Hearing threshold prediction with Auditory Steady State Responses and estimation of correction functions to compensate for differences with behavioral data, in adult subjects. Part 1: Audera and CHARTR EP devices.

Stavros Hatzopoulos1, Joseph Petruccelli, Lech Śliwa, Wiesław W Jędrzejczak, Krzysztof Kochanek, Henryk Skarżyński.   

Abstract

BACKGROUND: The objective of the study was the evaluation and comparison of hearing threshold values extrapolated from Auditory Steady State Responses, using 2 commercial systems and the estimation of correction factors applicable to the ASSR data. MATERIAL/
METHODS: One hundred ten subjects participated to the study. All subjects were initially examined with otoscopy, pure-tone audiometry and admittance. Data were acquired by 2 clinical systems the Audera (Viasys) and the CHARTR EP (ICS), using identical protocols. The acoustic stimuli consisted of single carrier frequencies at 1000, 2000 and 4000 Hz modulated at 40 Hz.
RESULTS: The data show that the threshold estimates from both devices differ significantly from the measured behavioral thresholds. The ICS device presented significantly larger mean-ASSR estimated hearing level values at the tested frequencies, implying an underestimation of the hearing threshold. Both sets of prediction errors overestimated hearing levels for the normal group. The prediction errors were in all cases greater for the Audera than for the ICS.
CONCLUSIONS: The errors encountered in the estimates of the 2 widely-used commercial devices suggest that the current ASSR protocols are not ready for a wide-range use and that significant developments in the area of threshold prediction/precision are necessary. If, on the other-hand, the ASSR predicted threshold is used on a purely consulting basis, as in hearing-aid fitting, then such errors might be acceptable in a clinical setting.

Entities:  

Mesh:

Year:  2012        PMID: 22739744      PMCID: PMC3560776          DOI: 10.12659/msm.883195

Source DB:  PubMed          Journal:  Med Sci Monit        ISSN: 1234-1010


Background

The last decade auditory steady state responses (ASSRs) have been proposed as a clinical tool for use in gaining more information about the hearing threshold, especially at the low frequencies threshold [1,2]. The ASSRs are recorded in response to pure tones, which are modulated in amplitude or frequency, and are frequency-specific. It is assumed that the electrical sources of the ASSR arise from multiple sources, partly in the cortex, partly in the brainstem, with different contributions depending on the temporal pattern of the stimuli [3-5]. Recent studies have demonstrated that the ASSR responses can be recorded at each frequency, with stimuli at levels close to the behavioral threshold [1,2,6,7]. This represents an important development in pediatric audiology, given the possibility that this technique could evaluate hearing in the low-middle frequency range. Indeed, as the most popular evoked responses (auditory brainstem response) are elicited by acoustic transients, they may precisely estimate the hearing at 2–4 kHz only. ASSR has multiple applications with regard to hearing screenings and diagnostic audiology. ASSR facilitates a rapid and statistically valid estimate of hearing thresholds, and can add significant threshold information while evaluating patients unable or unwilling to participate in behavioral tests [2]. The relationship between hearing threshold and ASSR is quite complicated and depends on numerous parameters such as the hearing level, the audiometric frequency, the ASSR modulation frequency, and the ASSR detection algorithm, which is the algorithm that provides the threshold values from ASSR values. Data from the literature suggest that ASSR protocols can predict hearing levels from subjects with hearing deficits in the frequencies 2000 and 4000 Hz, while threshold prediction in normal hearing adults presents significant errors [8-11]. In a previous paper [12], we suggested that the difference between ASSR and behavioral threshold significantly decreases with the amount of hearing loss. The relationship followed a linear form of the type y=ax+30, where y is the ASSR threshold and x the behavioral threshold. Our data showed that for a10 dB increment of the PTA behavioral threshold, the ASSR threshold increased by 7 dB. This means that the difference between the 2 measurements (30 dB as predicted by the intercept) tends to cancel-out when the hearing loss reaches 95–100 dB (severe-to-profound hearing loss cases). In this context it is advantageous to examine the clinical performance of ASSR threshold estimation in cases presenting normal hearing levels, or mild-to-moderate hearing losses. The objective of the study was two-fold. First, the hearing threshold was assessed with 2 widely-used clinical ASSR devices and inter-subject and inter-group differences were calculated. Second, to improve the clinical applicability of the ASSR estimated hearing levels, correction functions were computed from the available behavioral data and applied to the ASSR values.

