Literature DB >> 16111598

Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators.

Coen W Hurkmans1, Egon Scheepers, Bob G F Springorum, Hans Uiterwaal.   

Abstract

PURPOSE: Radiotherapy can influence the functioning of pacemakers and implantable cardioverter-defibrillators (ICDs). ICDs offer the same functionality as pacemakers, but are also able to deliver a high-voltage shock to the heart if needed. Guidelines for radiotherapy treatment of patients with an implanted rhythm device have been published in 1994 by The American Association of Physicists in Medicine, and are based only on experience with pacemakers. Data on the influence of radiotherapy on ICDs are limited. The objective of our study is to determine the influence of radiotherapy on the latest generation of ICDs. METHODS AND MATERIALS: Eleven modern ICDs have been irradiated in our department. The irradiation was performed with a 6-MV photon beam. The given dose was fractionated up to a cumulative dose of 120 Gy. Two to 5 days passed between consecutive irradiations. Frequency, output, sensing, telemetry, and shock energy were monitored.
RESULTS: Sensing interference by ionizing radiation on all ICDs has been demonstrated. For four ICDs, this would have caused the inappropriate delivery of a shock because of interference. At the end of the irradiation sessions, all devices had reached their point of failure. Complete loss of function was observed for four ICDs at dose levels between 0.5 Gy and 1.5 Gy.
CONCLUSIONS: The effect of radiation therapy on the newest generation of ICDs varies widely. If tachycardia monitoring and therapy are functional (programmed on) during irradiation, the ICD might inappropriately give antitachycardia therapy, often resulting in a shock. Although most ICDs did not fail below 80 Gy, some devices had already failed at doses below 1.5 Gy. Guidelines are formulated for the treatment of patients with an ICD.

Entities:  

Mesh:

Year:  2005        PMID: 16111598     DOI: 10.1016/j.ijrobp.2005.04.047

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  27 in total

Review 1.  [Perioperative management of patients with implanted pacemakers or cardioverter/defibrillators. Recommendations of the Austrian Society for Anaesthesiology, Resuscitation and Intensive Care Medicine, the Austrian Society for Cardiology and the Austrian Society for Surgery].

Authors:  H Gombotz; M Anelli Monti; N Leitgeb; M Nürnberg; B Strohmer
Journal:  Anaesthesist       Date:  2009-05       Impact factor: 1.041

2.  Radiation tolerance of contemporary implantable cardioverter-defibrillators.

Authors:  Michael Mollerus; Leslee Naslund; Margaret Lipinski; Anne Meyer; Bruce Libey; Ken Dornfeld
Journal:  J Interv Card Electrophysiol       Date:  2013-12-07       Impact factor: 1.900

Review 3.  Arrhythmias and Other Electrophysiology Issues in Cancer Patients Receiving Chemotherapy or Radiation.

Authors:  Federico Viganego; Robin Singh; Michael G Fradley
Journal:  Curr Cardiol Rep       Date:  2016-06       Impact factor: 2.931

4.  A randomized in vitro evaluation of transient and permanent cardiac implantable electronic device malfunctions following direct exposure up to 10 Gy.

Authors:  Maria Daniela Falco; Domenico Genovesi; Luciana Caravatta; Clelia Di Carlo; Ekaterina Bliakharskaia; Marianna Appignani; Massimiliano Faustino; Nanda Furia; Enrico Di Girolamo
Journal:  Strahlenther Onkol       Date:  2020-06-19       Impact factor: 3.621

Review 5.  [Radiation therapy in patients with cardiac pacemakers or implantable cardioverter defibrillators. Interdisciplinary safety recommendations].

Authors:  M Dorenkamp; C Stromberger; C von Heymann; W Haverkamp; P Wust; M Roser
Journal:  Strahlenther Onkol       Date:  2012-11-11       Impact factor: 3.621

6.  Effect of X-ray dose rates higher than 8 Gy/min on the functioning of cardiac implantable electronic devices.

Authors:  Kazuhiko Nakamura; Takahiro Aoyama; Naoki Kaneda; Masashi Otsuji; Yoshitaka Minami; Ami Sakuragi; Masaru Nakamura
Journal:  J Radiat Res       Date:  2020-05-22       Impact factor: 2.724

7.  DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices.

Authors:  Benjamin Gauter-Fleckenstein; Carsten W Israel; Marc Dorenkamp; Jürgen Dunst; Mattias Roser; Rainer Schimpf; Volker Steil; Jörg Schäfer; Ulrike Höller; Frederik Wenz
Journal:  Strahlenther Onkol       Date:  2015-03-05       Impact factor: 3.621

8.  Radiotherapy of patients with cardiac implantable electronic devices according to the DEGRO/DGK guideline-is the risk of relevant errors overestimated?

Authors:  Felix Steger; Matthias G Hautmann; Christoph Süß; Ute Hubauer; Ekrem Ücer; Lars Maier; Oliver Kölbl; Carsten Jungbauer
Journal:  Strahlenther Onkol       Date:  2019-08-09       Impact factor: 3.621

9.  A clinical example of extreme dose exposure for an implanted cardioverter-defibrillator : Beyond the DEGRO guidelines.

Authors:  Yoana Hristova; Janett Köhn; Stefanie Preuß; Claus Rödel; Panagiotis Balermpas
Journal:  Strahlenther Onkol       Date:  2017-05-31       Impact factor: 3.621

10.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690