Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine.

J Herzen1, M Todorovic, F Cremers, V Platz, D Albers, A Bartels, R Schmidt.   

Abstract

In this paper we present the results of a dosimetric evaluation of a 2D ionization chamber array with the objective of its implementation for quality assurance in clinical routine. The pixel ionization chamber MatriXX (Scanditronix Wellhofer, Germany) consists of 32x32 chambers with a distance of 7.6 mm between chamber centres. The effective depth of measurement under the surface of the detector was determined. The dose and energy dependence, the behaviour of the device during its initial phase and its time stability as well as the lateral response of a single chamber of the detector in cross-plane and diagonal directions were analysed. It could be shown, that the detector's response is linear with dose and energy independent. Taking the lateral response into account, two different dose profiles, for a pyramidal and an IMRT dose distribution, were applied to compare the data generated by a treatment planning system with measurements. From these investigations it can be concluded that the detector is a suitable device for quality assurance and 2D dose verifications.

Entities:  

Mesh:

Year:  2007        PMID: 17264380     DOI: 10.1088/0031-9155/52/4/023

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  35 in total

1.  Simulation for improvement of system sensitivity of radiochromic film dosimetry with different band-pass filters and scanner light intensities.

Authors:  Takeshi Kamomae; Yuki Miyabe; Akira Sawada; Osamu Matoba; Manabu Nakata; Shinsuke Yano; Toru Takakura; Takashi Mizowaki; Akio Itoh; Masahiro Hiraoka
Journal:  Radiol Phys Technol       Date:  2011-03-17

2.  Validation of a quick three-dimensional dose verification system for pre-treatment IMRT QA.

Authors:  Yuji Nakaguchi; Fujio Araki; Takeshi Ono; Yuki Tomiyama; Masato Maruyama; Nozomu Nagasue; Yoshinobu Shimohigashi; Yudai Kai
Journal:  Radiol Phys Technol       Date:  2014-09-27

3.  Reducing Dose Uncertainty for Spot-Scanning Proton Beam Therapy of Moving Tumors by Optimizing the Spot Delivery Sequence.

Authors:  Heng Li; X Ronald Zhu; Xiaodong Zhang
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-06-18       Impact factor: 7.038

4.  Sensitivity of the IQM and MatriXX detectors in megavolt photon beams.

Authors:  Oluwaseyi Michael Oderinde; Freek du Plessis
Journal:  Rep Pract Oncol Radiother       Date:  2019-08-02

5.  Quality assurance of TomoDirect treatment plans using I'mRT MatriXX.

Authors:  Cw Kong; Sk Yu; Ky Cheung; H Geng; Yw Ho; Ww Lam; Wk Wong
Journal:  Biomed Imaging Interv J       Date:  2012-04-01

6.  Implementation and validation of a commercial portal dosimetry software for intensity-modulated radiation therapy pre-treatment verification.

Authors:  C Varatharaj; Eugenia Moretti; M Ravikumar; Maria Rosa Malisan; Sanjay S Supe; Renato Padovani
Journal:  J Med Phys       Date:  2010-10

7.  Dose verification of helical tomotherapy intensity modulated radiation therapy planning using 2D-array ion chambers.

Authors:  S Xu; C Xie; Z Ju; X Dai; H Gong; L Wang; J Yang
Journal:  Biomed Imaging Interv J       Date:  2010-04-01

8.  Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array.

Authors:  Daniela Wagner; Hilke Vorwerk
Journal:  Radiat Oncol       Date:  2011-02-22       Impact factor: 3.481

9.  Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX.

Authors:  A Manikandan; Sarkar Biplab; Perianayagam A David; R Holla; T R Vivek; N Sujatha
Journal:  J Med Phys       Date:  2011-07

10.  A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy.

Authors:  Jaeman Son; Taesung Baek; Boram Lee; Dongho Shin; Sung Yong Park; Jeonghoon Park; Young Kyung Lim; Se Byeong Lee; Jooyoung Kim; Myonggeun Yoon
Journal:  Radiol Oncol       Date:  2015-08-21       Impact factor: 2.991

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.