Small bowel capsule endoscopy in patients with cardiac pacemakers and implantable cardioverter defibrillators: Outcome analysis using telemetry review.

Justin R Cuschieri1, Mohammed N Osman, Richard Ck Wong, Amitabh Chak, Gerard A Isenberg.   

Abstract

AIM: To determine if there were any interactions between cardiac devices and small bowel capsules secondary to electromagnetic interference (EMI) in patients who have undergone small bowel capsule endoscopy (SBCE).
METHODS: Authors conducted a chart review of 20 patients with a cardiac pacemaker (CP) or implantable cardioverter defibrillator (ICD) who underwent continuous electrocardiographic monitoring during their SBCE from 2003-2008. authors searched for unexplained electrocardiogram (ECG) findings, changes in CP and ICD set parameters, any abnormality in transmitted capsule data, and adverse clinical events.
RESULTS: There were no adverse events or hemodynamically significant arrhythmias reported. CP and ICD set parameters were preserved. The majority of ECG abnormalities were also found in pre- or post- SBCE ECG tracings and the CP behavior during arrhythmias appeared appropriate. Two patients seemed to have episodes of undersensing by the CP. However, similar findings were documented in ECGs taken outside the time frame of the SBCE. One patient was observed to have a low signal encountered from the capsule resulting in lack of localization, but no images were lost.
CONCLUSION: Capsule-induced EMI remains a possibility but is unlikely to be clinically important. CP-induced interference of SBCE is also possible, but is infrequent and does not result in loss of images transmitted by the capsule.

Entities:  

Keywords:  Cardiac pacemakers; Electromagnetic interference; Implantable cardioverter defibrillators; Small bowel capsule endoscopy; Telemetry review

Year:  2012        PMID: 22442746      PMCID: PMC3309898          DOI: 10.4253/wjge.v4.i3.87

Source DB:  PubMed          Journal:  World J Gastrointest Endosc


  20 in total

1.  Establishment of a Medical Implant Communications Service in the 402-405 MHz band. Federal Communications Commission. Final rule.

Authors: 
Journal:  Fed Regist       Date:  1999-12-15

2.  Reliability of electromagnetic filters of cardiac pacemakers tested by cellular telephone ringing.

Authors:  Alexandre Trigano; Olivier Blandeau; Christian Dale; Man-Fai Wong; Joe Wiart
Journal:  Heart Rhythm       Date:  2005-08       Impact factor: 6.343

3.  The effects of percutaneous catheter ablation on preexisting permanent pacemakers.

Authors:  G Vanerio; J Maloney; R Rashidi; R McCowan; L Castle; V Morant; B Wilkoff; T Simmons
Journal:  Pacing Clin Electrophysiol       Date:  1990-12       Impact factor: 1.976

4.  Effects of capsule endoscopy on cardiac pacemakers.

Authors:  G Payeras; J Piqueras; V J Moreno; A Cabrera; D Menéndez; R Jiménez
Journal:  Endoscopy       Date:  2005-12       Impact factor: 10.093

5.  Electromagnetic interference with implantable cardiac pacemakers by video capsule.

Authors:  Sergio Dubner; Yael Dubner; Sebastian Gallino; Liliana Spallone; David Zagalsky; Horacio Rubio; Joseph Zimmerman; Eran Goldin
Journal:  Gastrointest Endosc       Date:  2005-02       Impact factor: 9.427

6.  [Capsule endoscopy in patients with cardiac pacemakers: current situation in Germany].

Authors:  D Bandorski; K-L Diehl; D Jaspersen
Journal:  Z Gastroenterol       Date:  2005-08       Impact factor: 2.000

7.  Safety of wireless capsule endoscopy in patients with implantable cardiac defibrillators.

Authors:  Jonathan A Leighton; Komandoor Srivathsan; Elizabeth J Carey; Virender K Sharma; Russell I Heigh; Janice K Post; Paula J Erickson; Sarah R Robinson; Jane L Bazzell; David E Fleischer
Journal:  Am J Gastroenterol       Date:  2005-08       Impact factor: 10.864

