The inter- and intrafraction reproducibilities of three common IMRT delivery techniques.

Courtney R Buckey1, Sotirios Stathakis, Niko Papanikolaou.   

Abstract

PURPOSE: Intensity modulated radiation therapy (IMRT) treatment delivery requires higher precision than conventional 3D treatment delivery because of the sensitivity of the resulting dose to small geometric misalignment of the modulated beamlets. The chosen treatment delivery technique will affect the treatment precision in different ways, based on the characteristics of the delivery method. Delivery using a multileaf collimator (MLC) can reduce treatment time and therapist workload, but typically requires a greater number of monitor units and the fields are prone to both systematic and random leaf positioning errors. An alternative to MLC-based fields, patient specific brass compensators, do not suffer from these leaf positioning errors. In our study, we set out to investigate which delivery method will provide the highest levels of dosimetric reproducibility and the minimum amount of interfraction variability.
METHODS: In our study, a seven field IMRT plan for a head and neck treatment was created using the Pinnacle3 treatment planning system and the intensity maps for each field were obtained. The intensity maps of the fields were delivered with a Varian 2100C/D linear accelerator, using solid compensators and sliding window (SW) and step-and-shoot (SS) MLC segments. Three fields were selected from the seven-beam IMRT plan for comparison. Analysis was carried out using the MatriXX ion chamber array, radiochromic film, and Varian dynalog files.
RESULTS: Our results show that the error in MLC leaf positioning has no gantry angle dependence. The compensator and SW deliveries showed excellent agreement, even when stricter than usual gamma criteria were applied. However, we noted that under these strict conditions, the SS field had at least ten times more pixels out of range than did the compensators. When using step-and-shoot MLC fields, it was observed that the increase in dose rate or the increase of MU/segment degrades the quality of the plan. Analysis of the dynalog files showed that while each individual field had its own propensity for error, all fields showed the same trend: a greater percentage of time the leaves are out of position as dose rate increases, MUs decrease, or both.
CONCLUSIONS: The compensator-based field and both types of MLC-based fields have MatriXX results that are within the clinically acceptable tolerance of 3% dose difference and 2 mm DTA. However, when the criteria are tightened, it becomes evident that the compensators have a definite advantage over their comparable MLC-based competitors in terms of interfraction reproducibility. Fewer monitor units are required to deliver each portal, potentially improving patient outcomes and reducing unwanted side effects to both patients and therapists. In centers without MLC, compensators represent a simple and cost effective way to offer patients state of the art treatment. Based on the results of this study, compensator-based IMRT is a reliable, viable option for use in clinics both with and without MLC-equipped linacs.

Entities:  

Mesh:

Year:  2010        PMID: 20964202     DOI: 10.1118/1.3476413

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

Review 1.  Influence of segment width on plan quality for volumetric modulated arc based stereotactic body radiotherapy.

Authors:  Karthikeyan Nithiyanantham; Ganesh Kadirampatti Mani; Vikraman Subramani; Karrthick Karukkupalayam Palaniappan; Mohanraj Uthiran; Sennniandavar Vellengiri; Sambasivaselli Raju; Sanjay S Supe; Tejinder Kataria
Journal:  Rep Pract Oncol Radiother       Date:  2014-04-16

2.  Volumetric modulated arc planning for lung stereotactic body radiotherapy using conventional and unflattened photon beams: a dosimetric comparison with 3D technique.

Authors:  Geoffrey G Zhang; Lichung Ku; Thomas J Dilling; Craig W Stevens; Ray R Zhang; Weiqi Li; Vladimir Feygelman
Journal:  Radiat Oncol       Date:  2011-11-09       Impact factor: 3.481

3.  Quantitative analysis of brass compensators for commissioning of the Pinnacle planning system for IMRT.

Authors:  Larry L Gates; David J Gladstone
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

4.  Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans.

Authors:  Enrica Seravalli; Antonetta C Houweling; Leo Van Battum; Thom A Raaben; Marc Kuik; Jacco A de Pooter; Marion P R Van Gellekom; Jochem Kaas; Wilfred de Vries; Erik A Loeff; Jeroen B Van de Kamer
Journal:  Phys Imaging Radiat Oncol       Date:  2018-02-02

5.  An Investigation of Multileaf Collimator Performance Dependence on Gantry Angle Using Machine Log Files.

Authors:  Jeremy L Hughes; Martin A Ebert; Conor K McGarry; Christina E Agnew; Mahsheed Sabet; Pejman Rowshanfarzad
Journal:  J Med Phys       Date:  2021-12-31
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.