Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT.

Petr Symersky1, Jesse Habets, Paul Westers, Bas A J M de Mol, Mathias Prokop, Ricardo P J Budde.   

Abstract

OBJECTIVES: Multidetector computed tomography (MDCT) has diagnostic value for the evaluation of prosthetic heart valve (PHV) dysfunction but it is hampered by artefacts. We hypothesised that image acquisition using prospective triggering instead of retrospective gating would reduce artefacts related to pulsating PHV.
METHODS: In a pulsatile in vitro model, a mono- and bileaflet PHV were imaged using 256 MDCT at 60, 75 and 90 beats per minute (BPM) with either retrospective gating (120 kV, 600 mAs, pitch 0.2, CTDI(vol) 39.8 mGy) or prospective triggering (120 kV, 200 mAs, CTDI(vol) 13.3 mGy). Two thresholds (>175 and <-45HU), derived from the density of surrounding structures, were used for quantification of hyper- and hypodense artefacts. Image noise and artefacts were compared between protocols.
RESULTS: Prospective triggering reduced hyperdense artefacts for both valves at every BPM (P = 0.001 all comparisons). Hypodense artefacts were reduced for the monoleaflet valve at 60 (P = 0.009), 75 (P = 0.016) and 90 BPM (P = 0.001), and for the bileaflet valves at 60 (P = 0.001), 90 (P = 0.001) but not at 75 BPM (P = 0.6). Prospective triggering reduced image noise at 60 (P = 0.001) and 75 (P < 0.03) but not at 90 BPM.
CONCLUSIONS: Compared with retrospective gating, prospective triggering reduced most artefacts related to pulsating PHV in vitro. KEY POINTS: • Computed tomographic images are often degraded by prosthetic heart valve-induced artefacts • Prospective triggering reduces prosthetic heart valve-induced artefacts in vitro • Artefact reduction at 90 beats per minute occurs without image noise reduction • Prospective triggering may improve CT image quality of moving hyperdense structures.

Entities:  

Mesh:

Year:  2011        PMID: 22205446      PMCID: PMC3348472          DOI: 10.1007/s00330-011-2358-1

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


Introduction

Electrocardiography (ECG)-gated multidetector computed tomography (MDCT) can identify the morphological substrates of prosthetic heart valve (PHV) dysfunction [1-4]. Echocardiography, which is the conventional method of monitoring PHV function, may not always identify the cause of dysfunction because of acoustic shadowing [4-6]. In regions obscured by acoustic shadowing, MDCT has been shown to identify obstructive masses, thrombus, vegetations and degenerated thickened leaflets of biological prostheses [1–4, 7]. The identification of these causes of dysfunction may guide clinical management and may even form an indication for the surgical replacement of dysfunctional valves [1, 2, 4, 8, 9]. These imaging results have been achieved mostly with 64-slice systems using retrospectively ECG-gated coronary CT angiography protocols. With retrospectively gated acquisition, the in vitro and in vivo image quality of modern mechanical PHV was found to be generally good despite variable amounts of artefacts [1, 10–12]. The artefacts seemed to be related to the radiopaque parts of the PHV, such as the metal alloy of the prosthetic ring and the tungsten impregnated carbon leaflets, and consisted of low attenuation (hypodense) and high attenuation (hyperdense) components [10, 12]. Because of their resemblance to metal artefacts, probably similar causal mechanisms such as scatter, photon starvation, edge effects, beam hardening and motion may play a role [13-15]. Some of these interactions can be influenced by modifying acquisition parameters, other factors (such as motion) are patient-dependent. A pragmatic approach has been the use of modified coronary protocols with retrospective gating but with increased tube currents and voltages [3, 7]. Although these reports yielded good image quality, it was at the expense of sizable radiation exposure, which may curtail the use of CT for PHV evaluation and withhold the potential advantages of the technique to all but highly selected patients. Other adaptations for enhanced PHV imaging have not been studied and no specific PHV MDCT protocols have been developed to date. With the introduction of 256- and 320-detector systems, PHV can be imaged within one rotation using prospective ECG-triggering. Experiments with an in vitro model of coronary in-stent restenosis suggested that prospective triggering may be associated with improved image quality compared with retrospective gating [16]. In addition, prospective triggering may achieve this with a fraction of the radiation dose compared with modified coronary protocols that have been used for PHV imaging. Because PHV also appear as high density objects on CT, a similar improvement may be possible for prospectively triggered imaging of PHV. In order to test this hypothesis we compared a retrospectively gated and a prospectively triggered protocol using a 256-detector system for imaging the two most common bileaflet and tilting disc mechanical PHVs in a pulsatile in vitro model.

