Dose reduction and image quality in CT angiography for cerebral aneurysm with various tube potentials and current settings.

K Imai1, M Ikeda, C Kawaura, T Aoyama, Y Enchi, M Yamauchi.   

Abstract

OBJECTIVES: The purpose of this study is to investigate the image quality on both axial and three-dimensional CT angiograms of the brain at various tube potentials and currents, and to propose the use of descriptors for evaluating the image quality of three-dimensional CT angiograms using entropy analysis.
METHODS: A head phantom was used as a target object. Axial CT and three-dimensional CT angiograms were obtained at various effective milliampere-second values (49-350 mAs) and tube potentials (80-140 kVp) with a 64-row detector CT scanner. Lens doses were measured using a planar silicon pin-photodiode system. The signal-to-noise ratio (SNR) and streak artefacts on the axial CT angiograms were evaluated and the image quality of the three-dimensional CT angiograms was assessed using entropy analysis.
RESULTS: Lens doses increased with tube potential and effective milliampere-seconds. From the evaluation of SNR and streak artefacts on axial CT angiograms, we found that the image quality was improved by setting the tube potential at 100 kVp. However, there was little visual difference in the image quality for 100 kVp between 252 (effective value recommended by the manufacturer) and 350 mAs (maximum effective value). In the entropy analysis of the image quality of three-dimensional CT angiograms, the mutual information (information gain) per lens dose was largest at 80 kVp and 252 mAs.
CONCLUSION: Our results suggested that the suitable tube potentials for axial CT and three-dimensional CT angiograms were 100 and 80 kVp, respectively, and the effective milliampere-second value recommended by the manufacturer was appropriate.

Entities:  

Mesh:

Year:  2012        PMID: 22253346      PMCID: PMC3487084          DOI: 10.1259/bjr/63268088

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  20 in total

1.  CT of the head by use of reduced current and kilovoltage: relationship between image quality and dose reduction.

Authors:  M Cohnen; H Fischer; J Hamacher; E Lins; R Kötter; U Mödder
Journal:  AJNR Am J Neuroradiol       Date:  2000-10       Impact factor: 3.825

2.  Subtle lung nodules: influence of local anatomic variations on detection.

Authors:  Ehsan Samei; Michael J Flynn; Edward Peterson; William R Eyler
Journal:  Radiology       Date:  2003-05-15       Impact factor: 11.105

3.  Multi-detector row CT angiography of the brain at various kilovoltage settings.

Authors:  Birgit B Ertl-Wagner; Ralf-Thorsten Hoffmann; Roland Bruning; Karin Herrmann; Brad Snyder; Jeffrey D Blume; Maximilian F Reiser
Journal:  Radiology       Date:  2004-03-24       Impact factor: 11.105

4.  Comparison of image quality between conventional and low-dose nonenhanced head CT.

Authors:  Mark E Mullins; Michael H Lev; Peter Bove; Cara E O'Reilly; Sanjay Saini; James T Rhea; James H Thrall; George J Hunter; Leena M Hamberg; R Gilberto Gonzalez
Journal:  AJNR Am J Neuroradiol       Date:  2004-04       Impact factor: 3.825

5.  Improved vascular opacification in cerebral computed tomography angiography with 80 kVp.

Authors:  Malte L Bahner; Andreas Bengel; Gunnar Brix; Ivan Zuna; Hans-Ulrich Kauczor; Stefan Delorme
Journal:  Invest Radiol       Date:  2005-04       Impact factor: 6.016

6.  A detection method for streak artifacts and radiological noise in a non-uniform region in a CT image.

Authors:  Kuniharu Imai; Mitsuru Ikeda; Yukihiro Enchi; Takanaga Niimi
Journal:  Phys Med       Date:  2009-12-29       Impact factor: 2.685

7.  Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey.

Authors:  G Brix; H D Nagel; G Stamm; R Veit; U Lechel; J Griebel; M Galanski
Journal:  Eur Radiol       Date:  2003-04-10       Impact factor: 5.315

8.  Patient size and x-ray technique factors in head computed tomography examinations. II. Image quality.

Authors:  Walter Huda; Kristin A Lieberman; Jack Chang; Marsha L Roskopf
Journal:  Med Phys       Date:  2004-03       Impact factor: 4.071

9.  An in-phantom dosimetry system using pin silicon photodiode radiation sensors for measuring organ doses in x-ray CT and other diagnostic radiology.

Authors:  Takahiko Aoyama; Shuji Koyama; Chiyo Kawaura
Journal:  Med Phys       Date:  2002-07       Impact factor: 4.071

10.  Evaluation of cerebral aneurysms with helical CT: correlation with conventional angiography and MR angiography.

Authors:  R B Schwartz; H M Tice; S M Hooten; L Hsu; P E Stieg
Journal:  Radiology       Date:  1994-09       Impact factor: 11.105

View more
  6 in total

1.  Feasibility of ultra-low radiation dose digital subtraction angiography: Preliminary study in a simplified cerebral angiography phantom.

Authors:  Jun Young Maeng; Yunsun Song; Yu Sub Sung; Tae-Il Kim; Deok Hee Lee; Tae-Hyung Kim
Journal:  Interv Neuroradiol       Date:  2019-05-16       Impact factor: 1.610

2.  Low tube voltage and low contrast material volume cerebral CT angiography.

Authors:  Song Luo; Long Jiang Zhang; Felix G Meinel; Chang Sheng Zhou; Li Qi; Andrew D McQuiston; U Joseph Schoepf; Guang Ming Lu
Journal:  Eur Radiol       Date:  2014-05-04       Impact factor: 5.315

3.  Arterial contour detectability in head CT angiography.

Authors:  Yukihiro Enchi; Kuniharu Imai; Mitsuru Ikeda; Ikuko Takase; Chiyo Yamauchi-Kawaura; Masaki Mori
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-04-24       Impact factor: 2.924

4.  Subcutaneous tissue thickness is an independent predictor of image noise in cardiac CT.

Authors:  Henrique Lane Staniak; Rodolfo Sharovsky; Alexandre Costa Pereira; Cláudio Campi de Castro; Isabela M Benseñor; Paulo A Lotufo; Márcio Sommer Bittencourt
Journal:  Arq Bras Cardiol       Date:  2013-11-01       Impact factor: 2.000

5.  Contrast enhancement efficacy of iodinated contrast media: Effect of molecular structure on contrast enhancement.

Authors:  Kuniharu Imai; Mitsuru Ikeda; Yoshiki Satoh; Keisuke Fujii; Chiyo Kawaura; Takuya Nishimoto; Masaki Mori
Journal:  Eur J Radiol Open       Date:  2018-10-06

6.  Metal artifacts reduction in computed tomography: A phantom study to compare the effectiveness of metal artifact reduction algorithm, model-based iterative reconstruction, and virtual monochromatic imaging.

Authors:  Takuya Ishikawa; Shigeru Suzuki; Shingo Harashima; Rika Fukui; Masafumi Kaiume; Yoshiaki Katada
Journal:  Medicine (Baltimore)       Date:  2020-12-11       Impact factor: 1.817

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.