Practical calibration curve of small-type optically stimulated luminescence (OSL) dosimeter for evaluation of entrance skin dose in the diagnostic X-ray region.

Kazuki Takegami1, Hiroaki Hayashi, Hiroki Okino, Natsumi Kimoto, Itsumi Maehata, Yuki Kanazawa, Tohru Okazaki, Ikuo Kobayashi.   

Abstract

For X-ray diagnosis, the proper management of the entrance skin dose (ESD) is important. Recently, a small-type optically stimulated luminescence dosimeter (nanoDot OSL dosimeter) was made commercially available by Landauer, and it is hoped that it will be used for ESD measurements in clinical settings. Our objectives in the present study were to propose a method for calibrating the ESD measured with the nanoDot OSL dosimeter and to evaluate its accuracy. The reference ESD is assumed to be based on an air kerma with consideration of a well-known back scatter factor. We examined the characteristics of the nanoDot OSL dosimeter using two experimental conditions: a free air irradiation to derive the air kerma, and a phantom experiment to determine the ESD. For evaluation of the ability to measure the ESD, a calibration curve for the nanoDot OSL dosimeter was determined in which the air kerma and/or the ESD measured with an ionization chamber were used as references. As a result, we found that the calibration curve for the air kerma was determined with an accuracy of 5 %. Furthermore, the calibration curve was applied to the ESD estimation. The accuracy of the ESD obtained was estimated to be 15 %. The origin of these uncertainties was examined based on published papers and Monte-Carlo simulation. Most of the uncertainties were caused by the systematic uncertainty of the reading system and the differences in efficiency corresponding to different X-ray energies.

Entities:  

Mesh:

Year:  2015        PMID: 25975450     DOI: 10.1007/s12194-015-0318-1

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  15 in total

1.  Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries.

Authors:  Amy Berrington de González; Sarah Darby
Journal:  Lancet       Date:  2004-01-31       Impact factor: 79.321

2.  Semiempirical model for generating tungsten target x-ray spectra.

Authors:  D M Tucker; G T Barnes; D P Chakraborty
Journal:  Med Phys       Date:  1991 Mar-Apr       Impact factor: 4.071

3.  The energy dependence and dose response of a commercial optically stimulated luminescent detector for kilovoltage photon, megavoltage photon, and electron, proton, and carbon beams.

Authors:  Chester S Reft
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

4.  Angular dependence of the nanoDot OSL dosimeter.

Authors:  James R Kerns; Stephen F Kry; Narayan Sahoo; David S Followill; Geoffrey S Ibbott
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

5.  Characteristics of an OSLD in the diagnostic energy range.

Authors:  Rani M Al-Senan; Mustapha R Hatab
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

6.  Computation of bremsstrahlung X-ray spectra and comparison with spectra measured with a Ge(Li) detector.

Authors:  R Birch; M Marshall
Journal:  Phys Med Biol       Date:  1979-05       Impact factor: 3.609

7.  Backscatter factors for x-rays generated at voltages between 10 and 100 kV.

Authors:  B Grosswendt
Journal:  Phys Med Biol       Date:  1984-05       Impact factor: 3.609

8.  [Dose estimation for exposure conditions of diagnostic radiology acquired by a 2011 questionnaire in a phantom study].

Authors:  Yuta Matsunaga; Ai Kawaguchi; Kenichi Kobayashi; Yasuki Asada; Yukinori Takikawa; Masami Yamada; Shoichi Suzuki
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2013-12

9.  [Patient dose measurement with fluorescent glass dosimeter: characteristics evaluation and patient skin dose measurement in abdominal interventional radiology].

Authors:  Isao Komiya; Takashi Shirasaka; Yoshiyuki Umezu; Masayuki Tachibana; Takashi Izumi
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2004-02

10.  Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements.

Authors:  Paul A Jursinic
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

View more
  7 in total

1.  Energy dependence measurement of small-type optically stimulated luminescence (OSL) dosimeter by means of characteristic X-rays induced with general diagnostic X-ray equipment.

Authors:  Kazuki Takegami; Hiroaki Hayashi; Hiroki Okino; Natsumi Kimoto; Itsumi Maehata; Yuki Kanazawa; Tohru Okazaki; Takuya Hashizume; Ikuo Kobayashi
Journal:  Radiol Phys Technol       Date:  2015-11-20

2.  Practical method for determination of air kerma by use of an ionization chamber toward construction of a secondary X-ray field to be used in clinical examination rooms.

Authors:  Itsumi Maehata; Hiroaki Hayashi; Natsumi Kimoto; Kazuki Takegami; Hiroki Okino; Yuki Kanazawa; Masahide Tominaga
Journal:  Radiol Phys Technol       Date:  2016-03-18

3.  Entrance surface dose measurements using a small OSL dosimeter with a computed tomography scanner having 320 rows of detectors.

Authors:  Kazuki Takegami; Hiroaki Hayashi; Kenji Yamada; Yoshiki Mihara; Natsumi Kimoto; Yuki Kanazawa; Kousaku Higashino; Kazuta Yamashita; Fumio Hayashi; Tohru Okazaki; Takuya Hashizume; Ikuo Kobayashi
Journal:  Radiol Phys Technol       Date:  2016-06-24

4.  Estimation of identification limit for a small-type OSL dosimeter on the medical images by measurement of X-ray spectra.

Authors:  Kazuki Takegami; Hiroaki Hayashi; Hiroki Okino; Natsumi Kimoto; Itsumi Maehata; Yuki Kanazawa; Tohru Okazaki; Takuya Hashizume; Ikuo Kobayashi
Journal:  Radiol Phys Technol       Date:  2016-06-03

5.  Evaluation of surface dose and image quality using the half-scan mode in chest computed tomography-guided interventional radiology: a phantom study.

Authors:  Hiroaki Hasegawa; Jiro Sato; Ikuo Kobayashi
Journal:  Radiol Phys Technol       Date:  2018-02-27

6.  Energy dependence and angular dependence of an optically stimulated luminescence dosimeter in the mammography energy range.

Authors:  Ai Kawaguchi; Yuta Matsunaga; Shoichi Suzuki; Koichi Chida
Journal:  J Appl Clin Med Phys       Date:  2017-01-24       Impact factor: 2.102

7.  Direct measurement of radiation exposure dose to individual organs during diagnostic computed tomography examination.

Authors:  Kazuta Yamashita; Kosaku Higashino; Hiroaki Hayashi; Kazuki Takegami; Fumio Hayashi; Yoshihiro Tsuruo; Koichi Sairyo
Journal:  Sci Rep       Date:  2021-03-08       Impact factor: 4.379

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.