Materials and methods

Two mechanical PHVs, St Jude bileaflet (SJ, St Jude Medical Inc., St Paul, MN, USA, valve size 27 mm) and Medtronic Hall tilting disc (MH, Medtronic Inc., Minneapolis, MN, USA, valve size 27 mm, Fig. 1), were inserted into a pulsatile in vitro model, which was described by Symersky et al. [10]. In brief, the valves were mounted in a polymethylmethacrylate (PMMA) valve chamber (Fig. 2) that was placed in a thoracic phantom (QRM GmbH, Möhrendorf, Germany). The valve chamber was connected to a computer controlled piston pump and to an afterload reservoir. Water was used as a perfusate. The flow pulses were at a frequency of 60, 75 and 90 beats per minute (BPM) and produced identical pulsation cycles. The valves were opened from 30 to 50% of the artificial ECG interval generated by the computer.
Fig. 1

Photograph of a St Jude (a) and Medtronic Hall aortic valve (b), both pictured from the inflow (ventricular) side. (Images courtesy of St Jude Medical Inc., St Paul, MN, USA, and Medtronic Inc., Minneapolis, MN, USA)

Fig. 2

The PMMA cylindrical valve chamber in which a mechanical prosthetic valve was tightly mounted (arrow) under a 45° angle to the gantry (a). Only leaflet motion was possible. The valve chamber was then positioned in a thoracic phantom (b) and connected to the piston pump for leaflet motion

Photograph of a St Jude (a) and Medtronic Hall aortic valve (b), both pictured from the inflow (ventricular) side. (Images courtesy of St Jude Medical Inc., St Paul, MN, USA, and Medtronic Inc., Minneapolis, MN, USA) The PMMA cylindrical valve chamber in which a mechanical prosthetic valve was tightly mounted (arrow) under a 45° angle to the gantry (a). Only leaflet motion was possible. The valve chamber was then positioned in a thoracic phantom (b) and connected to the piston pump for leaflet motion Imaging was performed using 256 MDCT (iCT, Philips Medical Systems, Best, The Netherlands). Two imaging protocols were used: 1) a standard helical retrospectively ECG-gated protocol based on coronary CT angiography, which has been used for PHV imaging [2, 3, 7], and 2) a prospectively ECG-triggered protocol. CT parameters are presented in Table 1. Prospective triggering was performed with 128 × 0.625 mm collimation and axial imaging with no table movement and 8 cm exposed at the centre of the gantry. The maximal range that could be reconstructed from these data was 70 mm. No padding was used. Retrospective gating was also performed with 128 × 0.625 mm collimation at a pitch of 0.2. We chose the smallest possible anatomical range (80.1 mm), which could be covered with the use of this protocol. CT data acquisition only occurred during the 40% ECG interval (opened PHV) with prospective triggering and reconstructed with retrospective gating. Each valve was imaged 8 times using each protocol, yielding a total of 16 acquisitions per valve. Images were reconstructed at 0.9 mm thick slices with 0.45 mm reconstruction increment. These images were used for HU measurements as well as 3D volume rendering. Image acquisition with both protocols was performed at 60, 75 and 90 BPM.
Table 1

