentation obtained by automatic standard pre-set thresholds in two segmentation software programs: an in vitro study.

Authors:  Marcelo Lupion Poleti; Thais Maria Freire Fernandes; Otávio Pagin; Marcela Rodrigues Moretti; Izabel Regina Fischer Rubira-Bullen
Journal:  Clin Oral Investig       Date:  2015-05-13       Impact factor: 3.573

6.  Comparison of the accuracy of full head cone beam CT images obtained using a large field of view and stitched images.

Authors:  Mehmet Ozgur Ozemre; Ayse Gulsahi
Journal:  Dentomaxillofac Radiol       Date:  2018-06-19       Impact factor: 2.419

Review 7.  Applications of Cone Beam Computed Tomography in Orthodontics: A Review.

Authors:  Aditya Tadinada; Alana Marczak; Sumit Yadav; Padma M Mukherjee
Journal:  Turk J Orthod       Date:  2016-09-01

8.  Accuracy of peri-implant bone thickness and validity of assessing bone augmentation material using cone beam computed tomography.

Authors:  Dongyun Wang; Andreas Künzel; Vladimir Golubovic; Ilya Mihatovic; Gordon John; Zhuofan Chen; Jürgen Becker; Frank Schwarz
Journal:  Clin Oral Investig       Date:  2012-10-12       Impact factor: 3.573

9.  Three-dimensional localization of impacted canines and root resorption assessment using cone beam computed tomography.

Authors:  Eyad Almuhtaseb; Jing Mao; Derek Mahony; Rawan Bader; Zhi-Xing Zhang
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2014-06-18

10.  Cone-beam evaluation of pharyngeal airway space in adult skeletal Class II patients with different condylar positions.

Authors:  Jintao Xu; Ruonan Sun; Linna Wang; Xiaoying Hu
Journal:  Angle Orthod       Date:  2018-11-20       Impact factor: 2.079