Clinical performance of 2 dedicated PET scanners for breast imaging: initial evaluation.

Mami Iima1, Yuji Nakamoto, Shotaro Kanao, Tomoharu Sugie, Takayuki Ueno, Mayumi Kawada, Yoshiki Mikami, Masakazu Toi, Kaori Togashi.   

Abstract

UNLABELLED: The purpose of this study was to investigate the diagnostic performance of 2 newly developed dedicated breast PET scanners in patients with known or suspected breast cancer.
METHODS: Two types of scanner were evaluated, an O-shaped scanner and a C-shaped scanner. The O scanner was designed for imaging patients who were prone, and the C scanner was designed for those patients positioned leaning forward. Sixty-nine women with known or suspected breast carcinoma (80 lesions: 72 invasive carcinomas, 4 noninvasive carcinomas [ductal carcinoma in situ, or DCIS], 1 case of adenomatous ductal hyperplasia, and 3 benign lesions) were enrolled in this study. All patients underwent a conventional whole-body PET/CT scan, followed by breast scanning using both dedicated devices. The diagnostic performance of each scanner was assessed.
RESULTS: The maximal diameter of invasive tumors ranged from 4 to 112 mm, with an average of 26 mm. With the O scanner, 62 of 76 malignant lesions (including 3 DCIS) were detected, 5 lesions were not detected, and the remaining 9 lesions were outside the field of view. With the C scanner, 63 of 76 malignant lesions (including 2 DCIS) were detected, 7 lesions were not detected, and the remaining 6 lesions were outside the field of view. The lesion-based sensitivities of the O and C scanners were 82% (62/76) and 83% (63/76), respectively; sensitivities excluding lesions outside the field of view were 93% (62/67) and 90% (63/70), respectively. The sensitivity of conventional PET/CT was 92% (70/76). All lesions outside the field of view were close to the chest wall. The breast-based specificities of the O, C, and conventional scanners were 98% (48/49), 98% (56/57), and 100% (70/70), respectively.
CONCLUSION: Our preliminary study indicates that both dedicated breast PET scanners are clinically feasible and yield reasonably high sensitivity. More detailed information was obtained with these scanners than with the conventional scanner.

Entities:  

Mesh:

Year:  2012        PMID: 22933819     DOI: 10.2967/jnumed.111.100958

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  16 in total

Review 1.  Innovations in Instrumentation for Positron Emission Tomography.

Authors:  Eric Berg; Simon R Cherry
Journal:  Semin Nucl Med       Date:  2018-03-12       Impact factor: 4.446

Review 2.  Use of Breast-Specific PET Scanners and Comparison with MR Imaging.

Authors:  Deepa Narayanan; Wendie A Berg
Journal:  Magn Reson Imaging Clin N Am       Date:  2018-05       Impact factor: 2.266

Review 3.  Nuclear imaging of the breast: translating achievements in instrumentation into clinical use.

Authors:  Carrie B Hruska; Michael K O'Connor
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

4.  Effects of reflector and crystal surface on the performance of a depth-encoding PET detector with dual-ended readout.

Authors:  Silin Ren; Yongfeng Yang; Simon R Cherry
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

5.  Design Optimization of a TOF, Breast PET Scanner.

Authors:  Eunsin Lee; Matthew E Werner; Joel S Karp; Suleman Surti
Journal:  IEEE Trans Nucl Sci       Date:  2013-06       Impact factor: 1.679

6.  Effects of Detector Thickness on Geometric Sensitivity and Event Positioning Errors in the Rectangular PET/X Scanner.

Authors:  Lawrence R MacDonald; William C J Hunter; Paul E Kinahan; Robert S Miyaoka
Journal:  IEEE Trans Nucl Sci       Date:  2013-10       Impact factor: 1.679

7.  A dedicated breast-PET/CT scanner: Evaluation of basic performance characteristics.

Authors:  Raymond R Raylman; Will Van Kampen; Alexander V Stolin; Wenbo Gong; Gangadhar Jaliparthi; Peter F Martone; Mark F Smith; David Sarment; Neal H Clinthorne; Mark Perna
Journal:  Med Phys       Date:  2018-02-23       Impact factor: 4.071

Review 8.  Radionuclide methods and instrumentation for breast cancer detection and diagnosis.

Authors:  Suleman Surti
Journal:  Semin Nucl Med       Date:  2013-07       Impact factor: 4.446

9.  Preliminary performance characterization of DbPET2.1, a PET scanner dedicated to the imaging of the breast and extremities.

Authors:  Andrea Ferrero; Qiyu Peng; George W Burkett; Buddika Sumanasena; William W Moses; Ramsey D Badawi
Journal:  Biomed Phys Eng Express       Date:  2015-06-09

Review 10.  Dedicated Breast Gamma Camera Imaging and Breast PET: Current Status and Future Directions.

Authors:  Deepa Narayanan; Wendie A Berg
Journal:  PET Clin       Date:  2018-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.