CT parameters

ProspectiveRetrospective
One ECG intervalAll ECG intervals
Collimation128 × 0.625 mm128 × 0.625 mm
kV120120
mAs200600
Pitch00.2
Rotation time (ms)270270
FilterCardiac BCardiac B
Anatomical length (mm)8080.1
CTDIvol (mGy)13.339.8
DLP106.5517.8
CT parameters All image sets were transferred to a dedicated workstation for analysis (Extended Brilliance Workstation, Philips Medical Systems, Best, The Netherlands). Hyper- and hypodense artefact volumes were quantified in 3D-volume rendered images using two thresholds based on the densities of the surrounding structures according to a methodology described elsewhere for neurosurgical clips [13]. The PMMA in the valve chamber had a density of 130 ± 15 HU, and water 1 ± 16 HU (mean ± standard deviation). We chose threshold values that were approximately 3 standard deviations (SD) above the HU of PMMA measurement and below the HU of water, respectively. The chosen thresholds were 175 HU for hyperdense artefacts and −45 HU for hypodense artefacts. Because the PHVs are composed of various radiopaque components such as a titanium ring and tungsten impregnated leaflets, which have CT densities higher than 800 HU [12], the 175-HU threshold included the radiopaque components of the PHV. Hence, the percentage change of the volume measured with the 175-HU threshold underestimates the change in actual artefacts (i.e. the volume measured with the 175-HU threshold without the volume of radiopaque components with densities over 800 HU) [12]. Areas outside the valve chamber and other unrelated sources of artefacts were digitally excised in an identical manner for all images. Image noise (defined as the SD of CT attenuation) was measured in all images using a circular region of interest (diameter 1 cm) that was placed in an identical section of the PMMA structure of the valve chamber.

Statistical analysis

Data were analysed using SPSS software (SPSS Statistics Version 16.0, SPSS Inc, Chicago, IL, USA). Non-parametric data were presented as medians with interquartile range (IQR). A Mann–Whitney U test was used for the comparison of artefact volumes and image noise measurements. Statistical significance was defined as P < 0.05.

Results

The median artefact volumes for the two thresholds (−45 and 175 HU) at different frequencies are summarised in Table 2. For the St Jude valve, prospective triggering reduced hypodense artefacts (all volume <−45 HU) at 60 BPM (P = 0.001), 90 BPM (P = 0.001) but not 75 BPM (P = 0.6). Hyperdense artefacts (all volume >175 HU) were reduced with prospective triggering at 60 BPM (P = 0.001), 75 BPM (P = 0.001) and 90 BPM (P = 0.001). For the Medtronic Hall valve, hypodense artefacts were reduced with prospective triggering at 60 BPM (P = 0.009), 75 BPM (P = 0.016) and 90 BPM (P = 0.001). Hyperdense artefacts were also reduced at all frequencies (P = 0.001 for 60, 75, and 90 BPM, see Figs. 3 and 4). Volume rendered images illustrate the reduction of artefacts emanating from the leaflet and artefacts related to the prosthetic ring (Fig. 5). The image noise measurements are presented in Table 3. Although significant differences were found between retrospective gating and prospective triggering at 60 and 75 BPM, the image noise did not differ at 90 BPM.
Table 2

Artefacts measured in opened (systolic) valves

ValveBPMThresholdMedian artefact volume (mm3) (interquartile range)
RetrospectiveProspective
St Jude60<−45 HU2,056 (1,939–2,263)1,168 (1,146–1,201)
>175 HU9,000 (8,725–10,375)7,000 (6,925–7,150)
75<−45 HU1,609 (1,567–1,841)1,609 (1,558–1,641)
>175 HU8,400 (7,625–8,775)6,900 (6,700–7,275)
90<−45 HU2,742 (1,964–2,896)1,564 (1,527–1,589)
>175 HU9,700 (9,100–10,300)7,000 (6,800–7,450)
Medtronic hall60<−45 HU2,271 (1,873–2,630)1,571 (1,474–1,724)
>175 HU8,400 (8,300–9,225)6,150 (5,950–6,525)
75<−45 HU2,192 (1,788–2,522)1,608 (1,446–1,730)
>175 HU7,550 (7,400–8,075)5,800 (5,625–6,075)
90<−45 HU3,164 (2,547–3,770)1,774 (1,723–1,977)
>175 HU8,400 (7,575–10,000)5,950 (5,725–6,100)
Fig. 3