Material and Methods

Subjects

One hundred ten subjects – 37 males and 73 females – participated to the study. All subjects were initially examined with otoscopy, pure-tone audiometry and admittance. Eighty-five subjects presented normal hearing thresholds (ie, <20 dB HL) at the frequencies 125 to 8000 Hz and 25 presented steeply sloping, high-frequency, sensorineural hearing losses (>25 dB HL) at 4000 and 8000 Hz. The behavioral hearing thresholds were obtained with the TDH 49 earphone transducer.

ASSR recordings

All subjects were tested in a supine position within a sound-treated and electrically shielded room. The only instructions were to relax and to avoid excessive movements. Data were acquired by 2 clinical systems – the Audera (Viasys) and the CHARTR EP (ICS), using identical protocols. Additional details on the Audera protocol can be found in a previous publication [12]. The steady state potentials were collected by means of surface electrodes, with a vertex-mastoid montage, ipsilateral to the stimulated ear. The electrode impedance prior to the recordings was less than 5 kΩ. The acoustic stimuli consisted of single carrier frequencies at 1000, 2000, and 4000 Hz modulated at 40 Hz. The stimulus intensity was initially set to 30–40 dB above the behavioral threshold, and then changed by descending and ascending steps of 5 and 10 dB, respectively, until the minimal intensity corresponding to a significant ASSR was determined.

Statistical analyses

Since left and right ears showed ASSR responses with very high positive correlations (>0.9 in most cases), it was decided to average the L-R values and weight observations for all subjects equally. For each tested frequency, the pairwise differences for each subject were taken and 95% and 99% confidence intervals for the mean difference were computed. Three sets of difference data were generated: (i) Audera – PTA; (ii) ICS-PTA; and (iii) Audera-ICS. For the correction of the obtained ASSR data, a statistical 3-stage approach was devised and the full details are reported in the Appendix.

Results

Audera and PTA measurements

Pearson correlation estimates and stepdown-Bonferroni-adjusted p-values were obtained for the ASSR and PTA measurements, for both normal and hearing impaired subjects. For the normal group, significant and very significant correlations were observed at 1 and 2 kHz (−0.4611 p=0.0234, −0.6961, p<0.0001), while at 4 kHz the correlation was not significant (−0.3632, p=0.1064). For the hearing impaired group all correlations were very significant (0.8418, p<0.0001; 0.9139, p<0.0001; 0.9158, p<0.0001), with the highest value at 4 kHz. For each tested frequency, the pairwise differences between the pure tone audiometry hearing thresholds and the Audera-derived hearing levels for each subject were taken, and level 95% and 99% confidence intervals for the mean difference were computed. All intervals showed a higher mean Audera value compared with mean PTA. The data are summarized in Table 1A.
Table 1

Comparison of hearing levels between the ASSR devices and behavioral values: (A) Comparison between Audera and behavioral values (PTA). The second column reports the tested frequencies. The other two columns report the confidence intervals of the mean differences at 95% and 99% accuracy. Starred values indicate statistically significant differences. (B) Comparison between the ICS and behavioral values. The format follows that of (A). (C) Comparison between the Audera and the ICS estimated hearing levels. The format follows that of (A).

