Effects of MR surface coils on PET quantification.

Lawrence R MacDonald1, Steve Kohlmyer, Chi Liu, Thomas K Lewellen, Paul E Kinahan.   

Abstract

PURPOSE: The goal of this work was to investigate the effects of MRI surface coils on attenuation-corrected PET emission data. The authors studied the cases where either an MRI or a CT scan would be used to provide PET attenuation correction (AC). Combined MR/PET scanners that use the MRI for PET AC (MR-AC) face the challenge of absent surface coils in MR images and thus cannot directly account for attenuation in the coils. Combining MR and PET images could be achieved by transporting the subject on a stereotactically registered table between independent MRI and PET scanners. In this case, conventional PET CT-AC methods could be used. A challenge here is that high atomic number materials within MR coils cause artifacts in CT images and CT based AC is typically not validated for coil materials.
METHODS: The authors evaluated PET artifacts when MR coils were absent from AC data (MR-AC), or when coil attenuation was measured by CT scanning (CT-AC). They scanned PET phantoms with MR surface coils on a clinical PET/CT system and used CT-AC to reconstruct PET data. The authors then omitted the coil from the CT-AC image to mimic the MR-AC scenario. Images were acquired using cylinder and anthropomorphic phantoms. They evaluated and compared the following five scenarios: (1) A uniform cylinder phantom and head coil scanned and reconstructed using CT-AC; (2) similar emission data (with head coil present) were reconstructed without the head coil in the AC data; (3) the same cylinder scanned without the head coil present (reference scan); (4) a PET torso phantom with a full MR torso coil present in both PET and CT; (5) only half of the separable torso coil present in the PET/CT acquisition. The authors also performed analytic simulations of the first three scenarios.
RESULTS: Streak artifacts were present in CT images containing MR surface coils due to metal components. These artifacts persisted after the CT images were converted for PET AC. The artifacts were significantly reduced when half of the separable coil was removed during the scan. CT scans tended to over-estimate the linear attenuation coefficient (micro) of the metal components when using conventional methods for converting from CT number to micro(511 keV). Artifacts were visible outside the phantom in some of the PET emission images, corresponding to the MRI coil geometry. However, only subtle artifacts were apparent in the emission images inside the phantoms. On the other hand, the PET emission image quantitative accuracy was significantly affected: the activity was underestimated by 19% when AC did not include the head coil, and overestimated by 28% when the CT-AC included the head coil.
CONCLUSIONS: The presence of MR coils during PET or PET/CT scanning can cause subtle artifacts and potentially important quantification errors. Alternative CT techniques that mitigate artifacts should be used to improve AC accuracy. When possible, removing segments of an MR coil prior to the PET/CT exam is recommended. Further, MR coils could be redesigned to reduce artifacts by rearranging placement of the most attenuating materials.

Entities:  

Mesh:

Year:  2011        PMID: 21815368      PMCID: PMC3117895          DOI: 10.1118/1.3583697

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  Cranial thickness in relation to age, sex and general body build in a Danish forensic sample.

Authors:  N Lynnerup
Journal:  Forensic Sci Int       Date:  2001-03-01       Impact factor: 2.395

2.  An analytic study of the effects of attenuation on tumor detection in whole-body PET oncology imaging.

Authors:  Chuanyong Bai; Paul E Kinahan; David Brasse; Claude Comtat; David W Townsend; Carolyn C Meltzer; Victor Villemagne; Martin Charron; Michel Defrise
Journal:  J Nucl Med       Date:  2003-11       Impact factor: 10.057

3.  Normalized metal artifact reduction (NMAR) in computed tomography.

Authors:  Esther Meyer; Rainer Raupach; Michael Lell; Bernhard Schmidt; Marc Kachelriess
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

4.  Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI.

Authors:  Bernd J Pichler; Martin S Judenhofer; Ciprian Catana; Jeffrey H Walton; Manfred Kneilling; Robert E Nutt; Stefan B Siegel; Claus D Claussen; Simon R Cherry
Journal:  J Nucl Med       Date:  2006-04       Impact factor: 10.057

5.  Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner.

