Markerless attenuation correction for carotid MRI surface receiver coils in combined PET/MR imaging.

Mootaz Eldib1, Jason Bini, Philip M Robson, Claudia Calcagno, David D Faul, Charalampos Tsoumpas, Zahi A Fayad.   

Abstract

The purpose of the study was to evaluate the effect of attenuation of MR coils on quantitative carotid PET/MR exams. Additionally, an automated attenuation correction method for flexible carotid MR coils was developed and evaluated. The attenuation of the carotid coil was measured by imaging a uniform water phantom injected with 37 MBq of 18F-FDG in a combined PET/MR scanner for 24 min with and without the coil. In the same session, an ultra-short echo time (UTE) image of the coil on top of the phantom was acquired. Using a combination of rigid and non-rigid registration, a CT-based attenuation map was registered to the UTE image of the coil for attenuation and scatter correction. After phantom validation, the effect of the carotid coil attenuation and the attenuation correction method were evaluated in five subjects. Phantom studies indicated that the overall loss of PET counts due to the coil was 6.3% with local region-of-interest (ROI) errors reaching up to 18.8%. Our registration method to correct for attenuation from the coil decreased the global error and local error (ROI) to 0.8% and 3.8%, respectively. The proposed registration method accurately captured the location and shape of the coil with a maximum spatial error of 2.6 mm. Quantitative analysis in human studies correlated with the phantom findings, but was dependent on the size of the ROI used in the analysis. MR coils result in significant error in PET quantification and thus attenuation correction is needed. The proposed strategy provides an operator-free method for attenuation and scatter correction for a flexible MRI carotid surface coil for routine clinical use.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26020273      PMCID: PMC4495953          DOI: 10.1088/0031-9155/60/12/4705

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  25 in total

1.  Evaluation of the attenuation properties of MR equipment for its use in a whole-body PET/MR scanner.

Authors:  G Delso; A Martinez-Möller; R A Bundschuh; R Ladebeck; Y Candidus; D Faul; S I Ziegler
Journal:  Phys Med Biol       Date:  2010-07-20       Impact factor: 3.609

2.  Method for transforming CT images for attenuation correction in PET/CT imaging.

Authors:  Jonathan P J Carney; David W Townsend; Vitaliy Rappoport; Bernard Bendriem
Journal:  Med Phys       Date:  2006-04       Impact factor: 4.071

3.  Diffeomorphic demons: efficient non-parametric image registration.

Authors:  Tom Vercauteren; Xavier Pennec; Aymeric Perchant; Nicholas Ayache
Journal:  Neuroimage       Date:  2008-11-07       Impact factor: 6.556

4.  Three-dimensional ultrashort echo time imaging of solid polymers on a 3-Tesla whole-body MRI scanner.

Authors:  Fabian Springer; Petros Martirosian; Nina F Schwenzer; Michael Szimtenings; Peter Kreisler; Claus D Claussen; Fritz Schick
Journal:  Invest Radiol       Date:  2008-11       Impact factor: 6.016

5.  Image matching as a diffusion process: an analogy with Maxwell's demons.

Authors:  J P Thirion
Journal:  Med Image Anal       Date:  1998-09       Impact factor: 8.545

6.  Integrated PET/MR imaging: automatic attenuation correction of flexible RF coils.

Authors:  René Kartmann; Daniel H Paulus; Harald Braun; Bassim Aklan; Susanne Ziegler; Bharath K Navalpakkam; Markus Lentschig; Harald H Quick
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

7.  Attenuation correction for flexible magnetic resonance coils in combined magnetic resonance/positron emission tomography imaging.

Authors:  Mootaz Eldib; Jason Bini; Claudia Calcagno; Philip M Robson; Venkatesh Mani; Zahi A Fayad
Journal:  Invest Radiol       Date:  2014-02       Impact factor: 6.016

8.  Performance measurements of the Siemens mMR integrated whole-body PET/MR scanner.

Authors:  Gaspar Delso; Sebastian Fürst; Björn Jakoby; Ralf Ladebeck; Carl Ganter; Stephan G Nekolla; Markus Schwaiger; Sibylle I Ziegler
Journal:  J Nucl Med       Date:  2011-11-11       Impact factor: 10.057

9.  Toward simultaneous PET/MR breast imaging: systematic evaluation and integration of a radiofrequency breast coil.

Authors:  Bassim Aklan; Daniel H Paulus; Evelyn Wenkel; Harald Braun; Bharath K Navalpakkam; Susanne Ziegler; Christian Geppert; Eric E Sigmund; Amy Melsaether; Harald H Quick
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

10.  Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data.