Frequency95% interval99% Interval
Audera vs. PTA: A1 kHz−7.57 to −3.14*−8.31 to −2.41**
2 kHz−9.45 to −3.68*−10.40 to −2.72**
4 kHz−37.23 to −28.93*−38.58 to −27.58**
ICS vs. PTA: B1 kHz23.06 to 28.75*22.13 to 29.68**
2 kHz23.84 to 30.56*22.74 to 31.66**
4 kHz28.93 to 37.23*27.58 to 38.58**
Audera vs. ICS: C1 kHz−7.57 to −3.14*−8.31 to −2.41**
2 kHz−9.45 to −3.68*−10.40 to −2.72**
4 kHz−37.23 to −28.93*−38.58 to −27.58**

ICS and PTA measurements

Pearson correlation estimates and stepdown-Bonferroni-adjusted p-values were obtained for the ASSR and PTA measurements, for both normal and hearing impaired subjects. For the normal group, no significant correlation values were observed at 1 and 4 kHz (−0.2947 p=0.0801, −0.0398 p=0.7648), while very significant values were observed at 2 kHz (−0.3773 p=0.0080). For the hearing impaired group, all correlations were very significant (0.8191, p<0.0001; 0.9123, p<0.0001; 0.8868, p<0.0001), with the highest value at 2 kHz. As in the previous section, significant and very significant mean pairwise differences were observed between the pure tone audiometry hearing thresholds and the ICS derived hearing levels, at all tested frequencies. All intervals show a higher mean ICS value compared with mean PTA. The data are summarized in Table 1B.

Audera and ICS measurements

The mean pairwise differences in hearing thresholds from the 2 devices were found to be statistically different in all 3 tested frequencies. All intervals showed a higher mean ICS value compared with mean AUDERA. The data are summarized in Table 1C.

Correction functions and the Audera, ICS-derived hearing thresholds

The Appendix describes the 3-stage procedure used to obtain correction functions for the Audera and ICS-derived hearing thresholds. The problem can be thought of as a prediction problem: how can one best predict the PTA threshold from an “Audera or ICS” measurement? The reason for the chosen procedure can perhaps best be explained by looking at Figures 1–6, which show the PTA versus Audera and ICS thresholds. The black plusses show values for the subjects with normal hearing, the gray circles for the hearing impaired. A simple regression function (linear or nonlinear) of the PTA on the Audera values would provide a prediction that would be highly dependent on the relative numbers of normal and impaired subjects in the study. To avoid this problem, the likelihood of an impaired subject being given the Audera or ICS, the threshold was first estimated. This quantity would be less dependent on the normal/impaired mix in the sample. Prediction functions for each group (normal and impaired) were computed by simple linear regression, and the final prediction was an average of these weighted by the likelihood of a subject being impaired or normal. The grey curves in Figures 1–3 show the resulting prediction functions, which serve as correction functions for the device at various frequency combinations. Table 2 summarizes these data. The first column of Table 2 shows the ASSR-predicted hearing level, while the other columns show the correction of this value per frequency and per device. For example, for a predicted hearing level value of 50 dB HL, the corrected values for Audera are: 29.67, 24,23 and 21.05 dB HL for 1.0, 2.0 and 4.0 kHz, respectively. For ICS the corrected values are 13.81, 9.86, 9,13 dB HL for 1.0, 2.0 and 4.0 kHz, respectively. The data show that the adjustments were more severe for ICS than for Audera.
Figure 1

Audera correction function for 1 kHz responses: Both axis represent values in dB HL.

Figure 6

ICS correction function for 4 kHz responses: Both axis represent values in dB HL.

Figure 3

Audera correction function for 4 kHz responses: Both axis represent values in dB HL.

Table 2

Observed and corrected ASSR estimated hearing levels, per frequency and device.

Observed ASSR thresholdCorrected ASSR Threshold
AuderaICS Chartr EP
1 kHz2 kHz4 kHz1 kHz2 kHz4 kHz
1010.0210.239.077.147.174.76
157.798.078.356.106.224.71
205.886.237.765.185.404.70
254.945.077.444.514.794.77
306.005.157.604.284.514.96
359.847.068.614.784.695.35
4015.9911.1110.946.345.516.06
4522.9817.0915.039.307.187.25
5029.6724.2321.0513.819.869.13
5535.7331.7128.6119.7213.6511.95
6041.2538.9536.9326.6118.5815.96
6546.4245.7445.1633.9424.4921.29
7051.3852.0752.7841.2431.1727.91
7556.2358.0059.6248.2238.3235.55
8061.0163.6565.7754.7745.6443.78
8565.7769.1171.3660.9052.9052.13
9070.5074.4376.5766.6859.9660.22
9575.2379.6781.5272.1766.7267.82
10079.9684.8586.2977.4573.1774.83
10584.6890.0090.9582.5779.3181.29
11089.4095.1295.5487.5985.1887.27