Authors:  Ciprian Catana; Yibao Wu; Martin S Judenhofer; Jinyi Qi; Bernd J Pichler; Simon R Cherry
Journal:  J Nucl Med       Date:  2006-12       Impact factor: 10.057

6.  PET and NMR dual acquisition (PANDA): applications to isolated, perfused rat hearts.

Authors:  P B Garlick; P K Marsden; A C Cave; H G Parkes; R Slates; Y Shao; R W Silverman; S R Cherry
Journal:  NMR Biomed       Date:  1997-05       Impact factor: 4.044

7.  Simultaneous MRI and PET imaging of a rat brain.

Authors:  Raymond R Raylman; Stan Majewski; Susan K Lemieux; S Sendhil Velan; Brian Kross; Vladimir Popov; Mark F Smith; Andrew G Weisenberger; Carl Zorn; Gary D Marano
Journal:  Phys Med Biol       Date:  2006-11-23       Impact factor: 3.609

8.  Simultaneous PET and MR imaging.

Authors:  Y Shao; S R Cherry; K Farahani; K Meadors; S Siegel; R W Silverman; P K Marsden
Journal:  Phys Med Biol       Date:  1997-10       Impact factor: 3.609

9.  Do implanted pacemaker leads and ICD leads cause metal-related artifact in cardiac PET/CT?

Authors:  Frank P DiFilippo; Richard C Brunken
Journal:  J Nucl Med       Date:  2005-03       Impact factor: 10.057

10.  Thickness of the human cranial diploe in relation to age, sex and general body build.

Authors:  Niels Lynnerup; Jacob G Astrup; Birgitte Sejrsen
Journal:  Head Face Med       Date:  2005-12-20       Impact factor: 2.151

View more
  21 in total

Review 1.  Characterization of the impact to PET quantification and image quality of an anterior array surface coil for PET/MR imaging.

Authors:  Scott D Wollenweber; Gaspar Delso; Timothy Deller; David Goldhaber; Martin Hüllner; Patrick Veit-Haibach
Journal:  MAGMA       Date:  2013-06-26       Impact factor: 2.310

2.  Comparison of MR-based attenuation correction and CT-based attenuation correction of whole-body PET/MR imaging.

Authors:  David Izquierdo-Garcia; Stephen J Sawiak; Karin Knesaurek; Jagat Narula; Valentin Fuster; Joseph Machac; Zahi A Fayad
Journal:  Eur J Nucl Med Mol Imaging       Date:  2014-03-21       Impact factor: 9.236

3.  Quantitative carotid PET/MR imaging: clinical evaluation of MR-Attenuation correction versus CT-Attenuation correction in (18)F-FDG PET/MR emission data and comparison to PET/CT.

Authors:  Jason Bini; Philip M Robson; Claudia Calcagno; Mootaz Eldib; Zahi A Fayad
Journal:  Am J Nucl Med Mol Imaging       Date:  2015-02-15

4.  Quantitative and qualitative evaluation of sequential PET/MRI using a newly developed mobile PET system for brain imaging.

Authors:  Mizue Suzuki; Yasutaka Fushimi; Tomohisa Okada; Takuya Hinoda; Ryusuke Nakamoto; Yoshiki Arakawa; Nobukatsu Sawamoto; Kaori Togashi; Yuji Nakamoto
Journal:  Jpn J Radiol       Date:  2021-02-28       Impact factor: 2.374

5.  Comparison of lesion detection and quantitation of tracer uptake between PET from a simultaneously acquiring whole-body PET/MR hybrid scanner and PET from PET/CT.

Authors:  Marco Wiesmüller; Harald H Quick; Bharath Navalpakkam; Michael M Lell; Michael Uder; Philipp Ritt; Daniela Schmidt; Michael Beck; Torsten Kuwert; Carl C von Gall
Journal:  Eur J Nucl Med Mol Imaging       Date:  2012-10-06       Impact factor: 9.236

Review 6.  Potential Clinical Applications of 18F-Fluorodeoxyglucose Positron Emission Tomography/Magnetic Resonance Mammography in Breast Cancer.

Authors:  Ihn-Ho Cho; Eun-Jung Kong
Journal:  Nucl Med Mol Imaging       Date:  2016-08-30

7.  A 31-channel MR brain array coil compatible with positron emission tomography.

Authors:  Christin Y Sander; Boris Keil; Daniel B Chonde; Bruce R Rosen; Ciprian Catana; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2014-07-07       Impact factor: 4.668

8.  Markerless attenuation correction for carotid MRI surface receiver coils in combined PET/MR imaging.

Authors:  Mootaz Eldib; Jason Bini; Philip M Robson; Claudia Calcagno; David D Faul; Charalampos Tsoumpas; Zahi A Fayad
Journal:  Phys Med Biol       Date:  2015-05-28       Impact factor: 3.609

Review 9.  Attenuation Correction for Magnetic Resonance Coils in Combined PET/MR Imaging: A Review.

Authors:  Mootaz Eldib; Jason Bini; David D Faul; Niels Oesingmann; Charalampos Tsoumpas; Zahi A Fayad
Journal:  PET Clin       Date:  2015-11-27

Review 10.  MR Imaging-Guided Attenuation Correction of PET Data in PET/MR Imaging.

Authors:  David Izquierdo-Garcia; Ciprian Catana
Journal:  PET Clin       Date:  2016-01-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.