Multiplanar reformatted images of the St Jude valve in a view perpendicular to the leaflets acquired with retrospective gating (a) and prospective triggering (b) at 60 BPM. Note the increased deformation of the subprosthetic PMMA contours and increased hyper- and hypodense artefacts in (a)

Fig. 4

Changes in a hyperdense and b hypodense artefacts

Fig. 5

Volume rendered images of the 3D volume of a St Jude PHV at 60 BPM measured with the −45 HU threshold (including all volumes less than −45 HU) and with the 175 HU threshold (including all volumes with a density higher than 175 HU). Top panel artefact configurations with retrospective gating. Bottom panel images with prospective triggering. The 175 HU threshold volume includes all radiopaque parts of the valve such as the leaflets and the prosthetic ring

Table 3

Image noise (median and interquartile range) for retrospective gating and prospective triggering

ValveBeats per minuteRetrospective gatingProspective triggering P valuesa
St Jude6016.0 (14.5–17.0)11.0 (10.3–12.0)0.001
7518.0 (17.0–18.8)13.5 (12.3–14.8)0.001
9014.0 (12.3–16.0)13.0 (12.3–13.8)0.198
Medtronic hall6016.0 (14.3–17.0)11.0 (10.3–11.8)0.001
7517.5 (15.0–19.8)14.0 (12.5–16.0)0.026
9015.0 (14.3–15.8)14.5 (13.0–15.0)0.269

a The P values result from Mann–Whitney U test

Artefacts measured in opened (systolic) valves Multiplanar reformatted images of the St Jude valve in a view perpendicular to the leaflets acquired with retrospective gating (a) and prospective triggering (b) at 60 BPM. Note the increased deformation of the subprosthetic PMMA contours and increased hyper- and hypodense artefacts in (a) Changes in a hyperdense and b hypodense artefacts Volume rendered images of the 3D volume of a St Jude PHV at 60 BPM measured with the −45 HU threshold (including all volumes less than −45 HU) and with the 175 HU threshold (including all volumes with a density higher than 175 HU). Top panel artefact configurations with retrospective gating. Bottom panel images with prospective triggering. The 175 HU threshold volume includes all radiopaque parts of the valve such as the leaflets and the prosthetic ring Image noise (median and interquartile range) for retrospective gating and prospective triggering a The P values result from Mann–Whitney U test

