Comprehensive quality assurance for base of skull IMRT.

A Thomas1, J O'Daniel1, J Adamovics2, G Ibbott3, M Oldham1.   

Abstract

Six base of skull IMRT treatment plans were delivered to Presage dosimeters within the RPC Head and Neck Phantom for quality assurance (QA) verification. Isotropic 2mm 3D data were acquired by optical-CT scanning with the DLOS system (Duke Large Optical-CT Scanner) and compared to the Eclipse (Varian) treatment plan. Normalized Dose Distribution (NDD) pass rates were obtained for a number of criteria. High quality 3D dosimetry data was observed from the DLOS system, illustrated here through colormaps, isodose lines, and profiles. Excellent agreement with the planned dose distributions was also observed with NDD analysis revealing > 90% pass rates (with criteria 3%, 2mm), and noise < 0.5%. The results comprehensively confirm the high accuracy of base-of-skull IMRT treatment in our clinic.

Entities:  

Year:  2013        PMID: 24567740      PMCID: PMC3932065          DOI: 10.1088/1742-6596/444/1/012050

Source DB:  PubMed          Journal:  J Phys Conf Ser        ISSN: 1742-6588


  13 in total

1.  CERR: a computational environment for radiotherapy research.

Authors:  Joseph O Deasy; Angel I Blanco; Vanessa H Clark
Journal:  Med Phys       Date:  2003-05       Impact factor: 4.071

2.  On dose distribution comparison.

Authors:  Steve B Jiang; Greg C Sharp; Toni Neicu; Ross I Berbeco; Stella Flampouri; Thomas Bortfeld
Journal:  Phys Med Biol       Date:  2006-01-25       Impact factor: 3.609

3.  Fast laser scanning optical-CT apparatus for 3D radiation dosimetry.

Authors:  Nikola Krstajić; Simon J Doran
Journal:  Phys Med Biol       Date:  2007-05-04       Impact factor: 3.609

4.  3D dosimetry by optical-CT scanning.

Authors:  Mark Oldham
Journal:  J Phys Conf Ser       Date:  2006

5.  Investigation of the feasibility of relative 3D dosimetry in the Radiologic Physics Center Head and Neck IMRT phantom using presage/optical-CT.

Authors:  Harshad Sakhalkar; David Sterling; John Adamovics; Geoffrey Ibbott; Mark Oldham
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

6.  Investigation of radiological properties and water equivalency of PRESAGE dosimeters.

Authors:  Tina Gorjiara; Robin Hill; Zdenka Kuncic; John Adamovics; Stephen Bosi; Jung-Ha Kim; Clive Baldock
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

7.  Commissioning and benchmarking a 3D dosimetry system for clinical use.

Authors:  Andrew Thomas; Joseph Newton; John Adamovics; Mark Oldham
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

8.  A quality assurance method that utilizes 3D dosimetry and facilitates clinical interpretation.

Authors:  Mark Oldham; Andrew Thomas; Jennifer O'Daniel; Titania Juang; Geoffrey Ibbott; John Adamovics; John P Kirkpatrick
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-02-22       Impact factor: 7.038

Review 9.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

10.  Radiological properties of the PRESAGE and PAGAT polymer dosimeters.

Authors:  S Brown; A Venning; Y De Deene; P Vial; L Oliver; J Adamovics; C Baldock
Journal:  Appl Radiat Isot       Date:  2008-06-11       Impact factor: 1.513

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.