Discussion

The objective of the study was the evaluation of ASSR-estimated hearing levels at 1, 2 and 4 kHz, obtained with 2 commercial devices. The data show that the threshold estimates from both devices differ significantly from the measured behavioral thresholds. This fact suggests that the ASSR threshold values cannot be as useful as intended in the evaluation of hearing impairment in children and adults. One way to address the high-variance threshold estimates from the various ASSR devices is to introduce a correction procedure. Similar procedures are already incorporated in the majority of ASSR equipment, and the details of these procedures are not available to the public. It is assumed that the correction function is based on a regression-fit approach, which is very popular in the estimation of the hearing threshold as well. This approach presents a strong bias related to the data used to obtain the regression estimates. In this context, samples that well represent the properties of both groups (probably large in size) are required to calibrate the recorded ASSR potentials and extrapolate the hearing threshold values. Regarding sample sizes, large representative samples are preferable to small representative samples and to unrepresentative samples of any size, but small representative samples are preferable to unrepresentative samples of any size. In this study we addressed the issue of the large sample size requirements and its “population representativeness” with a 3-step procedure. Nevertheless, the accuracy of the calculated correction functions depends on the features of the ASSR responses. A key feature of the proposed procedure is the estimation of the likelihood of a subject presenting sensorineural deficits, given the Audera or ICS threshold measurement. Such a subject might have normal behavioral hearing levels at the low frequencies (ie, ≤1 kHz) and pronounced deficits in the midrange to high frequencies. The ASSR profile of such a case contains normative and hearing loss data. Summing up the profiles of all included cases, one can observe an overlap of frequency features in the low frequencies, which cause a bias in the estimated correction function. For example, Table 2 shows that for the observed Audera ASSR values at 1 kHz from 15 to 35 dB HL, the corrected values are :<=≤than 10 dB. To assess the quality of prediction, in-sample and cross-validated (leave 1 out) prediction errors of the form (y-E(y|x)) were computed for all the prediction functions (see Appendix) As expected, both sets of prediction errors display little bias for the combined data, but overestimate hearing levels for the normal group. The observed over-and under-estimates are more severe when the 2 groups overlap substantially in terms of the Audera or ICS measurements, as they do in Figure 6, and less so when there is less overlap, as in Figure 3. This is to be expected, since in that region it is hard to tell if the subject is normal or impaired. The prediction errors were in all cases greater for the Audera than for the ICS. There are numerous references in the literature [9-11,13-15] related to the relationship between the ASSR potentials, the extrapolated hearing levels and the behavioral pure tone thresholds. There is a consensus on the good correlation between the 2 measurements (ie, ASSR, PTA) for 2 and 4 kHz, whereas at the lower frequencies (0.5, 1. kHz) the agreement varies according to the study and the data analyzed. In addition, a number of studies [11,14] have used the correlation-metric to assess the relationship between ASSR potentials and behavioral hearing levels. The interpretation of the fact that 2 measurements are correlated can cause some confusion and lead to erroneous conclusions. The correlation suggests a degree of relationship between 2 variables, where the main issue is whether the ASSR estimate is accurate and precise. This scenario is well presented in the data reported by Ozdek et al. [14], where the ASSR responses were found to be well correlated to the respective behavioral hearing values, but significant mean differences were observed between the 2 sets of measurements. The data from the present study follow the pattern of the data reported by Ozdek et al. One of the difficulties in assessing data from previous studies is the fact that the protocols evoking the ASSR responses are quite different in terms of AM or FM carrier modulation in terms of the modulation frequency (higher modulation rates result in smaller ASSR amplitudes and lower threshold values), in terms of multiple or single frequency stimulation, and in terms of electrode montage. The latter was found to be the main factor for discrepancies on the ASSR-behavioral threshold relationship, as reported in various studies [15]. The present set of data show that the precision of the threshold estimates from the Audera and ICS devices depends (i) on the type of hearing impairment, as is well accepted in the literature [1,2]; and (ii) on the algorithm employed in the data extrapolation from “ASSR potential” to hearing threshold. The data from Results section 3 demonstrate that the ICS device presents significantly larger mean-ASSR estimated hearing level values at the tested frequencies, implying an underestimation of the hearing threshold. In this context, the Audera algorithm seems to be more precise, but the fact that the estimates from both devices were found to be significantly different from the corresponding behavioral values makes a precise evaluation of the extrapolation algorithm in the Audera device difficult. The developed correction functions offer the possibility to correct the estimated hearing levels of Audera and Chartr EP devices at frequencies which are normally required for hearing aid fitting. Nevertheless, 2 factors must be considered: (i) the corrections are protocol-dependent (modulation of 40 Hz) and additional testing is necessary to assess the performance of the correction procedure on ASSR potentials recorded with different modulation frequencies (>40 Hz); and (ii) the correction functions are not linear, especially in the low hearing level range, suggesting the need for additional sampling and additional studies to resolve issues in assessing subjects with normal (10–20 dB HL) and mild (25–40) hearing level ranges.

