Controlling the position and the dislocation of the middle ear transducer with high-resolution computed tomography and digital volume tomography: implications for the transducers' design.

Georgios Kontorinis1, Anja M Giesemann, Thomas Witt, Friedrich Goetz, Burkard Schwab.   

Abstract

A minimal tip dislocation of the middle ear transducer (MET(®), Otologics Ltd) may result in poor hearing performance. Our objective was to examine if a defined MET dislocation can be diagnosed by high-resolution computed tomography (HRCT) or digital volume tomography (DVT). A human cadaver head was sequentially implanted with different MET tips (incus application) including a ceramic tip (T 1c), a titanium tip (T 1t), a new, thinner titanium tip (T 2), and a spherical titanium tip (Ts). HRCT and DVT studies were performed. Afterward, the tips were pulled back 0.5 mm, so that they were not attached to the incus. HRCT and DVT scans were repeated to identify the dislocation. Using the best plain in HRCT images, the dislocation of the transducer could be measured reliably and reproducibly in half of the cases. In particular, the precise positioning and the dislocation could be identified when T 1t and Ts were implanted, with the Ts showing the best visibility. DVT failed in recognizing the dislocation in all cases. The identification of MET tip's dislocation with HRCT depends on the shape, size, and material of the tip. This knowledge is useful for the design of the implants, as determination of the right position of the middle ear transducer may be proven important for the hearing outcome. In some cases, however, surgical exploration may still be required. Although DVT represents a promising imaging method for the otologists, it can barely help when MET dislocation is suspected.

Entities:  

Mesh:

Year:  2011        PMID: 21938527     DOI: 10.1007/s00405-011-1772-y

Source DB:  PubMed          Journal:  Eur Arch Otorhinolaryngol        ISSN: 0937-4477            Impact factor:   2.503


  16 in total

1.  CT evaluation of prosthetic ossicular reconstruction procedures: what the otologist needs to know.

Authors:  J A Stone; S K Mukherji; B S Jewett; V N Carrasco; M Castillo
Journal:  Radiographics       Date:  2000 May-Jun       Impact factor: 5.333

2.  The Otologics MET ossicular stimulator.

Authors:  J F Kasic; J M Fredrickson
Journal:  Otolaryngol Clin North Am       Date:  2001-04       Impact factor: 3.346

3.  The value of digital volume tomography in assessing the position of cochlear implant arrays in temporal bone specimens.

Authors:  Thiemo Kurzweg; Carsten V Dalchow; Martin Bremke; Omid Majdani; Ingo Kureck; Reinald Knecht; Jochen A Werner; Afshin Teymoortash
Journal:  Ear Hear       Date:  2010-06       Impact factor: 3.570

4.  Anatomic limitations in implantation of middle ear transducer and carina middle ear implants.

Authors:  Georgios Kontorinis; Thomas Lenarz; Burkard Schwab
Journal:  Laryngoscope       Date:  2010-11       Impact factor: 3.325

5.  The effects of mass loading the ossicles with a floating mass transducer on middle ear transfer function.

Authors:  A J Needham; D Jiang; A Bibas; G Jeronimidis; A Fitzgerald O'Connor
Journal:  Otol Neurotol       Date:  2005-03       Impact factor: 2.311

6.  Fully implantable Otologics MET Carina device for the treatment of sensorineural hearing loss. Preliminary surgical and clinical results.

Authors:  L Bruschini; F Forli; A Santoro; P Bruschini; S Berrettini
Journal:  Acta Otorhinolaryngol Ital       Date:  2009-04       Impact factor: 2.124

7.  Otologics Middle Ear Transducer Ossicular Stimulator: performance results with varying degrees of sensorineural hearing loss.

Authors:  Herman A Jenkins; John K Niparko; William H Slattery; J Gail Neely; John M Fredrickson
Journal:  Acta Otolaryngol       Date:  2004-05       Impact factor: 1.494

8.  U.S. Phase I preliminary results of use of the Otologics MET Fully-Implantable Ossicular Stimulator.

Authors:  Herman A Jenkins; James S Atkins; Drew Horlbeck; Michael E Hoffer; Ben Balough; Joseph V Arigo; George Alexiades; William Garvis
Journal:  Otolaryngol Head Neck Surg       Date:  2007-08       Impact factor: 3.497

9.  Fully implantable hearing device with transducer on the round window as a treatment of mixed hearing loss.

Authors:  Stéphane Tringali; Nick Pergola; Paul Berger; Christian Dubreuil
Journal:  Auris Nasus Larynx       Date:  2008-11-13       Impact factor: 1.863

10.  Otologics fully implantable hearing system: Phase I trial 1-year results.

Authors:  Herman A Jenkins; James S Atkins; Drew Horlbeck; Michael E Hoffer; Ben Balough; George Alexiades; William Garvis
Journal:  Otol Neurotol       Date:  2008-06       Impact factor: 2.311

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.