Literature DB >> 9154765

Interference with cardiac pacemakers by cellular telephones.

D L Hayes1, P J Wang, D W Reynolds, M Estes, J L Griffith, R A Steffens, G L Carlo, G K Findlay, C M Johnson.   

Abstract

BACKGROUND: A growing body of evidence suggests that electromagnetic interference may occur between cardiac pacemakers and wireless hand-held (cellular) telephones, posing a potential public health problem. Electromagnetic interference may occur when the pacemaker is exposed to an electromagnetic field generated by the cellular telephone.
METHODS: In this multicenter, prospective, crossover study, we tested 980 patients with cardiac pacemakers with five types of telephones (one analogue and four digital) to assess the potential for interference. Telephones were tested in a test mode and were programmed to transmit at the maximal power, simulating the worst-case scenario; in addition, one telephone was tested during actual transmission to simulate actual use. Patients were electrocardiographically monitored while the telephones were tested at the ipsilateral ear and in a series of maneuvers directly over the pacemaker. Interference was classified according to the type and clinical significance of the effect.
RESULTS: The incidence of any type of interference was 20 percent in the 5533 tests, and the incidence of symptoms was 7.2 percent. The incidence of clinically significant interference was 6.6 percent. There was no clinically significant interference when the telephone was placed in the normal position over the ear. Interference that was definitely clinically significant occurred in only 1.7 percent of tests, and only when the telephone was held over the pacemaker. Interference was more frequent with dual-chamber pacemakers (25.3 percent) than with single-chamber pacemakers (6.8 percent, P<0.001) and more frequent with pacemakers without feed-through filters (28.9 to 55.8 percent) than with those with such filters (0.4 to 0.8 percent, P=0.01).
CONCLUSIONS: Cellular telephones can interfere with the function of implanted cardiac pacemakers. However, when telephones are placed over the ear, the normal position, this interference does not pose a health risk.

Entities:  

Mesh:

Year:  1997        PMID: 9154765     DOI: 10.1056/NEJM199705223362101

Source DB:  PubMed          Journal:  N Engl J Med        ISSN: 0028-4793            Impact factor:   91.245


  28 in total

1.  Possible electromagnetic interference with electronic medical equipment by radio waves coming from outside the hospital.

Authors:  E Hanada; K Kodama; K Takano; Y Watanabe; Y Nose
Journal:  J Med Syst       Date:  2001-08       Impact factor: 4.460

Review 2.  Theory and applications of biotelemetry.

Authors:  Nihal Fatma Güler; Elif Derya Ubeyli
Journal:  J Med Syst       Date:  2002-04       Impact factor: 4.460

3.  Infrared transmission of electronic information via LAN in the operating room.

Authors:  S Hagihira; M Takashina; T Mori; N Taenaka; T Mashimo; I Yoshiya
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

4.  Experimental study on malfunction of pacemakers due to exposure to different external magnetic fields.

Authors:  Maria Tiikkaja; Tommi Alanko; Harri Lindholm; Maila Hietanen; Juha Hartikainen; Lauri Toivonen
Journal:  J Interv Card Electrophysiol       Date:  2012-01-11       Impact factor: 1.900

5.  Evaluation of Electromagnetic Fields in a Hospital for Safe Use of Electronic Medical Equipment.

Authors:  Kai Ishida; Tomomi Fujioka; Tetsuo Endo; Ren Hosokawa; Tetsushi Fujisaki; Ryoji Yoshino; Minoru Hirose
Journal:  J Med Syst       Date:  2015-12-07       Impact factor: 4.460

6.  Radiofrequency identification: exploiting an old technology for measuring nurse time and motion.

Authors:  Terry L Jones
Journal:  Comput Inform Nurs       Date:  2012-09       Impact factor: 1.985

7.  Interference by cellular telephone with an implantable loop recorder.

Authors:  Alexandre Trigano; Olivier Blandeau; Samuel Levy
Journal:  J Interv Card Electrophysiol       Date:  2005-04       Impact factor: 1.900

8.  Use of mobile phones in hospitals.

Authors:  Stuart W G Derbyshire; Adam Burgess
Journal:  BMJ       Date:  2006-10-14

9.  Safety of capsule endoscopy using human body communication in patients with cardiac devices.

Authors:  Joo Won Chung; Hye Jin Hwang; Moon Jae Chung; Jeong Youp Park; Hui-Nam Pak; Si Young Song
Journal:  Dig Dis Sci       Date:  2012-02-07       Impact factor: 3.199

10.  Comparison of Injected and Radiated EMC Testing of Active Implanted Cardiac Medical Devices at the Boundary Frequency of 450 MHz.

Authors:  Howard Bassen; Gonzalo Mendoza
Journal:  IEEE Electromagn Compat Mag       Date:  2016 Fourth Quarter
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.