Electrostatic focal spot correction for x-ray tubes operating in strong magnetic fields.

Prasheel Lillaney1, Mihye Shin2, Waldo Hinshaw3, Rebecca Fahrig3.   

Abstract

PURPOSE: A close proximity hybrid x-ray/magnetic resonance (XMR) imaging system offers several critical advantages over current XMR system installations that have large separation distances (∼5 m) between the imaging fields of view. The two imaging systems can be placed in close proximity to each other if an x-ray tube can be designed to be immune to the magnetic fringe fields outside of the MR bore. One of the major obstacles to robust x-ray tube design is correcting for the effects of the MR fringe field on the x-ray tube focal spot. Any fringe field component orthogonal to the x-ray tube electric field leads to electron drift altering the path of the electron trajectories.
METHODS: The method proposed in this study to correct for the electron drift utilizes an external electric field in the direction of the drift. The electric field is created using two electrodes that are positioned adjacent to the cathode. These electrodes are biased with positive and negative potential differences relative to the cathode. The design of the focusing cup assembly is constrained primarily by the strength of the MR fringe field and high voltage standoff distances between the anode, cathode, and the bias electrodes. From these constraints, a focusing cup design suitable for the close proximity XMR system geometry is derived, and a finite element model of this focusing cup geometry is simulated to demonstrate efficacy. A Monte Carlo simulation is performed to determine any effects of the modified focusing cup design on the output x-ray energy spectrum.
RESULTS: An orthogonal fringe field magnitude of 65 mT can be compensated for using bias voltages of +15 and -20 kV. These bias voltages are not sufficient to completely correct for larger orthogonal field magnitudes. Using active shielding coils in combination with the bias electrodes provides complete correction at an orthogonal field magnitude of 88.1 mT. Introducing small fields (<10 mT) parallel to the x-ray tube electric field in addition to the orthogonal field does not affect the electrostatic correction technique. However, rotation of the x-ray tube by 30° toward the MR bore increases the parallel magnetic field magnitude (∼72 mT). The presence of this larger parallel field along with the orthogonal field leads to incomplete correction. Monte Carlo simulations demonstrate that the mean energy of the x-ray spectrum is not noticeably affected by the electrostatic correction, but the output flux is reduced by 7.5%.
CONCLUSIONS: The maximum orthogonal magnetic field magnitude that can be compensated for using the proposed design is 65 mT. Larger orthogonal field magnitudes cannot be completely compensated for because a pure electrostatic approach is limited by the dielectric strength of the vacuum inside the x-ray tube insert. The electrostatic approach also suffers from limitations when there are strong magnetic fields in both the orthogonal and parallel directions because the electrons prefer to stay aligned with the parallel magnetic field. These challenging field conditions can be addressed by using a hybrid correction approach that utilizes both active shielding coils and biasing electrodes.

Mesh:

Year:  2014        PMID: 25370658      PMCID: PMC4209011          DOI: 10.1118/1.4898099

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  26 in total

Review 1.  Cost-efficacy of MR-guided neurointerventions.

Authors:  J Kucharczyk; W A Hall; W C Broaddus; G T Gillies; C L Truwit
Journal:  Neuroimaging Clin N Am       Date:  2001-11       Impact factor: 2.264

Review 2.  Minimally invasive procedures. Interventional MR image-guided functional neurosurgery.

Authors:  R M Chu; R P Tummala; J Kucharczyk; C L Truwit; R E Maxwell
Journal:  Neuroimaging Clin N Am       Date:  2001-11       Impact factor: 2.264

3.  First use of a truly-hybrid X-ray/MR imaging system for guidance of brain biopsy.

Authors:  R Fahrig; G Heit; Z Wen; B L Daniel; K Butts; N J Pelc
Journal:  Acta Neurochir (Wien)       Date:  2003-11       Impact factor: 2.216

4.  Robust x-ray tubes for use within magnetic fields of MR scanners.

Authors:  Zhifei Wen; Rebecca Fahrig; Norbert J Pelc
Journal:  Med Phys       Date:  2005-07       Impact factor: 4.071

5.  Performance of a static-anode/flat-panel x-ray fluoroscopy system in a diagnostic strength magnetic field: a truly hybrid x-ray/MR imaging system.

Authors:  R Fahrig; Z Wen; A Ganguly; G DeCrescenzo; J A Rowlands; G M Stevens; R F Saunders; N J Pelc
Journal:  Med Phys       Date:  2005-06       Impact factor: 4.071

6.  Truly hybrid X-ray/MR imaging: toward a streamlined clinical system.

Authors:  Arundhuti Ganguly; Zhifei Wen; Bruce L Daniel; Kim Butts; Stephen T Kee; Viola Rieke; Huy M Do; Norbert J Pelc; Rebecca Fahrig
Journal:  Acad Radiol       Date:  2005-09       Impact factor: 3.173

7.  A fast-marching approach to cardiac electrophysiology simulation for XMR interventional imaging.

Authors:  M Sermesant; Y Coudière; V Moreau-Villèger; K S Rhode; D L G Hill; R S Razavi
Journal:  Med Image Comput Comput Assist Interv       Date:  2005

8.  MR-guided transjugular intrahepatic portosystemic shunt creation with use of a hybrid radiography/MR system.

Authors:  Stephen T Kee; Arundhuti Ganguly; Bruce L Daniel; Zhifei Wen; Kim Butts; Anne Shimikawa; Norbert J Pelc; Rebecca Fahrig; Michael D Dake
Journal:  J Vasc Interv Radiol       Date:  2005-02       Impact factor: 3.464

9.  Experimental renal artery embolization in a combined MR imaging/angiographic unit.

Authors:  Mark W Wilson; Nicholas Fidelman; Oliver M Weber; Alastair J Martin; Roy L Gordon; Jeanne M LaBerge; Robert K Kerlan; Kristen A Wolanske; Maythem Saeed
Journal:  J Vasc Interv Radiol       Date:  2003-09       Impact factor: 3.464

10.  Registration and tracking to integrate X-ray and MR images in an XMR facility.

Authors:  Kawal S Rhode; Derek L G Hill; Philip J Edwards; John Hipwell; Daniel Rueckert; Gerardo Sanchez-Ortiz; Sanjeet Hegde; Vithuran Rahunathan; Reza Razavi
Journal:  IEEE Trans Med Imaging       Date:  2003-11       Impact factor: 10.048

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.