| Literature DB >> 22070866 |
Geoffrey G Zhang1, Lichung Ku, Thomas J Dilling, Craig W Stevens, Ray R Zhang, Weiqi Li, Vladimir Feygelman.
Abstract
PURPOSE: Frequently, three-dimensional (3D) conformal beams are used in lung cancer stereotactic body radiotherapy (SBRT). Recently, volumetric modulated arc therapy (VMAT) was introduced as a new treatment modality. VMAT techniques shorten delivery time, reducing the possibility of intrafraction target motion. However dose distributions can be quite different from standard 3D therapy. This study quantifies those differences, with focus on VMAT plans using unflattened photon beams.Entities:
Mesh:
Substances:
Year: 2011 PMID: 22070866 PMCID: PMC3354344 DOI: 10.1186/1748-717X-6-152
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Dosimetric data of 3D, coplanar, non-coplanar VMAT and FFF-VMAT plans.
| Plan | 3D | Coplanar VMAT | Non-coplanar VMAT | Non-coplanar FFF-VMAT |
|---|---|---|---|---|
| CI | 1.44 ± 0.21 | 1.23 ± 0.20 | 1.22 ± 0.17 | 1.19 ± 0.13 |
| CN | 0.67 ± 0.09 | 0.79 ± 0.09 | 0.79 ± 0.08 | 0.81 ± 0.07 |
| GI | 7.12 ± 1.43 | 6.49 ± 1.90 | 6.41 ± 1.78 | 6.23 ± 1.59 |
| V20 (%) | 6.8 ± 2.9 | 5.8 ± 2.4 | 5.8 ± 2.3 | 5.6 ± 2.3 |
| V5 (%) | 23.5 ± 8.0 | 20.9 ± 7.4 | 21.0 ± 7.2 | 20.4 ± 7.1 |
| MLD (cGy) | 497 ± 153 | 458 ± 144 | 459 ± 140 | 449 ± 142 |
| MTD (cGy) | 5359 ± 101 | 5370 ± 116 | 5361 ± 104 | 5324 ± 83 |
| MU | 1528 ± 136 | 1708 ± 194 | 1660 ± 200 | 1805 ± 258 |
In the table, CI = conformity index, CN = conformity number, GI = gradient index, V20 is the percentage volume of lung-GTV exceeding 20 Gy, V5 is the percentage volume of lung-GTV exceeding 5 Gy, MLD = mean lung dose, MTD = mean target dose and MU = number of monitor units.
Compared to 3D plans, number of cases that is dosimetrically in favor of the VMAT or FFF plans.
| Index | CI | CN | GI | V5 | V20 | MLD | MTD | MU |
|---|---|---|---|---|---|---|---|---|
| Coplanar VMAT | 14 | 14 | 12 | 12 | 13 | 15 | 5 | 0 |
| Non-coplanar VMAT | 14 | 14 | 12 | 13 | 14 | 14 | 8 | 2 |
| Non-coplanar FFF-VMAT | 15 | 15 | 13 | 14 | 15 | 14 | 11 | 0 |
The total number of cases is 15. In the table, CI = conformity index, CN = conformity number, GI = gradient index, V20 is the percentage volume of lung-GTV exceeding 20 Gy, V5 is the percentage volume of lung-GTV exceeding 5 Gy, MLD = mean lung dose, MTD = mean target dose and MU = number of monitor units.
Figure 1Due to the limitation that the gantry of a Varian accelerator cannot pass from 180° to -180° and vice versa, the ideal arcs (dashed arc in this example) often have to be shortened to be deliverable (solid arc).