Precision and accuracy measurement of radiostereometric analysis applied to movement of the sacroiliac joint.

Thomas J Kibsgård1, Olav Røise, Britt Stuge, Stephan M Röhrl.   

Abstract

BACKGROUND: Different techniques have been used to quantify the movement of sacroiliac (SI) joints. These include radiostereometric analysis (RSA), but the accuracy and precision of this method have not been properly evaluated and it is unclear how many markers are required and where they should be placed to achieve proper accuracy and precision.
PURPOSE: The purpose of this study was to test accuracy and precision of RSA, applied to the SI joint, in a phantom model and in patients.
METHODS: We used a plastic phantom attached to a micrometer to obtain a true value of the movement of the SI joint and compared this value with the measured value obtained by RSA; the difference represented the accuracy. The precision of the system was measured by double examination in the phantom and in six patients, and was expressed by a limit of significance (LOS). We analyzed different marker distributions to find optimal marker placement and number of markers needed.
RESULTS: The accuracy was high and we identified no systematic errors. The precision of the phantom was high with a LOS less than 0.25° and 0.16 mm for all directions, and in patients, the precision was less than 0.71° for rotations and 0.47 mm translations. No markers were needed in the pubic symphysis to obtain good precision.
CONCLUSIONS: The accuracy and precision are high when RSA is used to measure movement in the SI joint and support the use of RSA in research of SI joint motion.

Entities:  

Mesh:

Year:  2012        PMID: 22695864      PMCID: PMC3462864          DOI: 10.1007/s11999-012-2413-5

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  22 in total

1.  Sacroiliac motion for extreme hip positions. A fresh cadaver study.

Authors:  G L Smidt; S H Wei; K McQuade; E Barakatt; T Sun; W Stanford
Journal:  Spine (Phila Pa 1976)       Date:  1997-09-15       Impact factor: 3.468

2.  Accuracy of radiographic and radiostereometric wear measurement of different hip prostheses: an experimental study.

Authors:  Thord von Schewelov; Lennart Sanzén; Niclas Börlin; Patrik Markusson; Ingemar Onsten
Journal:  Acta Orthop Scand       Date:  2004-12

3.  Palamed G compared with Palacos R with gentamicin in Charnley total hip replacement. A randomised, radiostereometric study of 60 HIPS.

Authors:  G Hallan; A Aamodt; O Furnes; A Skredderstuen; K Haugan; L I Havelin
Journal:  J Bone Joint Surg Br       Date:  2006-09

4.  Clinical validation of model-based RSA for a total knee prosthesis.

Authors:  Bart L Kaptein; Edward R Valstar; Berend C Stoel; Hans C Reiber; Rob G Nelissen
Journal:  Clin Orthop Relat Res       Date:  2007-11       Impact factor: 4.176

5.  Radiostereometry of hip prostheses. Review of methodology and clinical results.

Authors:  J Kärrholm; P Herberts; P Hultmark; H Malchau; B Nivbrant; J Thanner
Journal:  Clin Orthop Relat Res       Date:  1997-11       Impact factor: 4.176

6.  RSA wear measurements with or without markers in total hip arthroplasty.

Authors:  Niclas Börlin; Stephan M Röhrl; Charles R Bragdon
Journal:  J Biomech       Date:  2005-06-28       Impact factor: 2.712

7.  The mobility of the sacroiliac joints in healthy volunteers between 20 and 50 years of age.

Authors:  H A C Jacob; R O Kissling
Journal:  Clin Biomech (Bristol, Avon)       Date:  1995-10       Impact factor: 2.063

8.  Accuracy and precision of radiostereometric analysis in the measurement of three-dimensional micromotion in a fracture model of the distal radius.

Authors:  Rami Madanat; Tatu J Mäkinen; Niko Moritz; Kimmo T Mattila; Hannu T Aro
Journal:  J Orthop Res       Date:  2005-03       Impact factor: 3.494

9.  Guidelines for standardization of radiostereometry (RSA) of implants.

Authors:  Edward R Valstar; Richie Gill; Leif Ryd; Gunnar Flivik; Niclas Börlin; Johan Kärrholm
Journal:  Acta Orthop       Date:  2005-08       Impact factor: 3.717

Review 10.  Mobility of the pelvic joints in pregnancy-related lumbopelvic pain: a systematic review.

Authors:  Jan M A Mens; Annelies Pool-Goudzwaard; Henk J Stam
Journal:  Obstet Gynecol Surv       Date:  2009-03       Impact factor: 2.347

View more
  5 in total

1.  Physiological in vitro sacroiliac joint motion: a study on three-dimensional posterior pelvic ring kinematics.

Authors:  Niels Hammer; Mario Scholze; Thomas Kibsgård; Stefan Klima; Stefan Schleifenbaum; Thomas Seidel; Michael Werner; Ronny Grunert
Journal:  J Anat       Date:  2018-12-09       Impact factor: 2.610

2.  Experimental evaluation of precision and accuracy of RSA in the lumbar spine.

Authors:  Marie Christina Keller; Christof Hurschler; Michael Schwarze
Journal:  Eur Spine J       Date:  2020-12-04       Impact factor: 3.134

Review 3.  The sacroiliac joint: an overview of its anatomy, function and potential clinical implications.

Authors:  A Vleeming; M D Schuenke; A T Masi; J E Carreiro; L Danneels; F H Willard
Journal:  J Anat       Date:  2012-09-19       Impact factor: 2.610

4.  Authors' response to the letter to the editors by Professor M. T. Cibulka: a critical interpretation of sacroiliac joint movement studies.

Authors:  A Vleeming; M D Schuenke; B Sturesson; F H Willard
Journal:  J Anat       Date:  2013-03       Impact factor: 2.610

5.  Finite element analysis of load transition on sacroiliac joint during bipedal walking.

Authors:  Ryota Toyohara; Daisuke Kurosawa; Niels Hammer; Michael Werner; Keita Honda; Yusuke Sekiguchi; Shin-Ichi Izumi; Eiichi Murakami; Hiroshi Ozawa; Toshiro Ohashi
Journal:  Sci Rep       Date:  2020-08-13       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.