Hadeel Y Tarawneh1,2, Hamid R Sohrabi3,4,5, Wilhelmina H A M Mulders1, Ralph N Martins4,5, Dona M P Jayakody2,6.
Abstract
Behavioral measures, such as pure-tone audiometry (PTA), are commonly used to determine hearing thresholds, however, PTA does not always provide reliable hearing information in difficult to test individuals. Therefore, objective measures of hearing sensitivity that require little-to-no active participation from an individual are needed to facilitate the detection and treatment of hearing loss in difficult to test people. Investigation of the reliability of the auditory steady-state response (ASSR) for measuring hearing thresholds in older adults is limited. This study aimed to investigate if ASSR can be a reliable, objective measure of frequency specific hearing thresholds in older adults. Hearing thresholds were tested at 500 Hz, 1000 Hz, 2000 Hz, and 4000 Hz in 50 participants aged between 60 and 85 years old, using automated PTA and ASSR. Hearing thresholds obtained from PTA and ASSR were found to be significantly correlated (p < .001) in a cohort consisting of participants with normal hearing or mild hearing loss. ASSR thresholds were significantly higher as compared to PTA thresholds, but for the majority of cases the difference remained within the clinically acceptable range (15 dB). This study provides some evidence to suggest that ASSR can be a valuable tool for estimating objective frequency-specific hearing thresholds in older adults and indicate that ASSR could be useful in creating hearing treatment plans for older adults who are unable to complete behavioral PTA. Further research on older adults is required to improve the methodological features of ASSR to increase consistency and reliability, as well as minimize some of the limitations associated with this technique.Entities:
Keywords: auditory steady-state response (ASSR); hearing; objective audiometry; older adult; pure-tone audiometry (PTA)
Year: 2022 PMID: 35911911 PMCID: PMC9330634 DOI: 10.3389/fneur.2022.924096
Source DB: PubMed Journal: Front Neurol ISSN: 1664-2295 Impact factor: 4.086
Demographical characteristics of participants.
|
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|---|
| Combined | 50 | 72.1 ± 5.4 | 14.8 ± 2.5 | 26.9 ± 2.7 | 2.2 ± 2.1 | 2.2 ± 2.2 | 4.3 ± 2.9 |
| Females | 36 | 71.6 ± 5.1 | 14.9 ± 2.4 | 27.3 ± 2.9 | 2.1 ± 2 | 2 ± 1.9 | 4.5 ± 2.9 |
| Males | 14 | 72.9 ± 6 | 14.6 ± 2.7 | 26.2 ± 2.2 | 2.5 ± 2.3 | 2.6 ± 2.8 | 3.8 ± 2.8 |
Mean (± SD) pure tone audiometry (PTA) and auditory steady-state response (ASSR) hearing threshold values (in decibels dBHL) in normal hearing older adults.
|
|
|
|
|
| ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
| ||||||
| PTA | 13.9 ± 8.2 | 13.2 ± 7.2 | 13.6 ± 7.7 | 17.1 ± 8 | 16.6 ± 8.6 | 16.8 ± 8.2 | 15.5 ± 11.2 | 15.5 ± 11.7 | 15.5 ± 11.4 | 23.7 ± 16.3 | 24.5 ± 16.6 | 24.1 ± 16.3 |
| ASSR | 21.7 ± 12.4 | 20.6 ± 11.6 | 21.2 ± 11.9 | 22.8 ± 12.7 | 22.8 ± 11 | 22.8 ± 11.8 | 20.6 ± 12.1 | 21.8 ± 10.5 | 21.2 ± 11.3 | 30 ± 15.7 | 27.7 ± 16.7 | 28.8 ± 16.1 |
| Difference | 7.7 ± 11.6 | 7.3 ± 10.8 | 7.5 ± 11.2 | 5.7 ± 10.2 | 6.2 ± 10.2 | 6 ± 10.2 | 5.1 ± 8.3 | 6.3 ± 8.2 | 5.7 ± 8.2 | 6.2 ± 9.1 | 3.1 ± 9.2 | 4.7 ± 9.2 |
| P-value | ||||||||||||
R, right ear, L, left ear.
Figure 1Mean (± SD) pure tone audiometry (PTA; black lines A, B) and auditory steady-state response (ASSR; gray lines A, B) hearing thresholds (in decibels dBHL) for each carrier frequency in the right ear (A), left ear (B). (C) shows threshold differences between PTA and ASSR in left and right ears.
Figure 2Accuracy of ASSR estimations in percentage (%) presented for each frequency and for all frequencies combined.
Figure 3Distribution of ASSR and PTA differences for each frequency. Percentages presented for 0–5 dB, 10 dB, 15 dB and 20 dB+ (and over) threshold differences.
Figure 4Correlation between hearing thresholds (both ears combined) obtained from auditory steady-state response (ASSR) (x-axis) and pure-tone audiometry (PTA) (y-axis) in dB HL according to carrier frequency. Different panels show different frequencies. (A) 500Hz, (B) 1000 Hz, (C) 2000 Hz, and (D) 4000 Hz. Correlation coefficients (r) and p-value are presented on the top left corner of each panel for all tested frequencies. Black line indicates line of best fit, dotted line indicates 1:1 ratio line.