Discussion

Our findings demonstrate that axial imaging of PHVs using prospective triggering reduces artefacts with much lower radiation exposure. In our model, prospective triggering reduced artefact volumes for both valves and at all three frequencies with the exception of hypodense artefacts associated with the St Jude valve at 75 BPM. The imaging of PHV with MDCT is hampered by artefacts that depend on valve material and valve geometry [1–3, 10, 12]. Because of the radiopaque components, these artefacts have been interpreted as metal-related artefacts and efforts to improve image quality have included the increase in tube voltage and current [3, 7]. Unfortunately, this approach results in sizable dose exposure (and prohibitive exposure for serial evaluations) and therefore limits the usefulness of CT evaluation of PHV. In addition, retrospectively gated acquisition with reconstruction of 10 intervals (i.e. the full cardiac cycle) has been used to offset the variation of the PHV image quality during the cardiac cycle [2, 3, 7]. Because of the motion of the radiopaque PHV parts, which occurs due to the leaflet excursions and movement of the valve as a whole, reconstructions at each 10% of the ECG interval may generate only two or three image sets that are of diagnostic quality [2, 11, 12]. In vitro, rapid leaflet motion has been shown to increase hyperdense and hypodense artefacts, and rapid motion of the prosthetic ring, which has a higher density than the leaflets, may be expected to further increase artefacts [12]. In this study, we examined how prospective triggering and retrospective gating affected artefact behaviour and noise in PHV imaging. In our in vitro model, we tested the effect of prospective triggering at different frequencies. Axial acquisition reduced image noise only for 60 and 75 BPM but not for 90 BPM. This may be explained by differences in the efficiency of dose used for image reconstruction. For example, axial acquisition with no padding uses the full dose (200 mAs) for image reconstruction [17]. With equal tube voltage setting, the retrospectively gated protocol at 600 mAs with an R-R interval of 1,000 ms (frequency 60/min) and a nominal temporal resolution of 135 ms uses only approximately 80 mAs (=600 mAs × 135/1,000) for the reconstruction of a single ECG interval. At 75 BPM the dose efficiency increases to approximately 100 mAs. At 90 BPM, 120 mAs is used for the reconstruction of a single ECG interval which results in an image noise level comparable to axial acquisition. Axial acquisition reduced hyperdense and hypodense artefacts except hypodense artefacts of the St Jude valve at 75 BPM. The mechanisms responsible for this reduction are not yet fully understood. The important reduction in hyper- and hypodense artefacts for both valves at 60 and 90 BPM are in contrast to the lack of reduction of hypodense artefacts for the St Jude valve at 75 BPM. Possibly, this seems to be related to lower artefact volumes at 75 BPM with retrospective gating. Also, differences between the two valves may play a role. Earlier in vitro work showed that the opened tilting disc valve led to a sharper increase in hyper- and hypodense artefacts when compared to the St Jude valve [12]. Furthermore, the St Jude valve is composed of a nickel alloy prosthetic ring in contrast to the titanium alloy used for the Medtronic Hall. How these differences would cause a lack of reduction of hypodense artifacts at 75 BPM is not clear. Based on the differences between the two acquisition techniques one may postulate that factors associated with retrospective gating such as multisegment reconstructions and helical interpolation may play a role [14-20]. Although increased metal artefacts have been associated with increased noise [14, 15], we found a sharp reduction of hyper- and hypodense artefacts at 90 BPM for both valves despite equal image noise. This suggests that reduction of image noise may not be the primary mechanism responsible for artefacts reduction. However, small variations in leaflet position due to a fluttering motion of the leaflets may cause image discrepancies between consecutive cardiac cycles that increase artefacts with multisegment reconstructions [17, 18]. Furthermore, helical interpolation has been suggested as a mechanism for metal artefacts [16, 18–20]. Other experimental work comparing retrospective gating with prospective triggering for metal objects is scarce. Two reports using in vitro models of coronary in-stent restenosis found improved image quality and a reduction in artefacts [16, 20]. Because of the discrepancy in size and composition of coronary stents, our results are not readily comparable to these reports but similar mechanisms may explain the artefact reduction. There are several limitations to our study. Prospectively triggered acquisition allows reconstruction of only one pre-selected cardiac phase and therefore does not allow dynamic imaging. We used only two mechanical prostheses. Possibly, other prostheses consisting of other metal compounds might yield different results. However, the St Jude prosthesis is the most commonly implanted prosthesis worldwide and is the most likely to be encountered clinically. We used the Medtronic Hall valve because it is the most widely implanted tilting disc valve. Both the Medtronic Hall and St Jude valves have been associated with dysfunction due to tissue ingrowth [2, 8, 21] and for both prostheses CT imaging has been shown to be of additional value for the diagnosis [2, 4, 8]. Also, we used a phantom in which the valve was fixed and only the leaflets moved. Movement of the valve as a whole may induce more artefacts. Furthermore, the use of different reconstruction filters may change the amount of PHV-induced artefacts. Our results suggest a reduction of most PHV-related artefacts at different frequencies that appears independent of the image noise reduction with axial acquisition. Prerequisites, however, are a regular heart rate and the correct timing relative to the cardiac cycle.
  18 in total

1.  Usefulness of computed tomography scanning in the diagnosis of aortic prosthetic valve pannus.

Authors:  Dan Toledano; Christophe Acar
Journal:  J Heart Valve Dis       Date:  2010-09

2.  Evaluation of coronary artery in-stent restenosis with prospectively ECG-triggered axial CT angiography versus retrospective technique: a phantom study.

