Literature DB >> 11770916

Truly hybrid interventional MR/X-ray system: investigation of in vivo applications.

R Fahrig1, K Butts, Z Wen, R Saunders, S T Kee, D Y Sze, B L Daniel, F Laerum, N J Pelc.   

Abstract

RATIONALE AND
OBJECTIVES: The purpose of this study was to provide in vivo demonstrations of the functionality of a truly hybrid interventional x-ray/magnetic resonance (MR) system.
MATERIALS AND METHODS: A digital flat-panel x-ray system (1,024(2) array of 200 microm pixels, 30 frames per second) was integrated into an interventional 0.5-T magnet. The hybrid system is capable of MR and x-ray imaging of the same field of view without patient movement. Two intravascular procedures were performed in a 22-kg porcine model: placement of a transjugular intrahepatic portosystemic shunt (TIPS) (x-ray-guided catheterization of the hepatic vein, MR fluoroscopy-guided portal puncture, and x-ray-guided stent placement) and mock chemoembolization (x-ray-guided subselective catheterization of a renal artery branch and MR evaluation of perfused volume).
RESULTS: The resolution and frame rate of the x-ray fluoroscopy images were sufficient to visualize and place devices, including nitinol guidewires (0.016-0.035-inch diameter) and stents and a 2.3-F catheter. Fifth-order branches of the renal artery could be seen. The quality of both real-time (3.5 frames per second) and standard MR images was not affected by the x-ray system. During MR-guided TIPS placement, the trocar and the portal vein could be easily visualized, allowing successful puncture from hepatic to portal vein.
CONCLUSION: Switching back and forth between x-ray and MR imaging modalities without requiring movement of the patient was demonstrated. The integrated nature of the system could be especially beneficial when x-ray and MR image guidance are used iteratively.

Entities:  

Mesh:

Year:  2001        PMID: 11770916     DOI: 10.1016/S1076-6332(03)80702-X

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  8 in total

1.  Development of subject-specific geometric spine model through use of automated active contour segmentation and kinematic constraint-limited registration.

Authors:  Catherine G Strickland; Daniel E Aguiar; Eric A Nauman; Thomas M Talavage
Journal:  J Digit Imaging       Date:  2011-10       Impact factor: 4.056

2.  Design optimization of MR-compatible rotating anode x-ray tubes for stable operation.

Authors:  Mihye Shin; Prasheel Lillaney; Waldo Hinshaw; Rebecca Fahrig
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

3.  Shimming with permanent magnets for the x-ray detector in a hybrid x-ray/ MR system.

Authors:  Zhifei Wen; Rebecca Fahrig; Scott T Williams; Norbert J Pelc
Journal:  Med Phys       Date:  2008-09       Impact factor: 4.071

4.  Adaptive 4D MR imaging using navigator-based respiratory signal for MRI-guided therapy.

Authors:  Junichi Tokuda; Shigehiro Morikawa; Hasnine A Haque; Tetsuji Tsukamoto; Kiyoshi Matsumiya; Hongen Liao; Ken Masamune; Takeyoshi Dohi
Journal:  Magn Reson Med       Date:  2008-05       Impact factor: 4.668

5.  Electrostatic focal spot correction for x-ray tubes operating in strong magnetic fields.

Authors:  Prasheel Lillaney; Mihye Shin; Waldo Hinshaw; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

6.  Magnetostatic focal spot correction for x-ray tubes operating in strong magnetic fields using iterative optimization.

Authors:  Prasheel Lillaney; Mihye Shin; Steven M Conolly; Rebecca Fahrig
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

7.  Novel motor design for rotating anode x-ray tubes operating in the fringe field of a magnetic resonance imaging system.

Authors:  Prasheel Lillaney; Mihye Shin; Waldo Hinshaw; N Robert Bennett; Norbert Pelc; Rebecca Fahrig
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

Review 8.  Emerging roles for integrated imaging modalities in cardiovascular cell-based therapeutics: a clinical perspective.

Authors:  Peter J Psaltis; Robert D Simari; Martin Rodriguez-Porcel
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-09-08       Impact factor: 9.236

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.