| Literature DB >> 30255659 |
Carlos Ferrer1, Concepción Huertas1, Rodrigo Plaza1, Zulima Aza1, Eva Corredoira1.
Abstract
This work aims to study the effect on surface dose and dose distribution caused by the Elekta Fraxion cranial immobilization system. The effect of Fraxion inclusion in Elekta Monaco treatment planning system and its calculation accuracy is also checked. To study the dose attenuation, a cylindrical phantom was located over the Elekta Fraxion with an IBA CC13 ionization chamber placed in the central insert at the linac isocenter. Dose measurements at multiple gantry angles were performed for three open fields, 10 × 10 cm, 5 × 5 cm and other smaller 2 × 2 cm. Measured doses were compared with the ones calculated by Monaco. Surface dose and dose distribution in the buildup region were measured placing several Gafchromic Films EBT3 at linac CAX between the slabs of a RW3 phantom located over Fraxion and read using FilmQA Pro software. Measures were performed for two open field sizes and results were compared with Monaco calculations. Measurements show a 1% attenuation for 180° gantry angle but it can be as high as 3.4% (5 × 5 open field) for 150°/210° gantry angle, as with these angles the beam goes through the Fraxion's headrest twice. If Fraxion is not included in the calculation Monaco calculation can result in a 3% difference between measured and calculated doses, while with Fraxion in the calculation, the maximum difference is 0.9%. Fraxion increases 3.7 times the surface dose, which can be calculated by Monaco with a difference lower than 2%. Monaco also calculated correctly the PDD for both open fields (2%) when Fraxion is included in the calculation. This work shows that the attenuation varies with gantry angle. The inclusion of Fraxion in Monaco improves the calculation from 3% difference to 1% in the worst case. Furthermore, the surface dose increment and the dose in the buildup region are correctly calculated.Entities:
Keywords: Fraxion; attenuation; buildup; surface dose; treatment planning
Mesh:
Year: 2018 PMID: 30255659 PMCID: PMC6236818 DOI: 10.1002/acm2.12456
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Illustration of the Fraxion's headrest and posts. Reproduced with permission from the vendor.
Figure 2Cylindrical phantom placed over Fraxion's headrest for attenuation measurements.
Figure 3RW3 slabs phantom with radiochromic films placed between to measure the dose distribution.
Transmission/attenuation results
| Angle (°) | Field size 2 × 2 cm2 | Field size 5 × 5 cm2 | Field size 10 × 10 cm2 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| cGy | T (%) | Att (%) | cGy | T (%) | Att (%) | cGy | T (%) | Att (%) | |
| 0 | 73.18 | 100.00 | 0.00 | 81.05 | 100.00 | 0.00 | 88.00 | 100.00 | 0.00 |
| 90 | |||||||||
| 270 | |||||||||
| 135 | 73.15 | 99.96 | 0.05 | 80.99 | 99.92 | 0.08 | 87.99 | 99.99 | 0.01 |
| 140 | 73.17 | 99.98 | 0.02 | 80.98 | 99.91 | 0.09 | 88.01 | 100.01 | ‐0.01 |
| 150 | 70.78 | 96.72 | 3.28 | 78.32 | 96.63 | 3.37 | 85.24 | 96.86 | 3.14 |
| 160 | 71.93 | 98.29 | 1.71 | 79.44 | 98.01 | 1.99 | 86.47 | 98.26 | 1.74 |
| 170 | 72.23 | 98.69 | 1.31 | 79.80 | 98.45 | 1.55 | 86.84 | 98.68 | 1.32 |
| 180 | 72.43 | 98.97 | 1.03 | 80.04 | 98.75 | 1.25 | 87.11 | 98.99 | 1.01 |
Figure 4Percentage transmission for 10 × 10 and 5 × 5 cm2 field size as a function of the gantry angle.
Figure 5Difference between measured and calculated percentage attenuation when Fraxion is included or not in the calculation for the three field sizes.
Figure 6Measured and calculated PDD curves for 10 × 10 and 5 × 5 cm2 field sizes, with and without Fraxion.
Difference between measured and calculated surface dose, and difference when Fraxion is included or not in the calculation
| Field size (cm2) | Fraxion | Surface dose (% of max) | Difference calculated with/without fraxion (%) | ||
|---|---|---|---|---|---|
| Measured (%) | Calculated (%) | Difference measured/calulated (%) | |||
| 10 × 10 | Without fraxion | 20.4 | 22.6 | 2.1 | 49.7 |
| With fraxion | 74.3 | 72.2 | −2.1 | ||
| 5 × 5 | Without fraxion | 17.4 | 21.4 | 4.0 | 44.6 |
| With fraxion | 64.5 | 66.0 | 1.5 | ||
Figure 7Percentage dose variation due to the different thickness of Fraxion headrest at surface and two other depths.
Figure 8Attenuation variation with gantry angle for three field sizes.
Figure 9Treatment differences when Fraxion is not included in the calculation (left) or when it is (right).