Reliability of electromagnetic filters of cardiac pacemakers tested by cellular telephone ringing.

Alexandre Trigano1, Olivier Blandeau, Christian Dale, Man-Fai Wong, Joe Wiart.   

Abstract

BACKGROUND: State-of-the art cardiac pacemakers are protected against radiofrequency signals. Although there have been earlier clinical and in vitro reports of cellular phone interference with implantable devices, only a few studies have been performed in recent years. The ringing phase of digital GSM or PCS cellular phones includes a brief period of peak radiated power.
OBJECTIVES: This study tested the protection offered by electromagnetic filters of cardiac pacemakers against cellular phone ringing.
METHODS: We performed 330 consecutive tests in 158 patients at the time of routine examination in our pacemaker follow-up clinic. The programmed parameters remained unchanged before testing. During electrocardiographic monitoring, 2 single-band digital cellular phones consecutively placed over the pacemaker pocket each received a call. The phone systems tested were 1) GSM at a maximal power output of 2 W, operating on a 900 MHz carrier frequency, and 2) PCS at a maximal output of 1 W, operating on a 1800 MHz carrier frequency.
RESULTS: Interference was noted in only 5 tests, due to interaction by the GSM system with 4 unprotected pacemaker models. The GSM test was negative in 12 other tests of identical pulse generator models. The overall incidence of interference was 1.5% of tests.
CONCLUSIONS: Interference by cellular phone ringing occurred only with unprotected pacemaker models. Standard programming of these unprotected models was associated with a low incidence of interference.

Entities:  

Mesh:

Year:  2005        PMID: 16051120     DOI: 10.1016/j.hrthm.2005.03.011

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  4 in total

1.  Experimental study on malfunction of pacemakers due to exposure to different external magnetic fields.

Authors:  Maria Tiikkaja; Tommi Alanko; Harri Lindholm; Maila Hietanen; Juha Hartikainen; Lauri Toivonen
Journal:  J Interv Card Electrophysiol       Date:  2012-01-11       Impact factor: 1.900

2.  Small bowel capsule endoscopy in patients with cardiac pacemakers and implantable cardioverter defibrillators: Outcome analysis using telemetry review.

Authors:  Justin R Cuschieri; Mohammed N Osman; Richard Ck Wong; Amitabh Chak; Gerard A Isenberg
Journal:  World J Gastrointest Endosc       Date:  2012-03-16

Review 3.  [Electromagnetic interference : Pacemakers, cardiac resynchronization therapy devices, implantable cardioverter-defibrillator].

Authors:  Bernd Nowak; Oliver Przibille; Andreas Napp
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2022-07-04

4.  Interference between cellular telephones and implantable rhythm devices: a review on recent papers.

Authors:  Johnson Francis; Michael Niehaus
Journal:  Indian Pacing Electrophysiol J       Date:  2006-10-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.