Multidetector-row computed tomography imaging characteristics of mechanical prosthetic valves.

Petr Symersky1, Ricardo P J Budde, Mathias Prokop, Bas A J M de Mol.   

Abstract

BACKGROUND AND AIM OF THE STUDY: Electrocardiogram-gated multidetector-row computed tomography (MDCT) imaging may aid in the evaluation of prosthetic valve dysfunction. A pulsatile in vitro model was developed to study the MDCT imaging characteristics of mechanical heart valves (MHVs).
METHODS: Björk-Shiley (BS), St. Jude Medical (SJM), Medtronic-Hall (MH), CarboMedics (CM) and ON-X valves were inserted into an in vitro pulsatile model and scanned using a 64-detector row scanner. The image quality regarding visualization of the leaflets, prosthetic detail and periprosthetic detail, as well as the presence of image artifacts, was scored on a four-point scale.
RESULTS: The image quality for the BS valve was scored poor to moderate for all criteria, and was inferior to that seen with the other valves. Leaflet visualization was excellent for all the other valves. The prosthetic detail was good for MH and SJM valves, and excellent for the CM and ON-X valves. Periprosthetic detail was good for MH, SJM and CM valves, and excellent for ON-X valves. Artifacts were moderate for MH and SJM valves, minor for CM, and minimal for ON-X. All differences were shown to be statistically significant (p < 0.001).
CONCLUSION: The pulsatile in vitro model is an effective tool to detect differences in the MDCT imaging characteristics of MHVs. The image quality is determined by the prosthesis components and, to a lesser degree, by prosthesis design. Modern carbon-titanium MHVs yield good to excellent image quality on MDCT.

Entities:  

Mesh:

Year:  2011        PMID: 21560825

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  9 in total

1.  Factors affecting computed tomography image quality for assessment of mechanical aortic valves.

Authors:  Young Joo Suh; Young Jin Kim; Yoo Jin Hong; Hye-Jeong Lee; Jin Hur; Sae Rom Hong; Dong Jin Im; Yun Jung Kim; Byoung Wook Choi
Journal:  Int J Cardiovasc Imaging       Date:  2015-12-01       Impact factor: 2.357

2.  Assessment of a transcatheter heart valve prosthesis with multidetector computed tomography: in vitro and in vivo imaging characteristics.

Authors:  Linda M de Heer; Jesse Habets; Jolanda Kluin; Pieter R Stella; Willem P Th M Mali; Lex A van Herwerden; Ricardo P J Budde
Journal:  Int J Cardiovasc Imaging       Date:  2012-08-09       Impact factor: 2.357

Review 3.  Diagnostic evaluation of left-sided prosthetic heart valve dysfunction.

Authors:  Jesse Habets; Ricardo P Budde; Petr Symersky; Renee B van den Brink; Bas A de Mol; Willem P Mali; Lex A van Herwerden; Steven A Chamuleau
Journal:  Nat Rev Cardiol       Date:  2011-05-17       Impact factor: 32.419

4.  Artifact reduction strategies for prosthetic heart valve CT imaging.

Authors:  Jesse Habets; Petr Symersky; Tim Leiner; Bas A J M de Mol; Willem P Th M Mali; Ricardo P J Budde
Journal:  Int J Cardiovasc Imaging       Date:  2012-04-05       Impact factor: 2.357

5.  Coronary artery assessment by multidetector computed tomography in patients with prosthetic heart valves.

Authors:  Jesse Habets; Renee B A van den Brink; Ruben Uijlings; Anje M Spijkerboer; Willem P Th M Mali; Steven A J Chamuleau; Ricardo P J Budde
Journal:  Eur Radiol       Date:  2011-12-23       Impact factor: 5.315

6.  A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves.

Authors:  Jesse Habets; Petr Symersky; Bas A J M de Mol; Willem P Th M Mali; Tim Leiner; Ricardo P J Budde
Journal:  Int J Cardiovasc Imaging       Date:  2011-10-15       Impact factor: 2.357

7.  Multidetector CT imaging of mechanical prosthetic heart valves: quantification of artifacts with a pulsatile in-vitro model.

Authors:  Petr Symersky; Ricardo P J Budde; Paul Westers; Bas A J M de Mol; Mathias Prokop
Journal:  Eur Radiol       Date:  2011-05-15       Impact factor: 5.315

8.  Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT.

Authors:  Petr Symersky; Jesse Habets; Paul Westers; Bas A J M de Mol; Mathias Prokop; Ricardo P J Budde
Journal:  Eur Radiol       Date:  2011-12-30       Impact factor: 5.315

9.  Advanced CT acquisition protocol with a third-generation dual-source CT scanner and iterative reconstruction technique for comprehensive prosthetic heart valve assessment.

Authors:  Marguerite E Faure; Laurens E Swart; Marcel L Dijkshoorn; Jos A Bekkers; Marcel van Straten; Koen Nieman; Paul M Parizel; Gabriel P Krestin; Ricardo P J Budde
Journal:  Eur Radiol       Date:  2017-12-12       Impact factor: 5.315

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.