Conclusions

The errors encountered in the estimates of the 2 widely-used commercial devices suggest that the current ASSR protocols are not ready for wide-range use and that significant developments in the area of threshold prediction/precision are necessary. If, on the other-hand, the ASSR predicted threshold is used on a purely consulting basis, as in hearing-aid fitting, then such errors might be acceptable in a clinical setting.
Table 3

Model estimated-parameters.

β0β1δ0δ1γ0γ0
p1/a1−5.2270.15414.754−0.435−14.4350.944
p2/a2−4.3650.11015.217−0.434−16.7951.018
p4/a4−5.4870.10710.703−0.160−2.6090.894
p1/i1−4.6830.0919.606−0.212−17.0740.956
p2/i2−4.0750.0679.698−0.206−16.3350.956
p4/i4−5.7270.0804.955−0.017−7.3900.896
Table 4

In-sample prediction errors: means (std. deviation).

FitOverallNormal groupImpaired group
p1/a10.0000 (9.407)−1.597 (7.835)3.372 (11.616)
p2/a20.0000 (11.117)−2.986 (10.100)6.303 (10.764)
p4/a40.0000 (12.584)−4.471 (9.115)9.438 (13.875)
p1/i10.0000 (10.118)−2.610 (7.585)9.424 (12.535)
p2/i20.0000 (11.779)−3.446 (8.563)12.444 (13.559)
p4/i40.0000 (16.136)−5.713 (11.141)21.843 (13.492)
Table 5

Cross-validated prediction errors: means (std. deviation).

FitOverallNormal groupImpaired group
p1/a10.069 (10.182)−1.743 (8.339)3.896 (12.697)
p2/a2−0.059 (11.932)−3.236 (10.847)6.647 (11.590)
p4/a40.053 (13.399)−4.729 (9.638)10.147 (14.821)
p1/i10.064 (10.746)−2.707 (7.851)10.070 (13.770)
p2/i20.027 (12.398)−3.573 (8.873)13.027 (14.716)
p4/i4−0.031 (16.780)−5.928 (11.627)22.5193 (14.251)
  15 in total

1.  Auditory steady-state responses (ASSR): effects of modulation and carrier frequencies.

Authors:  Charles Petitot; Lionel Collett; John D Durrant
Journal:  Int J Audiol       Date:  2005-10       Impact factor: 2.117

2.  Threshold prediction using the auditory steady-state response and the tone burst auditory brain stem response: a within-subject comparison.

Authors:  Tiffany A Johnson; Carolyn J Brown
Journal:  Ear Hear       Date:  2005-12       Impact factor: 3.570

3.  Meta-analysis of variables that affect accuracy of threshold estimation via measurement of the auditory steady-state response (ASSR).

Authors:  Abreena I Tlumak; Elaine Rubinstein; John D Durrant
Journal:  Int J Audiol       Date:  2007-11       Impact factor: 2.117

4.  Comparison of pure tone audiometry and auditory steady-state responses in subjects with normal hearing and hearing loss.

