Literature DB >> 21614315

Optimal sensitometric curves of Kodak EDR2 film for dynamic intensity modulated radiation therapy verification.

S Suriyapee1, N Pitaxtarnin, S Oonsiri, C Jumpangern, I Israngkul Na Ayuthaya.   

Abstract

PURPOSE: To investigate the optimal sensitometric curves of extended dose range (EDR2) radiographic film in terms of depth, field size, dose range and processing conditions for dynamic intensity modulated radiation therapy (IMRT) dosimetry verification with 6 MV X-ray beams.
MATERIALS AND METHODS: A Varian Clinac 23 EX linear accelerator with 6 MV X-ray beam was used to study the response of Kodak EDR2 film. Measurements were performed at depths of 5, 10 and 15 cm in MedTec virtual water phantom and with field sizes of 2x2, 3x3, 10x10 and 15x15 cm(2). Doses ranging from 20 to 450 cGy were used. The film was developed with the Kodak RP X-OMAT Model M6B automatic film processor. Film response was measured with the Vidar model VXR-16 scanner. Sensitometric curves were applied to the dose profiles measured with film at 5 cm in the virtual water phantom with field sizes of 2x2 and 10x10 cm(2) and compared with ion chamber data. Scanditronix/Wellhofer OmniPro(TM) IMRT software was used for the evaluation of the IMRT plan calculated by Eclipse treatment planning.
RESULTS: Investigation of the reproducibility and accuracy of the film responses, which depend mainly on the film processor, was carried out by irradiating one film nine times with doses of 20 to 450 cGy. A maximum standard deviation of 4.9% was found which decreased to 1.9% for doses between 20 and 200 cGy. The sensitometric curves for various field sizes at fixed depth showed a maximum difference of 4.2% between 2x2 and 15x15 cm(2) at 5 cm depth with a dose of 450 cGy. The shallow depth tended to show a greater effect of field size responses than the deeper depths. The sensitometric curves for various depths at fixed field size showed slightly different film responses; the difference due to depth was within 1.8% for all field sizes studied. Both field size and depth effect were reduced when the doses were lower than 450 cGy. The difference was within 2.5% in the dose range from 20 to 300 cGy for all field sizes and depths studied. Dose profiles measured with EDR2 film were consistent with those measured with an ion chamber. The optimal sensitometric curve was acquired by irradiating film at a depth of 5 cm with doses ranging from 20 to 450 cGy with a 3×3 cm(2) multileaf collimator. The optimal sensitometric curve allowed accurate determination of the absolute dose distribution. In almost 200 cases of dynamic IMRT plan verification with EDR2 film, the difference between measured and calculated dose was generally less than 3% and with 3 mm distance to agreement when using gamma value verification.
CONCLUSION: EDR2 film can be used for accurate verification of composite isodose distributions of dynamic IMRT when the optimal sensitometric curve has been established.

Entities:  

Keywords:  EDR2 film; IMRT treatment plan verification; sensitometric curves

Year:  2008        PMID: 21614315      PMCID: PMC3097693          DOI: 10.2349/biij.4.1.e4

Source DB:  PubMed          Journal:  Biomed Imaging Interv J        ISSN: 1823-5530


  7 in total

1.  Investigation of Kodak extended dose range (EDR) film for megavoltage photon beam dosimetry.

Authors:  Indrin J Chetty; Paule M Charland
Journal:  Phys Med Biol       Date:  2002-10-21       Impact factor: 3.609

2.  Dosimetry of therapeutic photon beams using an extended dose range film.

Authors:  Jacqueline Esthappan; Sasa Mutic; William B Harms; James F Dempsey; Daniel A Low
Journal:  Med Phys       Date:  2002-10       Impact factor: 4.071

3.  A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery.

Authors:  Paul A Jursinic; Ben E Nelms
Journal:  Med Phys       Date:  2003-05       Impact factor: 4.071

4.  Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a static multileaf collimator.

Authors:  X R Zhu; P A Jursinic; D F Grimm; F Lopez; J J Rownd; M T Gillin
Journal:  Med Phys       Date:  2002-08       Impact factor: 4.071

5.  A quantitative evaluation of IMRT dose distributions: refinement and clinical assessment of the gamma evaluation.

Authors:  Tom Depuydt; Ann Van Esch; Dominique Pierre Huyskens
Journal:  Radiother Oncol       Date:  2002-03       Impact factor: 6.280

6.  Dosimetric performance of an enhanced dose range radiographic film for intensity-modulated radiation therapy quality assurance.

Authors:  Arthur J Olch
Journal:  Med Phys       Date:  2002-09       Impact factor: 4.071

7.  Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification.

Authors:  A Sankar; Komanduri M Ayyangar; R Mothilal Nehru; P G Gopalakrishna Kurup; V Murali; Charles A Enke; J Velmurugan
Journal:  Med Dosim       Date:  2006       Impact factor: 1.482

  7 in total
  1 in total

1.  Rounded leaf end modeling in Pinnacle VMAT treatment planning for fixed jaw linacs.

Authors:  Lori A Young; Fei Yang; Ning Cao; Juergen Meyer
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.