Accuracy of radiographic and radiostereometric wear measurement of different hip prostheses: an experimental study.

Thord von Schewelov1, Lennart Sanzén, Niclas Börlin, Patrik Markusson, Ingemar Onsten.   

Abstract

BACKGROUND: In vivo measurement of wear in the ball and socket articulation of total hip arthroplasties is of interest in the evaluation of both existing and new implants. Controversy reigns regarding the accuracy of different radiological measurement techniques and in particular how accuracy has been assessed.
MATERIAL AND METHODS: We assessed the accuracy of 2 radiostereometric (RSA) techniques for wear measurement and 3 standard radiographic techniques, namely Imagika (image analyzing software), Imagika corrected for head center displacement, and the Charnley Duo method. 5 custom-made adjustable phantoms with different prosthetic components were used.
RESULTS: In 20 measurements of all 5 phantoms at 3 levels of simulated wear (0.2 mm, 1.0 mm and 1.5 mm), the mean measurement error of the digital RSA examinations was 0.010 mm (accuracy 0.42). The corresponding error values for the three radiographic techniques were 0.19 (accuracy 1.3) for Charnley Duo, 0.13 (accuracy 1.3) for Imagika corrected, and 1.021 (accuracy 2.99) for Imagika. Measurement error decreased from 0.011 mm with ordinary RSA to 0.004 with RSA digital measurement. Head size, direction of wear in relation to the cup or type of prosthetic component did not influence the measurement error. The results of Charnley Duo and Imagika corrected were similar but the latter had an inexplicable systematic error in measuring one of the phantoms. Imagika had the worst results due to its inability to compensate for the out-of-head center effect. Alumina heads were difficult to analyze with all methods.
INTERPRETATION: By using the ISO standard for assessing accuracy, RSA can be expected to measure wear with an accuracy of about 0.4 mm irrespective of prosthetic component studied or direction of wear, whereas the best technique, in our study, based on standard radiographs can be accurate to about 1.3 mm.

Entities:  

Mesh:

Year:  2004        PMID: 15762258     DOI: 10.1080/00016470410004058

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  9 in total

1.  Precision and accuracy measurement of radiostereometric analysis applied to movement of the sacroiliac joint.

Authors:  Thomas J Kibsgård; Olav Røise; Britt Stuge; Stephan M Röhrl
Journal:  Clin Orthop Relat Res       Date:  2012-06-14       Impact factor: 4.176

2.  Second-generation highly cross-linked X3™ polyethylene wear: a preliminary radiostereometric analysis study.

Authors:  David G Campbell; John R Field; Stuart A Callary
Journal:  Clin Orthop Relat Res       Date:  2010-02-12       Impact factor: 4.176

3.  Analysis of polyethylene wear in plain radiographs.

Authors:  Maiken Stilling; Kjeld Søballe; Niels Trolle Andersen; Kristian Larsen; Ole Rahbek
Journal:  Acta Orthop       Date:  2009-12       Impact factor: 3.717

4.  Clinical comparison of polyethylene wear with zirconia or cobalt-chromium femoral heads.

Authors:  Maiken Stilling; Kjeld Anton Nielsen; Kjeld Søballe; Ole Rahbek
Journal:  Clin Orthop Relat Res       Date:  2009-03-27       Impact factor: 4.176

5.  The final follow-up plain radiograph is sufficient for clinical evaluation of polyethylene wear in total hip arthroplasty. A study of validity and reliability.

Authors:  Maiken Stilling; Kristian Larsen; Niels T Andersen; Kjeld Søballe; Søren Kold; Ole Rahbek
Journal:  Acta Orthop       Date:  2010-10       Impact factor: 3.717

6.  Validation of a 3D CT method for measurement of linear wear of acetabular cups.

Authors:  Anneli Jedenmalm; Fritjof Nilsson; Marilyn E Noz; Douglas D Green; Ulf W Gedde; Ian C Clarke; Andreas Stark; Gerald Q Maguire; Michael P Zeleznik; Henrik Olivecrona
Journal:  Acta Orthop       Date:  2011-02-01       Impact factor: 3.717

7.  Notes on the use and interpretation of radiostereometric analysis.

Authors:  Brian Derbyshire; Robin J Prescott; Martyn L Porter
Journal:  Acta Orthop       Date:  2009-02       Impact factor: 3.717

8.  Superior accuracy of model-based radiostereometric analysis for measurement of polyethylene wear: A phantom study.

Authors:  M Stilling; S Kold; S de Raedt; N T Andersen; O Rahbek; K Søballe
Journal:  Bone Joint Res       Date:  2012-08-01       Impact factor: 5.853

9.  A new automated way to measure polyethylene wear in THA using a high resolution CT scanner: method and analysis.

Authors:  Gerald Q Maguire; Marilyn E Noz; Henrik Olivecrona; Michael P Zeleznik; Lars Weidenhielm
Journal:  ScientificWorldJournal       Date:  2014-01-22
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.