Authors:  Ali Ozdek; Mahmut Karacay; Guleser Saylam; Emel Tatar; Nurdan Aygener; Mehmet Hakan Korkmaz
Journal:  Eur Arch Otorhinolaryngol       Date:  2010-01       Impact factor: 2.503

5.  Spectral characteristics of the responses of primary auditory-nerve fibers to amplitude-modulated signals.

Authors:  S M Khanna; M C Teich
Journal:  Hear Res       Date:  1989-05       Impact factor: 3.208

6.  Estimating audiometric thresholds using auditory steady-state responses.

Authors:  Terence W Picton; Andrew Dimitrijevic; Maria-Cecilia Perez-Abalo; Patricia Van Roon
Journal:  J Am Acad Audiol       Date:  2005-03       Impact factor: 1.664

7.  Predicting hearing thresholds and occupational hearing loss with multiple-frequency auditory steady-state responses.

Authors:  Ruey-Fen Hsu; Chi-Kung Ho; Sheng-Nan Lu; Shun-Sheng Chen
Journal:  J Otolaryngol Head Neck Surg       Date:  2010-10

8.  Scalp potentials of normal and hearing-impaired subjects in response to sinusoidally amplitude-modulated tones.

Authors:  S Kuwada; R Batra; V L Maher
Journal:  Hear Res       Date:  1986       Impact factor: 3.208

Review 9.  Human auditory steady-state responses.

Authors:  Terence W Picton; M Sasha John; Andrew Dimitrijevic; David Purcell
Journal:  Int J Audiol       Date:  2003-06       Impact factor: 2.117

10.  Threshold estimation in adult normal- and impaired-hearing subjects using auditory steady-state responses.

Authors:  Stavros Hatzopoulos; Silvano Prosser; Andrea Ciorba; Nadia Giarbini; Krzysztof Kochanek; Lech Sliwa; Henryk Skarzynski; Alessandro Martini
Journal:  Med Sci Monit       Date:  2010-01
View more
  5 in total

1.  Validity of correction factors applied to auditory steady-state responses (ASSRs) in normal hearing adults in chartr EP system.

Authors:  Zahra Ghasemahmad; Saeid Farahani
Journal:  Eur Arch Otorhinolaryngol       Date:  2019-05-21       Impact factor: 2.503

2.  Identifying congenital hearing impairment: preliminary results from a comparative study using objective and subjective audiometric protocols.

Authors:  A Ciorba; S Hatzopoulos; J Petruccelli; M Mazzoli; A Pastore; K Kochanek; P Skarzynski; A Wlodarczyk; H Skarzynski
Journal:  Acta Otorhinolaryngol Ital       Date:  2013-02       Impact factor: 2.124

3.  Towards neural correlates of auditory stimulus processing: a simultaneous auditory evoked potentials and functional magnetic resonance study using an odd-ball paradigm.

Authors:  Rafał Milner; Mateusz Rusiniak; Monika Lewandowska; Tomasz Wolak; Małgorzata Ganc; Ewa Piątkowska-Janko; Piotr Bogorodzki; Henryk Skarżyński
Journal:  Med Sci Monit       Date:  2014-01-13

4.  The Relationship Between Hearing Thresholds and Hyperandrogenism in Polycystic Ovary Syndrome.

Authors:  Mahfuz Turan; Rıfkı Ucler; Mehmet Fatih Garca; Zehra Kurdoglu; Hakan Cankaya; Abdurrahman Ayral; Nazım Bozan; Ahmet Faruk Kıroglu; Mertihan Kurdoglu
Journal:  Med Sci Monit       Date:  2016-11-15

5.  Comparison of Auditory Steady-State Responses With Conventional Audiometry in Older Adults.

Authors:  Hadeel Y Tarawneh; Hamid R Sohrabi; Wilhelmina H A M Mulders; Ralph N Martins; Dona M P Jayakody
Journal:  Front Neurol       Date:  2022-07-04       Impact factor: 4.086

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.