Rabih Tabet1, Najib Nassani2, Boutros Karam1, Youssef Shammaa1, Philippe Akhrass3, Liliane Deeb4.
Abstract
Background: To date, video capsule endoscopy (VCE) is still contraindicated by the FDA and the main manufacturers of Cardiac Implantable Electronic Devices (CIED) in patients with CIED, given a theoretical electromagnetic interference and possible device malfunction.Entities:
Mesh:
Year: 2019 PMID: 31236386 PMCID: PMC6545804 DOI: 10.1155/2019/3953807
Source DB: PubMed Journal: Can J Gastroenterol Hepatol ISSN: 2291-2789
Figure 1PRISMA diagram showing the selection process of the articles included in the final analysis.
Figure 2Number of published articles by year showing increasing interest in this topic.
Studies included in the analysis and the baseline characteristics of the patients.
| Study | Year | n | Mean age | Male | Female | Completion rate | Diagnostic rate |
|---|---|---|---|---|---|---|---|
| Leighton [ | 2004 | 5 | 73 | 4/5 (80%) | 1/5 (20%) | 5/5 (100%) | n/a |
| Payeras [ | 2005 | 20 | 78 | 14/20 (70%) | 6/20 (30%) | 20/20 (100%) | 14/20 (70%) |
| Dubner [ | 2005 | 6 | 61 | 4/6 (66%) | 2/6 (34%) | n/a | n/a |
| Leighton [ | 2005 | 5 | 72 | 1/5 (20%) | 4/5 (80%) | n/a | n/a |
| Dirks [ | 2008 | 5 | 45 | 0/5 (0%) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) |
| Elias [ | 2009 | 4 | 78 | 2/4 (50%) | 2/4 (50%) | n/a | n/a |
| Bandorski [ | 2011 | 59 | 73 | n/a | n/a | n/a | n/a |
| Chung [ | 2012 | 6 | 60.5 | 3/6 (50%) | 3/6 (50%) | 6/6 (100%) | n/a |
| Bandorski [ | 2012 | 380 | n/a | n/a | n/a | n/a | n/a |
| Cuschieri [ | 2012 | 20 | 71 | 13/20 (65%) | 7/20 (35%) | n/a | n/a |
| Harris [ | 2013 | 118 | 72 | 74/118 (63%) | 44/118 (37%) | 79/118 (67%) | n/a |
| Stanish [ | 2014 | 21 | 71.1 | 8/21 (38%) | 13/21 (62%) | 20/21 (95%) | 13/21 (62%) |
| Moneghini [ | 2016 | 14 | 66.4 | 11/14 (78%) | 3/14 (22%) | 13/14 (92.8%) | 9/14 (64.2%) |
| Amornsawadwattana [ | 2016 | 30 | 60.1 | 24/30 (80%) | 6/30 (20%) | 30/30 (100%) | 12/30 (40%) |
| Truss [ | 2016 | 8 | 63 | 8/8 (100%) | 0/8 (0%) | 8/8 (100%) | 8/8 (100%) |
| Hanson [ | 2016 | 34 | 67.1 | n/a | n/a | 34/34 (100%) | 31/34 (91%) |
|
|
|
|
|
|
|
|
Pooled results of the 16 studies included in the analysis [5–20].
| Type of Device | Total Number of cardiac devices (n) | Cardiac devices brand | VCE brand | Cardiac Outcome | Quality of transmitted images |
|---|---|---|---|---|---|
| PPM | 487 | Medtronic (77) | PillCam (431) | No adverse cardiac events | 4 cases of VCE interference with telemetry leads leading to few images loss |
|
| |||||
| ICD | 162 | Medtronic (15) | PillCam (147) | No adverse cardiac events | 2 cases of VCE interference with telemetry leads leading to few images loss |
|
| |||||
| LVAD | 86 | Thoratec HeartMate II (35) | PillCam (52) | No adverse cardiac events | 5 cases of possible interference leading to images loss/artifact when VCE was near the device |
Permanent Pacemaker (PPM); Implantable Cardiac Defibrillator (ICD); Left Ventricular Assist Device (LVAD); Video Capsule Endoscopy (VCE).