Authors:  W-J Yang; Z-L Pan; H Zhang; L-F Pang; Y Guo; K-M Chen
Journal:  Radiol Med       Date:  2010-10-06       Impact factor: 3.469

3.  Minimizing clip artifacts in multi CT angiography of clipped patients.

Authors:  I van der Schaaf; M van Leeuwen; A Vlassenbroek; B Velthuis
Journal:  AJNR Am J Neuroradiol       Date:  2006-01       Impact factor: 3.825

4.  Prospective ECG-triggered axial CT at 140-kV tube voltage improves coronary in-stent restenosis visibility at a lower radiation dose compared with conventional retrospective ECG-gated helical CT.

Authors:  Jun Horiguchi; Chikako Fujioka; Masao Kiguchi; Hideya Yamamoto; Toshiro Kitagawa; Shingo Kohno; Katsuhide Ito
Journal:  Eur Radiol       Date:  2009-05-09       Impact factor: 5.315

5.  Evaluation of two iterative techniques for reducing metal artifacts in computed tomography.

Authors:  F Edward Boas; Dominik Fleischmann
Journal:  Radiology       Date:  2011-02-25       Impact factor: 11.105

6.  Multidetector-row computed tomography imaging characteristics of mechanical prosthetic valves.

Authors:  Petr Symersky; Ricardo P J Budde; Mathias Prokop; Bas A J M de Mol
Journal:  J Heart Valve Dis       Date:  2011-03

7.  Artefacts in spiral-CT images and their relation to pitch and subject morphology.

Authors:  J E Wilting; J Timmer
Journal:  Eur Radiol       Date:  1999       Impact factor: 5.315

8.  Artifacts in CT: recognition and avoidance.

Authors:  Julia F Barrett; Nicholas Keat
Journal:  Radiographics       Date:  2004 Nov-Dec       Impact factor: 5.333

9.  Usefulness of a multidetector-row computed tomography scanner for detecting pannus formation.

Authors:  Hideki Teshima; Nobuhiko Hayashida; Shuji Fukunaga; Eiki Tayama; Takemi Kawara; Shigeaki Aoyagi; Masafumi Uchida
Journal:  Ann Thorac Surg       Date:  2004-02       Impact factor: 4.330

10.  Comparison of multidetector-row computed tomography to echocardiography and fluoroscopy for evaluation of patients with mechanical prosthetic valve obstruction.

Authors:  Petr Symersky; Ricardo P J Budde; Bas A J M de Mol; Mathias Prokop
Journal:  Am J Cardiol       Date:  2009-08-28       Impact factor: 2.778

View more
  3 in total

1.  Factors affecting computed tomography image quality for assessment of mechanical aortic valves.

Authors:  Young Joo Suh; Young Jin Kim; Yoo Jin Hong; Hye-Jeong Lee; Jin Hur; Sae Rom Hong; Dong Jin Im; Yun Jung Kim; Byoung Wook Choi
Journal:  Int J Cardiovasc Imaging       Date:  2015-12-01       Impact factor: 2.357

2.  Multidetector-row computed tomography for prosthetic heart valve dysfunction: is concomitant non-invasive coronary angiography possible before redo-surgery?

Authors:  Wilco Tanis; Dominika Suchá; Ward Laufer; Jesse Habets; Lex A van Herwerden; Petr Symersky; Steven Chamuleau; Ricardo P J Budde
Journal:  Eur Radiol       Date:  2014-12-14       Impact factor: 5.315

3.  Advanced CT acquisition protocol with a third-generation dual-source CT scanner and iterative reconstruction technique for comprehensive prosthetic heart valve assessment.

Authors:  Marguerite E Faure; Laurens E Swart; Marcel L Dijkshoorn; Jos A Bekkers; Marcel van Straten; Koen Nieman; Paul M Parizel; Gabriel P Krestin; Ricardo P J Budde
Journal:  Eur Radiol       Date:  2017-12-12       Impact factor: 5.315

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.