| Literature DB >> 24567740 |
A Thomas1, J O'Daniel1, J Adamovics2, G Ibbott3, M Oldham1.
Abstract
Six base of skull IMRT treatment plans were delivered to Presage dosimeters within the RPC Head and Neck Phantom for quality assurance (QA) verification. Isotropic 2mm 3D data were acquired by optical-CT scanning with the DLOS system (Duke Large Optical-CT Scanner) and compared to the Eclipse (Varian) treatment plan. Normalized Dose Distribution (NDD) pass rates were obtained for a number of criteria. High quality 3D dosimetry data was observed from the DLOS system, illustrated here through colormaps, isodose lines, and profiles. Excellent agreement with the planned dose distributions was also observed with NDD analysis revealing > 90% pass rates (with criteria 3%, 2mm), and noise < 0.5%. The results comprehensively confirm the high accuracy of base-of-skull IMRT treatment in our clinic.Entities:
Year: 2013 PMID: 24567740 PMCID: PMC3932065 DOI: 10.1088/1742-6596/444/1/012050
Source DB: PubMed Journal: J Phys Conf Ser ISSN: 1742-6588