Feasibility of ultra-low radiation dose digital subtraction angiography: Preliminary study in a simplified cerebral angiography phantom.

Jun Young Maeng1, Yunsun Song1, Yu Sub Sung1, Tae-Il Kim1, Deok Hee Lee1, Tae-Hyung Kim2.   

Abstract

INTRODUCTION: The objective of this article is to evaluate the feasibility of cerebral digital subtraction angiography (DSA) using ultra-low radiation dose settings in a simplified cerebral angiography phantom.
MATERIALS AND METHODS: We created a silicone phantom capable of producing a simplified cerebral DSA. A total of 18 DSA sets were obtained with gradual six-step reduction of the detector entrance dose (DED) from 1.82 to 0.08 μGy per frame, while standard, postprocessing algorithm (PPA) and copper filter (0.3 mm) with PPA (CwP) algorithm reconstruction protocols were applied. We quantitatively compared their signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) and qualitatively analyzed the images' qualities in terms of image sharpness, contrast, and noise as investigated by five observers.
RESULTS: The SNR and CNR, which decreased with lowering of the DED in the standard protocol group, were significantly compensated by using the PPA. The values were approximately double in the PPA (11.5 ± 2.9) and CwP (11.0 ± 2.5) groups compared with the standard (5.4 ± 1.1) group in the DED of 0.24 μGy per frame as well as in the other values. The total scores of the observers according to the protocols showed a tendency to decrease as the DED lowered. On average, the PPA (96.3 ± 34.6) and CwP (91.3 ± 29.9) groups yielded higher results than the standard protocol (83.7 ± 46.7).
CONCLUSION: Given that the current DED ranges from 1.82 to 3.60 μGy per frame for routine cerebral DSA, our results indicate that DED can be decreased to 15%-30% of the current dose level in vessels 2-4 mm in diameter if image-improvement algorithms are applied.

Entities:  

Keywords:  Cerebral angiography; dose reduction; image quality compensation; radiation dose; radiation exposure

Mesh:

Year:  2019        PMID: 31096837      PMCID: PMC6777105          DOI: 10.1177/1591019919850302

Source DB:  PubMed          Journal:  Interv Neuroradiol        ISSN: 1591-0199            Impact factor:   1.610


  19 in total

1.  Low-dose 2D X-ray angiography enhancement using 2-axis PCA for the preservation of blood-vessel region and noise minimization.

Authors:  Yong Geun Lee; Jeongjin Lee; Yeong-Gil Shin; Ho Chul Kang
Journal:  Comput Methods Programs Biomed       Date:  2015-09-28       Impact factor: 5.428

2.  Reduction of radiation dose for cerebral angiography using flat panel detector of direct conversion type: a vascular phantom study.

Authors:  Y Hatakeyama; S Kakeda; N Ohnari; J Moriya; N Oda; K Nishino; W Miyamoto; Y Korogi
Journal:  AJNR Am J Neuroradiol       Date:  2007-04       Impact factor: 3.825

3.  Radiation dose and image quality in neuroangiography: effects of increased tube voltage, added x-ray filtration and antiscatter grid removal.

Authors:  M Söderman; B Hansson; B Axelsson
Journal:  Interv Neuroradiol       Date:  2001-05-15       Impact factor: 1.610

4.  SYSTEM UPGRADE ON PHILIPS ALLURA FD20 ANGIOGRAPHY SYSTEMS: EFFECTS ON PATIENT SKIN DOSE AND STATIC IMAGE QUALITY.

Authors:  Nick Ryckx; Marta Sans-Merce; Reto Meuli; Jean-Baptiste Zerlauth; Francis R Verdun
Journal:  Radiat Prot Dosimetry       Date:  2015-11-29       Impact factor: 0.972

5.  Reduced Patient Radiation Exposure during Neurodiagnostic and Interventional X-Ray Angiography with a New Imaging Platform.

Authors:  K van der Marel; S Vedantham; I M J van der Bom; M Howk; T Narain; K Ty; A Karellas; M J Gounis; A S Puri; A K Wakhloo
Journal:  AJNR Am J Neuroradiol       Date:  2017-01-19       Impact factor: 3.825

6.  Patient radiation dose in diagnostic and interventional procedures for intracranial aneurysms: experience at a single center.

Authors:  Chang Woo Chun; Bum-Soo Kim; Cheol Hyoun Lee; Yon Kwon Ihn; Yong-Sam Shin
Journal:  Korean J Radiol       Date:  2014-11-07       Impact factor: 3.500

7.  Copper filtration in pediatric digital X-ray imaging: its impact on image quality and dose.

Authors:  Philippe Brosi; Anja Stuessi; Francis R Verdun; Peter Vock; Rainer Wolf
Journal:  Radiol Phys Technol       Date:  2011-03-23

8.  Patient Radiation Exposure During Diagnostic and Therapeutic Procedures for Intracranial Aneurysms: A Multicenter Study.

Authors:  Yon Kwon Ihn; Bum-Soo Kim; Jun Soo Byun; Sang Hyun Suh; Yoo Dong Won; Deok Hee Lee; Byung Moon Kim; Young Soo Kim; Pyong Jeon; Chang-Woo Ryu; Sang-Il Suh; Dae Seob Choi; See Sung Choi; Jin Wook Choi; Hyuk Won Chang; Jae-Wook Lee; Sang Heum Kim; Young Jun Lee; Shang Hun Shin; Soo Mee Lim; Woong Yoon; Hae Woong Jeong; Moon Hee Han
Journal:  Neurointervention       Date:  2016-09-03

9.  Radiation dose reduction during neurointerventional procedures by modification of default settings on biplane angiography equipment.

Authors:  Elyne N Kahn; Joseph J Gemmete; Neeraj Chaudhary; Byron Gregory Thompson; Kevin Chen; Emmanuel G Christodoulou; Aditya S Pandey
Journal:  J Neurointerv Surg       Date:  2015-08-05       Impact factor: 5.836

10.  Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614 patients.

Authors:  Michael Söderman; Maria Mauti; Sjirk Boon; Artur Omar; María Marteinsdóttir; Tommy Andersson; Staffan Holmin; Bart Hoornaert
Journal:  Neuroradiology       Date:  2013-09-05       Impact factor: 2.804

View more
  2 in total

1.  Feasibility of low-dose digital subtraction angiography protocols for the endovascular treatment of intracranial dural arteriovenous fistulas.

Authors:  Yunsun Song; Seongsik Han; Byung Jun Kim; Seong Heum Oh; Jin Su Kim; Tae Il Kim; Deok Hee Lee
Journal:  Neuroradiology       Date:  2020-08-28       Impact factor: 2.804

2.  Low-Dose Three-Dimensional Rotational Angiography for Evaluating Intracranial Aneurysms: Analysis of Image Quality and Radiation Dose.

Authors:  Hee Jong Ki; Bum-Soo Kim; Jun-Ki Kim; Jai Ho Choi; Yong Sam Shin; Yangsean Choi; Na-Young Shin; Jinhee Jang; Kook-Jin Ahn
Journal:  Korean J Radiol       Date:  2022-01-04       Impact factor: 3.500

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.