Evaluation of two iterative techniques for reducing metal artifacts in computed tomography.

F Edward Boas1, Dominik Fleischmann.   

Abstract

PURPOSE: To evaluate two methods for reducing metal artifacts in computed tomography (CT)--the metal deletion technique (MDT) and the selective algebraic reconstruction technique (SART)--and compare these methods with filtered back projection (FBP) and linear interpolation (LI).
MATERIALS AND METHODS: The institutional review board approved this retrospective HIPAA-compliant study; informed patient consent was waived. Simulated projection data were calculated for a phantom that contained water, soft tissue, bone, and iron. Clinical projection data were obtained retrospectively from 11 consecutively identified CT scans with metal streak artifacts, with a total of 178 sections containing metal. Each scan was reconstructed using FBP, LI, SART, and MDT. The simulated scans were evaluated quantitatively by calculating the average error in Hounsfield units for each pixel compared with the original phantom. Two radiologists who were blinded to the reconstruction algorithms used qualitatively evaluated the clinical scans, ranking the overall severity of artifacts for each algorithm. P values for comparisons of the image quality ranks were calculated from the binomial distribution.
RESULTS: The simulations showed that MDT reduces artifacts due to photon starvation, beam hardening, and motion and does not introduce new streaks between metal and bone. MDT had the lowest average error (76% less than FBP, 42% less than LI, 17% less than SART). Blinded comparison of the clinical scans revealed that MDT had the best image quality 100% of the time (95% confidence interval: 72%, 100%). LI had the second best image quality, and SART and FBP had the worst image quality. On images from two CT scans, as compared with images generated by the scanner, MDT revealed information of potential clinical importance.
CONCLUSION: For a wide range of scans, MDT yields reduced metal streak artifacts and better-quality images than does FBP, LI, or SART. SUPPLEMENTAL MATERIAL: http://radiology.rsna.org/lookup/suppl/doi:10.1148/radiol.11101782/-/DC1. RSNA, 2011

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21357521     DOI: 10.1148/radiol.11101782

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  57 in total

1.  Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software.

Authors:  Young Han Lee; Kwan Kyu Park; Ho-Taek Song; Sungjun Kim; Jin-Suck Suh
Journal:  Eur Radiol       Date:  2012-02-04       Impact factor: 5.315

2.  Factors affecting computed tomography image quality for assessment of mechanical aortic valves.

Authors:  Young Joo Suh; Young Jin Kim; Yoo Jin Hong; Hye-Jeong Lee; Jin Hur; Sae Rom Hong; Dong Jin Im; Yun Jung Kim; Byoung Wook Choi
Journal:  Int J Cardiovasc Imaging       Date:  2015-12-01       Impact factor: 2.357

3.  Quantitative evaluation of beam-hardening artefact correction in dual-energy CT myocardial perfusion imaging.

Authors:  Andreas M Bucher; Julian L Wichmann; U Joseph Schoepf; Christopher D Wolla; Christian Canstein; Andrew D McQuiston; Aleksander W Krazinski; Carlo N De Cecco; Felix G Meinel; Thomas J Vogl; Lucas L Geyer
Journal:  Eur Radiol       Date:  2015-12-09       Impact factor: 5.315

Review 4.  Computed tomography--old ideas and new technology.

Authors:  Dominik Fleischmann; F Edward Boas
Journal:  Eur Radiol       Date:  2011-01-20       Impact factor: 5.315

5.  CT imaging during microwave ablation: analysis of spatial and temporal tissue contraction.

Authors:  Dong Liu; Christopher L Brace
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

6.  Metal artifact reduction using virtual monochromatic images for patients with pedicle screws implants on CT.

Authors:  Yue Dong; Ai Jun Shi; Jian Lin Wu; Ru Xin Wang; Li Fei Sun; Ai Lian Liu; Yi Jun Liu
Journal:  Eur Spine J       Date:  2015-06-13       Impact factor: 3.134

7.  [Postoperative spine].

Authors:  R Schlaeger; J M Lieb; K Shariat; F J Ahlhelm
Journal:  Radiologe       Date:  2014-11       Impact factor: 0.635

8.  Model Image-Based Metal Artifact Reduction for Computed Tomography.

Authors:  Dmytro Luzhbin; Jay Wu
Journal:  J Digit Imaging       Date:  2020-02       Impact factor: 4.056

Review 9.  Imaging-based internal body temperature measurements: The journal Temperature toolbox.

Authors:  Juho Raiko; Kalle Koskensalo; Teija Sainio
Journal:  Temperature (Austin)       Date:  2020-05-29

10.  Clinical evaluation of the normalized metal artefact reduction algorithm caused by dental fillings in CT.

Authors:  X-Y Gong; E Meyer; X-J Yu; J-H Sun; L-P Sheng; K-H Huang; R-Z Wu
Journal:  Dentomaxillofac Radiol       Date:  2013-02-18       Impact factor: 2.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.