A quantitative study of motion estimation methods on 4D cardiac gated SPECT reconstruction.

Wenyuan Qi1, Yongyi Yang, Xiaofeng Niu, Michael A King.   

Abstract

PURPOSE: Motion-compensated temporal processing can have a major impact on improving the image quality in gated cardiac single photon emission computed tomography (SPECT). In this work, we investigate the effect of different optical flow estimation methods for motion-compensated temporal processing in gated SPECT. In particular, we explore whether better motion estimation can substantially improve reconstructed image quality, and how the estimated motion would compare to the ideal case of known motion in terms of reconstruction.
METHODS: We consider the following three methods for obtaining the image motion in 4D reconstruction: (1) the Horn-Schunck optical flow equation (OFE) method, (2) a recently developed periodic OFE method, and (3) known cardiac motion derived from the NURBS-based cardiac-torso (NCAT) phantom. The periodic OFE method is used to exploit the inherent periodic nature in cardiac gated images. In this method, the optical flow in a sequence is modeled by a Fourier harmonic representation, which is then estimated from the image data. We study the impact of temporal processing on 4D reconstructions when the image motion is obtained with the different methods above. For quantitative evaluation, we use simulated imaging with multiple noise realizations from the NCAT phantom, where different patient geometry and lesion sizes are also considered. To quantify the reconstruction results, we use the following measures of reconstruction accuracy and defect detection in the myocardium: (1) overall error level in the myocardium, (2) regional accuracy of the left ventricle (LV) wall, (3) accuracy of regional time activity curves of the LV, and (4) perfusion defect detectability with a channelized Hotelling observer (CHO). In addition, we also examine the effect of noise on the distortion in the reconstructed LV wall shape by detecting its contours. As a preliminary demonstration, these methods are also tested on two sets of clinical acquisitions.
RESULTS: For the different quantitative measures considered, the periodic OFE further improved the reconstruction accuracy of the myocardium compared to OFE in 4D reconstruction; its improvement in reconstruction almost matched that of the known motion. Specifically, the overall mean-squared error in the myocardium was reduced by over 20% with periodic OFE; with noise level fixed at 10%, the regional bias on the LV was reduced from 20% (OFE) to 14% (periodic OFE), compared to 11% by the known motion. In addition, the CHO results show that there was also improvement in lesion detectability with the periodic OFE. The regional time activity curves obtained with the periodic OFE were also observed to be more consistent with the reference; in addition, the contours of the reconstructed LV wall with the periodic OFE were demonstrated to show less degree of variations among different noise realizations. Such improvements were also consistent with the results obtained from the clinical acquisitions.
CONCLUSIONS: Use of improved optical flow estimation can further improve the accuracy of reconstructed images in 4D. The periodic OFE method not only can achieve improvements over the traditional OFE, but also can almost match that of the known motion in terms of the several quality measures considered.

Entities:  

Mesh:

Year:  2012        PMID: 22894443      PMCID: PMC3422358          DOI: 10.1118/1.4738377

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Effects of motion, attenuation, and scatter corrections on gated cardiac SPECT reconstruction.

Authors:  Xiaofeng Niu; Yongyi Yang; Mingwu Jin; Miles N Wernick; Michael A King
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Spatiotemporal processing of gated cardiac SPECT images using deformable mesh modeling.

Authors:  Jovan G Brankov; Yongyi Yang; Miles N Wernick
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

3.  Fast gradient-based algorithms for constrained total variation image denoising and deblurring problems.

Authors:  Amir Beck; Marc Teboulle
Journal:  IEEE Trans Image Process       Date:  2009-07-24       Impact factor: 10.856

4.  A computational approach to edge detection.

Authors:  J Canny
Journal:  IEEE Trans Pattern Anal Mach Intell       Date:  1986-06       Impact factor: 6.226

5.  Block-iterative techniques for fast 4D reconstruction using a priori motion models in gated cardiac SPECT.

Authors:  D S Lalush; B M Tsui
Journal:  Phys Med Biol       Date:  1998-04       Impact factor: 3.609

6.  A Monte Carlo program for the simulation of scintillation camera characteristics.

Authors:  M Ljungberg; S E Strand
Journal:  Comput Methods Programs Biomed       Date:  1989-08       Impact factor: 5.428

7.  Addition of a channel mechanism to the ideal-observer model.

Authors:  K J Myers; H H Barrett
Journal:  J Opt Soc Am A       Date:  1987-12       Impact factor: 2.129

8.  Use of a nonstationary temporal Wiener filter in nuclear medicine.

Authors:  M A King; T R Miller
Journal:  Eur J Nucl Med       Date:  1985

9.  Optical flow estimation for a periodic image sequence.

Authors:  Ling Li; Yongyi Yang
Journal:  IEEE Trans Image Process       Date:  2010-01       Impact factor: 10.856

10.  List-mode-based reconstruction for respiratory motion correction in PET using non-rigid body transformations.

Authors:  F Lamare; M J Ledesma Carbayo; T Cresson; G Kontaxakis; A Santos; C Cheze Le Rest; A J Reader; D Visvikis
Journal:  Phys Med Biol       Date:  2007-08-09       Impact factor: 3.609

  10 in total
  7 in total

1.  Limited-angle effect compensation for respiratory binned cardiac SPECT.

Authors:  Wenyuan Qi; Yongyi Yang; Miles N Wernick; P Hendrik Pretorius; Michael A King
Journal:  Med Phys       Date:  2016-01       Impact factor: 4.071

2.  Fast spatiotemporal image reconstruction based on low-rank matrix estimation for dynamic photoacoustic computed tomography.

Authors:  Kun Wang; Jun Xia; Changhui Li; Lihong V Wang; Mark A Anastasio
Journal:  J Biomed Opt       Date:  2014-05       Impact factor: 3.170

3.  4D numerical observer for lesion detection in respiratory-gated PET.

Authors:  Auranuch Lorsakul; Quanzheng Li; Cathryn M Trott; Christopher Hoog; Yoann Petibon; Jinsong Ouyang; Andrew F Laine; Georges El Fakhri
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

4.  Noise suppressed partial volume correction for cardiac SPECT/CT.

Authors:  Chung Chan; Hui Liu; Yariv Grobshtein; Mitchel R Stacy; Albert J Sinusas; Chi Liu
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

5.  4D non-local means post-filtering for cardiac gated SPECT.

Authors:  Chao Song; Yongyi Yang; P Hendrik Pretorius; Michael A King
Journal:  Phys Med Biol       Date:  2018-01-31       Impact factor: 3.609

6.  4-D Reconstruction With Respiratory Correction for Gated Myocardial Perfusion SPECT.

Authors:  Wenyuan Qi; Yongyi Yang; Chao Song; Miles N Wernick; P Hendrik Pretorius; Michael A King
Journal:  IEEE Trans Med Imaging       Date:  2017-04-04       Impact factor: 10.048

7.  Discrete imaging models for three-dimensional optoacoustic tomography using radially symmetric expansion functions.

Authors:  Kun Wang; Robert W Schoonover; Richard Su; Alexander Oraevsky; Mark A Anastasio
Journal:  IEEE Trans Med Imaging       Date:  2014-05       Impact factor: 10.048

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.