Mechanical analysis of conventional and small diameter conical implant abutments.

Izabela Cristina Maurício Moris1, Adriana Cláudia Lapria Faria, Maria da Gloria Chiarello de Mattos, Ricardo Faria Ribeiro, Renata Cristina Silveira Rodrigues.   

Abstract

PURPOSE: The aim of the present study was to evaluate if a smaller morse taper abutment has a negative effect on the fracture resistance of implant-abutment connections under oblique compressive loads compared to a conventional abutment.
MATERIALS AND METHODS: Twenty morse taper conventional abutments (4.8 mm diameter) and smaller abutments (3.8 mm diameter) were tightened (20 Ncm) to their respective implants (3.5 × 11 mm) and after a 10 minute interval, implant/abutment assemblies were subjected to static compressive test, performed in a universal test machine with 1 mm/min displacement, at 45° inclination. The maximum deformation force was determined. Data were statistically analyzed by student t test.
RESULTS: Maximum deformation force of 4.8 mm and 3.8 mm abutments was approximately 95.33 kgf and 95.25 kgf, respectively, but no fractures were noted after mechanical test. Statistical analysis demonstrated that the evaluated abutments were statistically similar (P=.230).
CONCLUSION: Abutment measuring 3.8 mm in diameter (reduced) presented mechanical properties similar to 4.8 mm (conventional) abutments, enabling its clinical use as indicated.

Entities:  

Keywords:  Compressive Strength; Dental Implant Abutment Design; Dental Implants; Morse Taper Dental Implant Abutment Connection

Year:  2012        PMID: 22977724      PMCID: PMC3439626          DOI: 10.4047/jap.2012.4.3.158

Source DB:  PubMed          Journal:  J Adv Prosthodont        ISSN: 2005-7806            Impact factor:   1.904


  8 in total

1.  Fatigue resistance of ITI implant-abutment connectors -- a comparison of the standard cone with a novel internally keyed design.

Authors:  Jean Perriard; W Anselm Wiskott; Aissa Mellal; Susanne S Scherrer; John Botsis; Urs C Belser
Journal:  Clin Oral Implants Res       Date:  2002-10       Impact factor: 5.977

2.  Fatigue resistance of two implant/abutment joint designs.

Authors:  Ameen Khraisat; Roxana Stegaroiu; Shuichi Nomura; Osamu Miyakawa
Journal:  J Prosthet Dent       Date:  2002-12       Impact factor: 3.426

Review 3.  New prosthetic restorative features of Ankylos implant system.

Authors:  Paul Weigl
Journal:  J Oral Implantol       Date:  2004       Impact factor: 1.779

4.  An in vitro evaluation of the strength of a 1-piece and 2-piece conical abutment joint in implant design.

Authors:  M R Norton
Journal:  Clin Oral Implants Res       Date:  2000-10       Impact factor: 5.977

5.  Mechanics of the taper integrated screwed-in (TIS) abutments used in dental implants.

Authors:  Dinçer Bozkaya; Sinan Müftü
Journal:  J Biomech       Date:  2005-01       Impact factor: 2.712

6.  Initial clinical efficacy of 3-mm implants immediately placed into function in conditions of limited spacing.

Authors:  Michael S Reddy; S Jean O'Neal; Sandra Haigh; Ruth Aponte-Wesson; Nico C Geurs
Journal:  Int J Oral Maxillofac Implants       Date:  2008 Mar-Apr       Impact factor: 2.804

7.  Implant-abutment interface design affects fatigue and fracture strength of implants.

Authors:  Lars Steinebrunner; Stefan Wolfart; Klaus Ludwig; Matthias Kern
Journal:  Clin Oral Implants Res       Date:  2008-12       Impact factor: 5.977

8.  Fracture resistance of the implant-abutment connection in implants with internal hex and internal conical connections under oblique compressive loading: an in vitro study.

Authors:  Abílio Ricciardi Coppedê; Edmilson Bersani; Maria da Gloria Chiarello de Mattos; Renata Cristina Silveira Rodrigues; Ivete Aparecida de Mattias Sartori; Ricardo Faria Ribeiro
Journal:  Int J Prosthodont       Date:  2009 May-Jun       Impact factor: 1.681

  8 in total
  3 in total

1.  Fractures of implant fixtures: a retroscccpective clinical study.

Authors:  Han-Chang Yu; Young-Kyun Kim
Journal:  Maxillofac Plast Reconstr Surg       Date:  2020-04-25

2.  Influence of Manufacturing Methods of Implant-Supported Crowns on External and Internal Marginal Fit: A Micro-CT Analysis.

Authors:  Izabela C M Moris; Silas Borges Monteiro; Raíssa Martins; Ricardo Faria Ribeiro; Erica A Gomes
Journal:  Biomed Res Int       Date:  2018-01-23       Impact factor: 3.411

3.  Finite Element Analysis of Stress in Anterior Prosthetic Rehabilitation with Zirconia Implants with and without Cantilever.

Authors:  Karina Tiemi Sanomya Tsumanuma; Ricardo Armini Caldas; Isaias Donizeti Silva; Milton Edson Miranda; William Cunha Brandt; Rafael Pino Vitti
Journal:  Eur J Dent       Date:  2021-08-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.