Magnetostatic focal spot correction for x-ray tubes operating in strong magnetic fields using iterative optimization.

Prasheel Lillaney1, Mihye Shin, Steven M Conolly, Rebecca Fahrig.   

Abstract

PURPOSE: Combining x-ray fluoroscopy and MR imaging systems for guidance of interventional procedures has become more commonplace. By designing an x-ray tube that is immune to the magnetic fields outside of the MR bore, the two systems can be placed in close proximity to each other. A major obstacle to robust x-ray tube design is correcting for the effects of the magnetic fields on the x-ray tube focal spot. A potential solution is to design active shielding that locally cancels the magnetic fields near the focal spot.
METHODS: An iterative optimization algorithm is implemented to design resistive active shielding coils that will be placed outside the x-ray tube insert. The optimization procedure attempts to minimize the power consumption of the shielding coils while satisfying magnetic field homogeneity constraints. The algorithm is composed of a linear programming step and a nonlinear programming step that are interleaved with each other. The coil results are verified using a finite element space charge simulation of the electron beam inside the x-ray tube. To alleviate heating concerns an optimized coil solution is derived that includes a neodymium permanent magnet. Any demagnetization of the permanent magnet is calculated prior to solving for the optimized coils. The temperature dynamics of the coil solutions are calculated using a lumped parameter model, which is used to estimate operation times of the coils before temperature failure.
RESULTS: For a magnetic field strength of 88 mT, the algorithm solves for coils that consume 588 A∕cm(2). This specific coil geometry can operate for 15 min continuously before reaching temperature failure. By including a neodymium magnet in the design the current density drops to 337 A∕cm(2), which increases the operation time to 59 min. Space charge simulations verify that the coil designs are effective, but for oblique x-ray tube geometries there is still distortion of the focal spot shape along with deflections of approximately 3 mm in the radial and circumferential directions on the anode.
CONCLUSIONS: Active shielding is an attractive solution for correcting the effects of magnetic fields on the x-ray focal spot. If extremely long fluoroscopic exposure times are required, longer operation times can be achieved by including a permanent magnet with the active shielding design.

Mesh:

Year:  2012        PMID: 22957623      PMCID: PMC3436919          DOI: 10.1118/1.4742060

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Truly hybrid interventional MR/X-ray system: investigation of in vivo applications.

Authors:  R Fahrig; K Butts; Z Wen; R Saunders; S T Kee; D Y Sze; B L Daniel; F Laerum; N J Pelc
Journal:  Acad Radiol       Date:  2001-12       Impact factor: 3.173

2.  First use of a truly-hybrid X-ray/MR imaging system for guidance of brain biopsy.

Authors:  R Fahrig; G Heit; Z Wen; B L Daniel; K Butts; N J Pelc
Journal:  Acta Neurochir (Wien)       Date:  2003-11       Impact factor: 2.216

3.  A truly hybrid interventional MR/X-ray system: feasibility demonstration.

Authors:  R Fahrig; K Butts; J A Rowlands; R Saunders; J Stanton; G M Stevens; B L Daniel; Z Wen; D L Ergun; N J Pelc
Journal:  J Magn Reson Imaging       Date:  2001-02       Impact factor: 4.813

4.  Truly hybrid X-ray/MR imaging: toward a streamlined clinical system.

Authors:  Arundhuti Ganguly; Zhifei Wen; Bruce L Daniel; Kim Butts; Stephen T Kee; Viola Rieke; Huy M Do; Norbert J Pelc; Rebecca Fahrig
Journal:  Acad Radiol       Date:  2005-09       Impact factor: 3.173

5.  Compatibility of interventional x-ray and magnetic resonance imaging: feasibility of a closed bore XMR (CBXMR) system.

Authors:  Lukasz Brzozowski; Arundhuti Ganguly; Mihaela Pop; Zhifei Wen; Robert Bennett; Rebecca Fahrig; J A Rowlands
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

6.  Investigation of electron trajectories of an x-ray tube in magnetic fields of MR scanners.

Authors:  Zhifei Wen; Rebecca Fahrig; Steven Conolly; Norbert J Pelc
Journal:  Med Phys       Date:  2007-06       Impact factor: 4.071

7.  MR-guided transjugular intrahepatic portosystemic shunt creation with use of a hybrid radiography/MR system.

Authors:  Stephen T Kee; Arundhuti Ganguly; Bruce L Daniel; Zhifei Wen; Kim Butts; Anne Shimikawa; Norbert J Pelc; Rebecca Fahrig; Michael D Dake
Journal:  J Vasc Interv Radiol       Date:  2005-02       Impact factor: 3.464

8.  Closed bore XMR (CBXMR) systems for aortic valve replacement: active magnetic shielding of x-ray tubes.

Authors:  John A Bracken; Giovanni DeCrescenzo; Philip Komljenovic; Prasheel V Lillaney; Rebecca Fahrig; J A Rowlands
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

9.  Experimental renal artery embolization in a combined MR imaging/angiographic unit.

Authors:  Mark W Wilson; Nicholas Fidelman; Oliver M Weber; Alastair J Martin; Roy L Gordon; Jeanne M LaBerge; Robert K Kerlan; Kristen A Wolanske; Maythem Saeed
Journal:  J Vasc Interv Radiol       Date:  2003-09       Impact factor: 3.464

10.  Registration and tracking to integrate X-ray and MR images in an XMR facility.

Authors:  Kawal S Rhode; Derek L G Hill; Philip J Edwards; John Hipwell; Daniel Rueckert; Gerardo Sanchez-Ortiz; Sanjeet Hegde; Vithuran Rahunathan; Reza Razavi
Journal:  IEEE Trans Med Imaging       Date:  2003-11       Impact factor: 10.048

View more
  3 in total

1.  Design optimization of MR-compatible rotating anode x-ray tubes for stable operation.

Authors:  Mihye Shin; Prasheel Lillaney; Waldo Hinshaw; Rebecca Fahrig
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

2.  Electrostatic focal spot correction for x-ray tubes operating in strong magnetic fields.

Authors:  Prasheel Lillaney; Mihye Shin; Waldo Hinshaw; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

3.  Novel motor design for rotating anode x-ray tubes operating in the fringe field of a magnetic resonance imaging system.

Authors:  Prasheel Lillaney; Mihye Shin; Waldo Hinshaw; N Robert Bennett; Norbert Pelc; Rebecca Fahrig
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.