Literature DB >> 15729234

Electromagnetic interference with implantable cardiac pacemakers by video capsule.

Sergio Dubner, Yael Dubner, Sebastian Gallino, Liliana Spallone, David Zagalsky, Horacio Rubio, Joseph Zimmerman, Eran Goldin.   

Abstract

BACKGROUND: Electromagnetic fields, such as those generated by cellular phones and metal detectors, may interfere with normal pacemaker function. However, it remains unclear whether the wireless capsule endoscope interacts with implanted pacemakers. This prospective study evaluated potential interactions between the M2A video capsule (Given Imaging, Yoknam, Israel) and implanted pacemakers.
METHODS: A total of 100 consecutive patients (70 men, 30 women) with an implanted pacemaker (95 on bipolar mode) were studied. The testing was performed with a functional testing device (Test Cap) for the Given Diagnostic System that reproduces the effect of the video capsule by transmitting at exactly the same frequency. During continuous electrocardiographic monitoring and recording, 100 tests were carried out without changing the pacemaker settings. Those with a positive result were retested 1 week later.
RESULTS: The 100 pacemakers evaluated in the study population included the following: 70 dual chamber (11 DDD, 56 DDDR, 3 VDD) and 30 ventricular inhibited (12 VVI, 18 VVIR). In 4 of the 100 patients, pacemaker interference (noise-mode function forcing a synchronous mode) was registered during the Test Cap operation. Three patients had a dual-chamber pacemaker, and one had a single-chamber pacemaker. The interference was reproducible in all cases 1 week later. None of the implanted pacemakers tested was affected by oversensing.
CONCLUSIONS: Electromagnetic interferences with pacemakers from the M2A video capsule can occur, but this is without clinical significance. No potentially dangerous pacemaker inhibition was observed.

Entities:  

Mesh:

Year:  2005        PMID: 15729234     DOI: 10.1016/s0016-5107(04)02834-2

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  15 in total

Review 1.  Endoscopic techniques to detect small-bowel neuroendocrine tumors: A literature review.

Authors:  Roberta Elisa Rossi; Dario Conte; Luca Elli; Federica Branchi; Sara Massironi
Journal:  United European Gastroenterol J       Date:  2016-07-08       Impact factor: 4.623

2.  Safety of capsule endoscopy using human body communication in patients with cardiac devices.

Authors:  Joo Won Chung; Hye Jin Hwang; Moon Jae Chung; Jeong Youp Park; Hui-Nam Pak; Si Young Song
Journal:  Dig Dis Sci       Date:  2012-02-07       Impact factor: 3.199

3.  Small bowel capsule endoscopy in patients with cardiac pacemakers and implantable cardioverter defibrillators: Outcome analysis using telemetry review.

Authors:  Justin R Cuschieri; Mohammed N Osman; Richard Ck Wong; Amitabh Chak; Gerard A Isenberg
Journal:  World J Gastrointest Endosc       Date:  2012-03-16

Review 4.  Endoscopy and implantable electronic devices.

Authors:  G D Corbett; P C Buttery; P J Pugh; E A B Cameron
Journal:  Frontline Gastroenterol       Date:  2012-02-05

5.  Capsule endoscopy in the evaluation of obscure gastrointestinal bleeding: a comprehensive review.

Authors:  Feng Li; Jonathan A Leighton; Virender K Sharma
Journal:  Gastroenterol Hepatol (N Y)       Date:  2007-10

Review 6.  Small bowel capsule endoscopy: Where are we after almost 15 years of use?

Authors:  Cedric Van de Bruaene; Danny De Looze; Pieter Hindryckx
Journal:  World J Gastrointest Endosc       Date:  2015-01-16

7.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

8.  Interference between pacemakers/implantable cardioverter defibrillators and video capsule endoscopy.

Authors:  Dirk Bandorski; Johannes Gehron; Reinhard Höltgen
Journal:  World J Gastrointest Endosc       Date:  2013-04-16

9.  Capsule endoscopy in patients with cardiac pacemakers, implantable cardioverter defibrillators, and left heart devices: a review of the current literature.

Authors:  Dirk Bandorski; Martin Keuchel; Martin Brück; Reinhard Hoeltgen; Marcus Wieczorek; Ralf Jakobs
Journal:  Diagn Ther Endosc       Date:  2011-04-27

10.  Capsule Endoscopy in Patients with Cardiac Pacemakers and Implantable Cardioverter Defibrillators: (Re)evaluation of the Current State in Germany, Austria, and Switzerland 2010.

Authors:  Dirk Bandorski; Ralf Jakobs; Martin Brück; Reinhard Hoeltgen; Marcus Wieczorek; Martin Keuchel
Journal:  Gastroenterol Res Pract       Date:  2011-12-29       Impact factor: 2.260