8.  The effects of electronic article surveillance systems on permanent cardiac pacemakers: an in vitro study.

Authors:  E H Lucas; D Johnson; B P McElroy
Journal:  Pacing Clin Electrophysiol       Date:  1994-11       Impact factor: 1.976

9.  Interference with cardiac pacemakers by cellular telephones.

Authors:  D L Hayes; P J Wang; D W Reynolds; M Estes; J L Griffith; R A Steffens; G L Carlo; G K Findlay; C M Johnson
Journal:  N Engl J Med       Date:  1997-05-22       Impact factor: 91.245

10.  [Could C- and D-network mobile phones endanger patients with pacemakers?].

Authors:  F Hofgärtner; T Müller; H Sigel
Journal:  Dtsch Med Wochenschr       Date:  1996-05-17       Impact factor: 0.628

View more
  13 in total

Review 1.  Small-bowel capsule endoscopy: a ten-point contemporary review.

Authors:  Anastasios Koulaouzidis; Emanuele Rondonotti; Alexandros Karargyris
Journal:  World J Gastroenterol       Date:  2013-06-28       Impact factor: 5.742

2.  Lack of interference between small bowel capsule endoscopy and implantable cardiac defibrillators: an 'in vivo' electrophysiological study.

Authors:  Dario Moneghini; Alessandro Lipari; Guido Missale; Luigi Minelli; Gianpaolo Cengia; Luca Bontempi; Antonio Curnis; Renzo Cestari
Journal:  United European Gastroenterol J       Date:  2015-09-24       Impact factor: 4.623

Review 3.  Small bowel capsule endoscopy: Where are we after almost 15 years of use?

Authors:  Cedric Van de Bruaene; Danny De Looze; Pieter Hindryckx
Journal:  World J Gastrointest Endosc       Date:  2015-01-16

4.  Interference between pacemakers/implantable cardioverter defibrillators and video capsule endoscopy.

Authors:  Dirk Bandorski; Johannes Gehron; Reinhard Höltgen
Journal:  World J Gastrointest Endosc       Date:  2013-04-16

Review 5.  Capsule endoscopy in patients with cardiac pacemakers, implantable cardioverter defibrillators and left heart assist devices.

Authors:  Dirk Bandorski; Reinhard Höltgen; Dominik Stunder; Martin Keuchel
Journal:  Ann Gastroenterol       Date:  2014

6.  Is It Safe to Perform Therapeutic Endoscopic Procedures in Patients with Implanted Electronic Cardiac Devices?

Authors:  Eun Young Kim
Journal:  Clin Endosc       Date:  2016-03-25

Review 7.  Contraindications for video capsule endoscopy.

Authors:  Dirk Bandorski; Niehls Kurniawan; Peter Baltes; Reinhard Hoeltgen; Matthias Hecker; Dominik Stunder; Martin Keuchel
Journal:  World J Gastroenterol       Date:  2016-12-07       Impact factor: 5.742

8.  Pooled Analysis of the Efficacy and Safety of Video Capsule Endoscopy in Patients with Implantable Cardiac Devices.

Authors:  Rabih Tabet; Najib Nassani; Boutros Karam; Youssef Shammaa; Philippe Akhrass; Liliane Deeb
Journal:  Can J Gastroenterol Hepatol       Date:  2019-05-19

9.  Capsule Endoscopy in Patients with Implantable Electromedical Devices is Safe.

Authors:  Lucinda A Harris; Stephanie L Hansel; Elizabeth Rajan; Komandoor Srivathsan; Robert Rea; Michael D Crowell; David E Fleischer; Shabana F Pasha; Suryakanth R Gurudu; Russell I Heigh; Arthur D Shiff; Janice K Post; Jonathan A Leighton
Journal:  Gastroenterol Res Pract       Date:  2013-04-03       Impact factor: 2.260

10.  Clinical use of capsule endoscopy in a patient with cardiac pacemaker.

Authors:  Osman Ersoy; Ebru Akin; Aylin Demirezer
Journal:  Saudi J Gastroenterol       Date:  2013 Sep-Oct       Impact factor: 2.485

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.