Authors:  Axel Martinez-Möller; Michael Souvatzoglou; Gaspar Delso; Ralph A Bundschuh; Christophe Chefd'hotel; Sibylle I Ziegler; Nassir Navab; Markus Schwaiger; Stephan G Nekolla
Journal:  J Nucl Med       Date:  2009-03-16       Impact factor: 10.057

View more
  9 in total

1.  High-risk plaque features can be detected in non-stenotic carotid plaques of patients with ischaemic stroke classified as cryptogenic using combined (18)F-FDG PET/MR imaging.

Authors:  Fabien Hyafil; Andreas Schindler; Dominik Sepp; Tilman Obenhuber; Anna Bayer-Karpinska; Tobias Boeckh-Behrens; Sabine Höhn; Marcus Hacker; Stephan G Nekolla; Axel Rominger; Martin Dichgans; Markus Schwaiger; Tobias Saam; Holger Poppert
Journal:  Eur J Nucl Med Mol Imaging       Date:  2015-10-03       Impact factor: 9.236

Review 2.  Attenuation Correction for Magnetic Resonance Coils in Combined PET/MR Imaging: A Review.

Authors:  Mootaz Eldib; Jason Bini; David D Faul; Niels Oesingmann; Charalampos Tsoumpas; Zahi A Fayad
Journal:  PET Clin       Date:  2015-11-27

Review 3.  Emerging role of integrated PET-MRI in osteoarthritis.

Authors:  Amarnath Jena; Sangeeta Taneja; Prerana Rana; Nidhi Goyal; Abhishek Vaish; Rajesh Botchu; Raju Vaishya
Journal:  Skeletal Radiol       Date:  2021-06-29       Impact factor: 2.199

Review 4.  Applications of PET-MRI in musculoskeletal disease.

Authors:  Feliks Kogan; Stephen M Broski; Daehyun Yoon; Garry E Gold
Journal:  J Magn Reson Imaging       Date:  2018-07       Impact factor: 4.813

Review 5.  Synergistic motion compensation strategies for positron emission tomography when acquired simultaneously with magnetic resonance imaging.

Authors:  Irene Polycarpou; Georgios Soultanidis; Charalampos Tsoumpas
Journal:  Philos Trans A Math Phys Eng Sci       Date:  2021-07-05       Impact factor: 4.226

6.  Evaluation of a Dedicated Radiofrequency Carotid PET/MRI Coil.

Authors:  Mueez Aizaz; Jochem A J van der Pol; Roel Wierts; Hans Zwart; Abe J van der Werf; Joachim E Wildberger; Jan A Bucerius; Rik P M Moonen; Marianne Eline Kooi
Journal:  J Clin Med       Date:  2022-05-04       Impact factor: 4.241

7.  MLAA-based attenuation correction of flexible hardware components in hybrid PET/MR imaging.

Authors:  Thorsten Heußer; Christopher M Rank; Yannick Berker; Martin T Freitag; Marc Kachelrieß
Journal:  EJNMMI Phys       Date:  2017-03-01

8.  Evaluation of 511 keV photon attenuation by a novel 32-channel phased array prospectively designed for cardiovascular hybrid PET/MRI imaging.

Authors:  Adam Farag; R Terry Thompson; Jonathan D Thiessen; Heather Biernaski; Frank S Prato; Jean Théberge
Journal:  Eur J Hybrid Imaging       Date:  2020-05-12

9.  Assessment of a novel 32-channel phased array for cardiovascular hybrid PET/MRI imaging: MRI performance.

Authors:  Adam Farag; R Terry Thompson; Jonathan D Thiessen; John Butler; Frank S Prato; Jean Théberge
Journal:  Eur J Hybrid Imaging       Date:  2019-08-22
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.