A quality assurance method that utilizes 3D dosimetry and facilitates clinical interpretation.

Mark Oldham1, Andrew Thomas, Jennifer O'Daniel, Titania Juang, Geoffrey Ibbott, John Adamovics, John P Kirkpatrick.   

Abstract

PURPOSE: To demonstrate a new three-dimensional (3D) quality assurance (QA) method that provides comprehensive dosimetry verification and facilitates evaluation of the clinical significance of QA data acquired in a phantom. Also to apply the method to investigate the dosimetric efficacy of base-of-skull (BOS) intensity-modulated radiotherapy (IMRT) treatment. METHODS AND MATERIALS: Two types of IMRT QA verification plans were created for 6 patients who received BOS IMRT. The first plan enabled conventional 2D planar IMRT QA using the Varian portal dosimetry system. The second plan enabled 3D verification using an anthropomorphic head phantom. In the latter, the 3D dose distribution was measured using the DLOS/Presage dosimetry system (DLOS = Duke Large-field-of-view Optical-CT System, Presage Heuris Pharma, Skillman, NJ), which yielded isotropic 2-mm data throughout the treated volume. In a novel step, measured 3D dose distributions were transformed back to the patient's CT to enable calculation of dose-volume histograms (DVH) and dose overlays. Measured and planned patient DVHs were compared to investigate clinical significance.
RESULTS: Close agreement between measured and calculated dose distributions was observed for all 6 cases. For gamma criteria of 3%, 2 mm, the mean passing rate for portal dosimetry was 96.8% (range, 92.0%-98.9%), compared to 94.9% (range, 90.1%-98.9%) for 3D. There was no clear correlation between 2D and 3D passing rates. Planned and measured dose distributions were evaluated on the patient's anatomy, using DVH and dose overlays. Minor deviations were detected, and the clinical significance of these are presented and discussed.
CONCLUSIONS: Two advantages accrue to the methods presented here. First, treatment accuracy is evaluated throughout the whole treated volume, yielding comprehensive verification. Second, the clinical significance of any deviations can be assessed through the generation of DVH curves and dose overlays on the patient's anatomy. The latter step represents an important development that advances the clinical relevance of complex treatment QA.
Copyright © 2012. Published by Elsevier Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22361085      PMCID: PMC3832997          DOI: 10.1016/j.ijrobp.2011.12.015

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  13 in total

1.  CERR: a computational environment for radiotherapy research.

Authors:  Joseph O Deasy; Angel I Blanco; Vanessa H Clark
Journal:  Med Phys       Date:  2003-05       Impact factor: 4.071

2.  Monte carlo evaluation of the AAA treatment planning algorithm in a heterogeneous multilayer phantom and IMRT clinical treatments for an Elekta SL25 linear accelerator.

Authors:  E Sterpin; M Tomsej; B De Smedt; N Reynaert; S Vynckier
Journal:  Med Phys       Date:  2007-05       Impact factor: 4.071

3.  3D dosimetry by optical-CT scanning.

Authors:  Mark Oldham
Journal:  J Phys Conf Ser       Date:  2006

4.  Investigation of the feasibility of relative 3D dosimetry in the Radiologic Physics Center Head and Neck IMRT phantom using presage/optical-CT.

Authors:  Harshad Sakhalkar; David Sterling; John Adamovics; Geoffrey Ibbott; Mark Oldham
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

Review 5.  Dosimetry tools and techniques for IMRT.

Authors:  Daniel A Low; Jean M Moran; James F Dempsey; Lei Dong; Mark Oldham
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

6.  Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors.

Authors:  Benjamin E Nelms; Heming Zhen; Wolfgang A Tomé
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

7.  Commissioning and benchmarking a 3D dosimetry system for clinical use.

Authors:  Andrew Thomas; Joseph Newton; John Adamovics; Mark Oldham
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

8.  A method to correct for stray light in telecentric optical-CT imaging of radiochromic dosimeters.

Authors:  Andrew Thomas; Joseph Newton; Mark Oldham
Journal:  Phys Med Biol       Date:  2011-06-30       Impact factor: 3.609

9.  Establishing action levels for EPID-based QA for IMRT.

Authors:  Rebecca M Howell; Iris P N Smith; Christie S Jarrio
Journal:  J Appl Clin Med Phys       Date:  2008-06-23       Impact factor: 2.102

10.  A survey on planar IMRT QA analysis.

Authors:  Benjamin E Nelms; Jeff A Simon
Journal:  J Appl Clin Med Phys       Date:  2007-07-17       Impact factor: 2.102

View more
  19 in total

1.  A comprehensive investigation of the accuracy and reproducibility of a multitarget single isocenter VMAT radiosurgery technique.

Authors:  Andrew Thomas; Michael Niebanck; Titania Juang; Zhiheng Wang; Mark Oldham
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

2.  An investigation of a PRESAGE® in vivo dosimeter for brachytherapy.

Authors:  A K Vidovic; T Juang; S Meltsner; J Adamovics; J Chino; B Steffey; O Craciunescu; M Oldham
Journal:  Phys Med Biol       Date:  2014-06-24       Impact factor: 3.609

3.  Comprehensive quality assurance for base of skull IMRT.

Authors:  A Thomas; J O'Daniel; J Adamovics; G Ibbott; M Oldham
Journal:  J Phys Conf Ser       Date:  2013

4.  On the feasibility of optical-CT imaging in media of different refractive index.

Authors:  Leith Rankine; Mark Oldham
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

5.  On the feasibility of comprehensive high-resolution 3D remote dosimetry.

Authors:  Titania Juang; Ryan Grant; John Adamovics; Geoffrey Ibbott; Mark Oldham
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

6.  Comparison of 2D and 3D gamma analyses.

Authors:  Kiley B Pulliam; Jessie Y Huang; Rebecca M Howell; David Followill; Ryan Bosca; Jennifer O'Daniel; Stephen F Kry
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

7.  Utility and validation of biomechanical deformable image registration in low-contrast images.

Authors:  Michael Velec; Titania Juang; Joanne L Moseley; Mark Oldham; Kristy K Brock
Journal:  Pract Radiat Oncol       Date:  2015-03-29

8.  An investigation of PRESAGE® 3D dosimetry for IMRT and VMAT radiation therapy treatment verification.

Authors:  Jake Jackson; Titania Juang; John Adamovics; Mark Oldham
Journal:  Phys Med Biol       Date:  2015-02-16       Impact factor: 3.609

9.  Investigating the accuracy of microstereotactic-body-radiotherapy utilizing anatomically accurate 3D printed rodent-morphic dosimeters.

Authors:  Steven T Bache; Titania Juang; Matthew D Belley; Bridget F Koontz; John Adamovics; Terry T Yoshizumi; David G Kirsch; Mark Oldham
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

10.  Investigating end-to-end accuracy of image guided radiation treatment delivery using a micro-irradiator.

Authors:  L J Rankine; J Newton; S T Bache; S K Das; J Adamovics; D G Kirsch; M Oldham
Journal:  Phys Med Biol       Date:  2013-10-18       Impact factor: 3.609