Physical validation of a Monte Carlo-based, phantom-derived approach to computed tomography organ dosimetry under tube current modulation.

Elliott J Stepusin1, Daniel J Long2, Kayla R Ficarrotta1, David E Hintenlang1, Wesley E Bolch1.   

Abstract

PURPOSE: To physically validate the accuracy of a Monte Carlo-based, phantom-derived methodology for computed tomography (CT) dosimetry that utilizes organ doses from precomputed axial scans and that accounts for tube current modulation (TCM).
METHODS: The output of a Toshiba Aquilion ONE CT scanner was modeled, based on physical measurement, in the Monte Carlo radiation transport code MCNPX (v2.70). CT examinations were taken of two anthropomorphic phantoms representing pediatric and adult patients (15-yr-old female and adult male) at various energies, in which physical organ dose measurements were made using optically stimulated luminescence dosimeters (OSLDs). These exams (chest-abdomen-pelvis) were modeled using organ dose data obtained from the computationally equivalent phantom of each anthropomorphic phantom. TCM was accounted for by weighting all organ dose contributions by both the relative attenuation of the phantom and the image-derived mA value (from the DICOM header) at the same z-extent (cranial-caudal direction) of the axial dose data.
RESULTS: The root mean squares of percent difference in organ dose when comparing the physical organ dose measurements to the computational estimates were 21.2, 12.1, and 15.1% for the uniform (no attenuation weighting), weighted (computationally derived), and image-based methodologies, respectively.
CONCLUSIONS: Overall, these data suggest that the Monte Carlo-based dosimetry presented in this work is viable for CT dosimetry. Additionally, for CT exams with TCM, local attenuation weighting of organ dose contributions from precomputed axial dosimetry libraries increases organ dose accuracy.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  Monte Carlo; anthropomorphic phantom; computed tomography; organ dose; tube current modulation

Mesh:

Year:  2017        PMID: 28688151      PMCID: PMC6343853          DOI: 10.1002/mp.12461

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  24 in total

1.  Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies.

Authors:  M Gies; W A Kalender; H Wolf; C Suess
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

2.  Dose conversion coefficients for CT examinations of adults with automatic tube current modulation.

Authors:  H Schlattl; M Zankl; J Becker; C Hoeschen
Journal:  Phys Med Biol       Date:  2010-09-30       Impact factor: 3.609

3.  Spektr: a computational tool for x-ray spectral analysis and imaging system optimization.

Authors:  J H Siewerdsen; A M Waese; D J Moseley; S Richard; D A Jaffray
Journal:  Med Phys       Date:  2004-11       Impact factor: 4.071

Review 4.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

Review 5.  CT dose reduction and dose management tools: overview of available options.

Authors:  Cynthia H McCollough; Michael R Bruesewitz; James M Kofler
Journal:  Radiographics       Date:  2006 Mar-Apr       Impact factor: 5.333

6.  A method to generate equivalent energy spectra and filtration models based on measurement for multidetector CT Monte Carlo dosimetry simulations.

Authors:  Adam C Turner; Di Zhang; Hyun J Kim; John J DeMarco; Chris H Cagnon; Erin Angel; Dianna D Cody; Donna M Stevens; Andrew N Primak; Cynthia H McCollough; Michael F McNitt-Gray
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

7.  Characteristics of an OSLD in the diagnostic energy range.

Authors:  Rani M Al-Senan; Mustapha R Hatab
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

8.  Characterization of a commercially-available, optically-stimulated luminescent dosimetry system for use in computed tomography.

Authors:  Lindsey Lavoie; Monica Ghita; Libby Brateman; Manuel Arreola
Journal:  Health Phys       Date:  2011-09       Impact factor: 1.316

9.  [CT-expo--a novel program for dose evaluation in CT].

Authors:  G Stamm; H D Nagel
Journal:  Rofo       Date:  2002-12

10.  Construction of anthropomorphic phantoms for use in dosimetry studies.

Authors:  James F Winslow; Daniel E Hyer; Ryan F Fisher; Christopher J Tien; David E Hintenlang
Journal:  J Appl Clin Med Phys       Date:  2009-08-06       Impact factor: 2.102