Otologics Middle Ear Transducer Ossicular Stimulator: performance results with varying degrees of sensorineural hearing loss.

Herman A Jenkins1, John K Niparko, William H Slattery, J Gail Neely, John M Fredrickson.   

Abstract

OBJECTIVES: This study was conducted to demonstrate the safety and efficacy of the Otologics Middle Ear Transducer (MET) Ossicular Stimulator and, in particular, to compare the audiologic benefits of this novel form of electromechanical stimulation with those of conventional acoustical amplification.
MATERIAL AND METHODS: A total of 282 patients were implanted with the device in Europe and the USA. Pure-tone audiometry, speech recognition and subjective assessment of benefit were tested before the surgery and 2, 3, 6 and 12 months afterwards. The US patients were fitted with a digital hearing aid for a minimum of 4 weeks prior to surgery, and the same benefit measures were performed with the digital hearing aid and their "walk-in" hearing aid.
RESULTS: Group mean postoperative bone and air conduction thresholds did not change significantly from preoperative levels. Postoperative air conduction thresholds decreased slightly in some individual patients, due to the mass loading effect exerted by the transducer on the ossicles. Sufficient gain was available to reach target prescription levels for moderate to severely impaired hearing individuals. Speech and subjective assessment of patient preference indicated that patients did as well or better with the MET Ossicular Stimulator than with their "walk-in" aid or the standardized digital aid.
CONCLUSION: The capability of the MET Ossicular Stimulator to provide appropriate gain as a function of degree of hearing loss indicates that the device is a viable treatment for moderate to severe sensorineural hearing loss in adults.

Entities:  

Mesh:

Year:  2004        PMID: 15224860     DOI: 10.1080/00016480410016298

Source DB:  PubMed          Journal:  Acta Otolaryngol        ISSN: 0001-6489            Impact factor:   1.494


  22 in total

1.  Controlling the position and the dislocation of the middle ear transducer with high-resolution computed tomography and digital volume tomography: implications for the transducers' design.

Authors:  Georgios Kontorinis; Anja M Giesemann; Thomas Witt; Friedrich Goetz; Burkard Schwab
Journal:  Eur Arch Otorhinolaryngol       Date:  2011-09-22       Impact factor: 2.503

Review 2.  [Implantable hearing aids].

Authors:  J C Luers; D Beutner; K-B Hüttenbrink
Journal:  HNO       Date:  2011-10       Impact factor: 1.284

Review 3.  [Semi-implantable hearing aids for sensorineural hearing loss and combined hearing loss: experiences at the German Armed Forces Hospital in Ulm].

Authors:  M Tisch; H Maier
Journal:  HNO       Date:  2009-03       Impact factor: 1.284

4.  Round window membrane implantation with an active middle ear implant: a study of the effects on the performance of round window exposure and transducer tip diameter in human cadaveric temporal bones.

Authors:  Stéphane Tringali; Kanthaiah Koka; Arnaud Deveze; N Julian Holland; Herman A Jenkins; Daniel J Tollin
Journal:  Audiol Neurootol       Date:  2010-02-11       Impact factor: 1.854

5.  Minimal Reporting Standards for Active Middle Ear Hearing Implants.

Authors:  Hannes Maier; Uwe Baumann; Wolf-Dieter Baumgartner; Dirk Beutner; Marco D Caversaccio; Thomas Keintzel; Martin Kompis; Thomas Lenarz; Astrid Magele; Torsten Mewes; Alexander Müller; Tobias Rader; Torsten Rahne; Sebastian P Schraven; Burkard Schwab; Georg Mathias Sprinzl; Bernd Strauchmann; Ingo Todt; Thomas Wesarg; Barbara Wollenberg; Stefan K Plontke
Journal:  Audiol Neurootol       Date:  2018-09-07       Impact factor: 1.854

6.  [A fully-implantable active hearing device in congenital auricular atresia].

Authors:  R Siegert; C Neumann
Journal:  HNO       Date:  2014-07       Impact factor: 1.284

7.  Fully implantable Otologics MET Carina device for the treatment of sensorineural hearing loss. Preliminary surgical and clinical results.

Authors:  L Bruschini; F Forli; A Santoro; P Bruschini; S Berrettini
Journal:  Acta Otorhinolaryngol Ital       Date:  2009-04       Impact factor: 2.124

8.  A tri-coil bellows-type round window transducer with improved frequency characteristics for middle-ear implants.

Authors:  Dong Ho Shin; Ki Woong Seong; Sunil Puria; Kyu-Yup Lee; Jin-Ho Cho
Journal:  Hear Res       Date:  2016-09-02       Impact factor: 3.208

Review 9.  Middle ear implantable hearing devices: an overview.

Authors:  David S Haynes; Jadrien A Young; George B Wanna; Michael E Glasscock
Journal:  Trends Amplif       Date:  2009-09

Review 10.  [Differential indication of active middle ear implants].

Authors:  K Braun; H-P Zenner; N Friese; A Tropitzsch
Journal:  HNO       Date:  2015-06       Impact factor: 1.284

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.