Z Li1, T P Mitchell, J R Palta, C Liu.
Abstract
A QA test tool is designed to quantitatively measure the HDR source positioning error, and to facilitate quick and dependable HDR timer linearity test and daily output constancy check. The test tool consists of two concentric disks. The lower disk has a cutout for inserting an HDR catheter, and the upper disk accepts a diode or miniature ionization chamber and can rotate relative to the lower disk. Ionization readings from the source (transferred to the center of the disks) are obtained at two rotational positions of the upper disk which houses the detector. The ratio of the readings is used to determine the source-positioning error of the HDR unit relative to the nominal source position by a simple triangulation principle. Experimental measurements confirm that the QA test tool is sensitive to approximately 0.2 mm variance in source positioning errors. In addition, the QA test tool is suitable for other common HDR QA tests such as the source travel step size test, the daily HDR unit output constancy check, and the timer linearity test. Its simple and robust design permits routine clinical use and provides a high confidence level in the accurate operation of HDR units.Entities:
Mesh:
Year: 1998 PMID: 9507484 DOI: 10.1118/1.598185
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071