The effects of percutaneous catheter ablation on preexisting permanent pacemakers.

G Vanerio1, J Maloney, R Rashidi, R McCowan, L Castle, V Morant, B Wilkoff, T Simmons.   

Abstract

STUDY
OBJECTIVE: Determine the effect of percutaneous catheter ablation (CA) on permanent pacemakers. MEASUREMENTS AND
RESULTS: Twenty-three patients who underwent CA at The Cleveland Clinic Foundation from September 1983 to January 1990, and had a previously implanted pacemaker were studied. Electrocardiographic data during the CA procedure and clinic data including pacemaker evaluations were analyzed. Fifty-two percent (12/23) of the pacemakers malfunctioned: five developed transient ventricular loss of capture; two undersensing; one oversensing; three could not be interrogated or programmed, and one did not respond to the magnet test. Four patients developed syncopal episodes and two severe dizziness after the procedure. All had their pacemakers replaced. In total, seven were explanted. Destructive analysis by the individual manufacturer identified pacemaker circuitry failure in five. Unipolar pacemakers and anodal ablation procedures had more frequent and severe malfunctions, but the difference was not statistically significant.
CONCLUSIONS: Pacemaker malfunction is frequent during CA. It may be prevented by programming the pacemaker, when possible, to the nonfunctioning mode (000 mode). Temporarily disconnecting the pacemaker during ablation requires further evaluation as an alternative approach. Close follow-up can detect pacemaker malfunction and prevent complications.

Entities:  

Mesh:

Year:  1990        PMID: 1704517     DOI: 10.1111/j.1540-8159.1990.tb06866.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Small bowel capsule endoscopy in patients with cardiac pacemakers and implantable cardioverter defibrillators: Outcome analysis using telemetry review.

Authors:  Justin R Cuschieri; Mohammed N Osman; Richard Ck Wong; Amitabh Chak; Gerard A Isenberg
Journal:  World J Gastrointest Endosc       Date:  2012-03-16

2.  Capsule Endoscopy in Patients with Implantable Electromedical Devices is Safe.

Authors:  Lucinda A Harris; Stephanie L Hansel; Elizabeth Rajan; Komandoor Srivathsan; Robert Rea; Michael D Crowell; David E Fleischer; Shabana F Pasha; Suryakanth R Gurudu; Russell I Heigh; Arthur D Shiff; Janice K Post; Jonathan A Leighton
Journal:  Gastroenterol Res Pract       Date:  2013-04-03       Impact factor: 2.260

3.  Safety of wireless capsule endoscopy in patients with implantable cardiac devices.

Authors:  Christopher Kasia; Anoop Appannagari; Anjali Joshi; Mukund Venu
Journal:  JGH Open       Date:  2019-09-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.