A phantom-based calibration method for digital x-ray tomosynthesis.

Hui Miao1, Xizeng Wu, Huijuan Zhao, Hong Liu.   

Abstract

OBJECTIVE: The purpose of this study was to develop a phantom-based experimental calibration method to minimize the reconstruction artifacts for the geometric misalignments of the digital tomosynthesis prototype.
METHODS: A calibration phantom with ten fiducial markers was designed. Using this calibration phantom, the projection matrices of an experimental digital tomosynthesis prototype were acquired from each projection view under a series of misalignment conditions. The American College of Radiology mammography phantom was imaged and reconstructed with and without using the correction of the corresponding calibration projection matrices. The effectiveness of the calibration technique was then quantitatively analyzed through comparison of the calibrated and uncalibrated images.
RESULTS: As the isocenter horizontal-shift increases, the reconstruction artifacts become clearly distinguishable. Using the calibration technique, the reconstruction artifacts resulting from the isocenter horizontal-shift were effectively minimized for the prototype.
CONCLUSIONS: For the specific experimental conditions utilized in this study, the phantom-based calibration method effectively reduced reconstruction artifacts for the prototype investigated in this study. The calibration method holds potential to benefit other tomosynthesis applications.

Entities:  

Mesh:

Year:  2012        PMID: 22398585      PMCID: PMC3313675          DOI: 10.3233/XST-2012-0316

Source DB:  PubMed          Journal:  J Xray Sci Technol        ISSN: 0895-3996            Impact factor:   1.535


  12 in total

1.  Three-dimensional artifact induced by projection weighting and misalignment.

Authors:  J Hsieh
Journal:  IEEE Trans Med Imaging       Date:  1999-04       Impact factor: 10.048

2.  Enhanced 3-D-reconstruction algorithm for C-arm systems suitable for interventional procedures.

Authors:  K Wiesent; K Barth; N Navab; P Durlak; T Brunner; O Schuetz; W Seissler
Journal:  IEEE Trans Med Imaging       Date:  2000-05       Impact factor: 10.048

3.  Implementation of a cone-beam reconstruction algorithm for the single-circle source orbit with embedded misalignment correction using homogeneous coordinates.

Authors:  M Karolczak; S Schaller; K Engelke; A Lutz; U Taubenreuther; K Wiesent; W Kalender
Journal:  Med Phys       Date:  2001-10       Impact factor: 4.071

Review 4.  Digital x-ray tomosynthesis: current state of the art and clinical potential.

Authors:  James T Dobbins; Devon J Godfrey
Journal:  Phys Med Biol       Date:  2003-10-07       Impact factor: 3.609

5.  A generic geometric calibration method for tomographic imaging systems with flat-panel detectors--a detailed implementation guide.

Authors:  Xinhua Li; Zhang Da; Bob Liu
Journal:  Med Phys       Date:  2010-07       Impact factor: 4.071

6.  Digital tomosynthesis with an on-board kilovoltage imaging device.

Authors:  Devon J Godfrey; Fang-Fang Yin; Mark Oldham; Sua Yoo; Christopher Willett
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-05-01       Impact factor: 7.038

Review 7.  Breast tomosynthesis: present considerations and future applications.

Authors:  Jeong Mi Park; Edmund A Franken; Megha Garg; Laurie L Fajardo; Loren T Niklason
Journal:  Radiographics       Date:  2007-10       Impact factor: 5.333

8.  A convenient alignment approach for x-ray imaging experiments based on laser positioning devices.

Authors:  Zhang Da; Molly Donovan; Xizeng Wu; Hong Liu
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

9.  Digital tomosynthesis in breast imaging.

Authors:  L T Niklason; B T Christian; L E Niklason; D B Kopans; D E Castleberry; B H Opsahl-Ong; C E Landberg; P J Slanetz; A A Giardino; R Moore; D Albagli; M C DeJule; P F Fitzgerald; D F Fobare; B W Giambattista; R F Kwasnick; J Liu; S J Lubowski; G E Possin; J F Richotte; C Y Wei; R F Wirth
Journal:  Radiology       Date:  1997-11       Impact factor: 11.105

10.  Compensating the intensity fall-off effect in cone-beam tomography by an empirical weight formula.

Authors:  Zikuan Chen; Vince D Calhoun; Shengjiang Chang
Journal:  Appl Opt       Date:  2008-11-10       Impact factor: 1.980

View more
  5 in total

1.  Characterization of a high-energy in-line phase contrast tomosynthesis prototype.

Authors:  Di Wu; Aimin Yan; Yuhua Li; Molly D Wong; Bin Zheng; Xizeng Wu; Hong Liu
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

2.  Perspective pinhole model with planar source for augmented reality surgical navigation based on C-arm imaging.

Authors:  Ho-Gun Ha; Sangseo Jeon; Seongpung Lee; Hyunseok Choi; Jaesung Hong
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-07-16       Impact factor: 2.924

3.  Image quality and dose efficiency of high energy phase sensitive x-ray imaging: phantom studies.

Authors:  Molly Donovan Wong; Xizeng Wu; Hong Liu
Journal:  J Xray Sci Technol       Date:  2014       Impact factor: 1.535

4.  Detectability comparison of simulated tumors in digital breast tomosynthesis using high-energy X-ray inline phase sensitive and commercial imaging systems.

Authors:  Muhammad U Ghani; Molly D Wong; Farid H Omoumi; Bin Zheng; Laurie L Fajardo; Aimin Yan; Xizeng Wu; Hong Liu
Journal:  Phys Med       Date:  2018-02-23       Impact factor: 2.685

5.  A geometric calibration method for the digital chest tomosynthesis with dual-axis scanning geometry.

Authors:  Chia-Hao Chang; Yu-Ching Ni; Syuan-Ya Huang; Ho-Hui Hsieh; Sheng-Pin Tseng; Fan-Pin Tseng
Journal:  PLoS One       Date:  2019-04-25       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.