Truly hybrid X-ray/MR imaging: toward a streamlined clinical system.

Arundhuti Ganguly1, Zhifei Wen, Bruce L Daniel, Kim Butts, Stephen T Kee, Viola Rieke, Huy M Do, Norbert J Pelc, Rebecca Fahrig.   

Abstract

RATIONALE AND
OBJECTIVE: We have installed an improved X-ray/MR (XMR) truly hybrid system with higher imaging signal-to-noise ratio (SNR) and versatility than our first prototype. In our XMR design, a fixed anode X-ray fluoroscopy system is positioned between the two donut-shaped magnetic poles of a 0.5T GE Signa-SP magnet (SP-XMR). This paper describes the methods for increased compatibility between the upgraded x-ray and MR systems that have helped improve patient management.
MATERIALS AND METHODS: A GE OEC 9800 system (GE OEC Salt Lake City, UT) was specially reconfigured for permitting X-ray fluoroscopy inside the interventional magnet. A higher power X-ray tube, a new permanent tube mounting system, automatic exposure control (AEC), remote controlled collimators, choice of multiple frame rates, DICOM image compatibility, magnetically shimmed X-ray detector, X-ray compatible MR coil, and better RF shielding are the highlights of the new system. A total of 23 clinical procedures have been conducted with SP-XMR guidance of which five were performed using the new system.
RESULTS: The 70% increased power for fluoroscopy, and a new 6 times higher power single frame imaging mode, has improved imaging capability. The choice of multiple imaging frame rates, AEC, and collimator control allow reduction in X-ray exposure to the patient. The DICOM formatting has permitted easy transfer of clinical images over the hospital PACS network. The increased MR compatibility of the detector and the X-ray transparent MR coil has enabled faster switching between X-ray and MR imaging modes.
CONCLUSION: The improvements introduced in our SP-XMR system have further streamlined X-ray/MR hybrid imaging. Additional clinical procedures could benefit from the new SP-XMR imaging.

Entities:  

Mesh:

Year:  2005        PMID: 16099685     DOI: 10.1016/j.acra.2005.03.076

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  12 in total

1.  Design optimization of MR-compatible rotating anode x-ray tubes for stable operation.

Authors:  Mihye Shin; Prasheel Lillaney; Waldo Hinshaw; Rebecca Fahrig
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

2.  Shimming with permanent magnets for the x-ray detector in a hybrid x-ray/ MR system.

Authors:  Zhifei Wen; Rebecca Fahrig; Scott T Williams; Norbert J Pelc
Journal:  Med Phys       Date:  2008-09       Impact factor: 4.071

3.  Robust automatic rigid registration of MRI and X-ray using external fiducial markers for XFM-guided interventional procedures.

Authors:  Ashvin K George; Merdim Sonmez; Robert J Lederman; Anthony Z Faranesh
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

4.  Electrostatic focal spot correction for x-ray tubes operating in strong magnetic fields.

Authors:  Prasheel Lillaney; Mihye Shin; Waldo Hinshaw; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

5.  Magnetostatic focal spot correction for x-ray tubes operating in strong magnetic fields using iterative optimization.

Authors:  Prasheel Lillaney; Mihye Shin; Steven M Conolly; Rebecca Fahrig
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

6.  Novel motor design for rotating anode x-ray tubes operating in the fringe field of a magnetic resonance imaging system.

Authors:  Prasheel Lillaney; Mihye Shin; Waldo Hinshaw; N Robert Bennett; Norbert Pelc; Rebecca Fahrig
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

7.  Use of perfluorocarbon nanoparticles for non-invasive multimodal cell tracking of human pancreatic islets.

Authors:  Brad P Barnett; Jesus Ruiz-Cabello; Partha Hota; Ronald Ouwerkerk; Michael J Shamblott; Cal Lauzon; Piotr Walczak; Wesley D Gilson; Vadappuram P Chacko; Dara L Kraitchman; Aravind Arepally; Jeff W M Bulte
Journal:  Contrast Media Mol Imaging       Date:  2011 Jul-Aug       Impact factor: 3.161

8.  Feasibility of using real-time MRI to measure joint kinematics in 1.5T and open-bore 0.5T systems.

Authors:  Christine E Draper; Juan M Santos; Lampros C Kourtis; Thor F Besier; Michael Fredericson; Gary S Beaupre; Garry E Gold; Scott L Delp
Journal:  J Magn Reson Imaging       Date:  2008-07       Impact factor: 4.813

9.  Technical Note: A Monte Carlo study of magnetic-field-induced radiation dose effects in mice.

Authors:  Ashley E Rubinstein; Zhongxing Liao; Adam D Melancon; Michele Guindani; David S Followill; Ramesh C Tailor; John D Hazle; Laurence E Court
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

10.  In vivo assessment of cold adaptation in insect larvae by magnetic resonance imaging and magnetic resonance spectroscopy.

Authors:  Daniel Mietchen; Bertram Manz; Frank Volke; Kenneth Storey
Journal:  PLoS One       Date:  2008-12-05       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.