Material separation in x-ray CT with energy resolved photon-counting detectors.

Xiaolan Wang1, Dirk Meier, Katsuyuki Taguchi, Douglas J Wagenaar, Bradley E Patt, Eric C Frey.   

Abstract

PURPOSE: The objective of the study was to demonstrate that, in x-ray computed tomography (CT), more than two types of materials can be effectively separated with the use of an energy resolved photon-counting detector and classification methodology. Specifically, this applies to the case when contrast agents that contain K-absorption edges in the energy range of interest are present in the object. This separation is enabled via the use of recently developed energy resolved photon-counting detectors with multiple thresholds, which allow simultaneous measurements of the x-ray attenuation at multiple energies.
METHODS: To demonstrate this capability, we performed simulations and physical experiments using a six-threshold energy resolved photon-counting detector. We imaged mouse-sized cylindrical phantoms filled with several soft-tissue-like and bone-like materials and with iodine-based and gadolinium-based contrast agents. The linear attenuation coefficients were reconstructed for each material in each energy window and were visualized as scatter plots between pairs of energy windows. For comparison, a dual-kVp CT was also simulated using the same phantom materials. In this case, the linear attenuation coefficients at the lower kVp were plotted against those at the higher kVp.
RESULTS: In both the simulations and the physical experiments, the contrast agents were easily separable from other soft-tissue-like and bone-like materials, thanks to the availability of the attenuation coefficient measurements at more than two energies provided by the energy resolved photon-counting detector. In the simulations, the amount of separation was observed to be proportional to the concentration of the contrast agents; however, this was not observed in the physical experiments due to limitations of the real detector system. We used the angle between pairs of attenuation coefficient vectors in either the 5-D space (for non-contrast-agent materials using energy resolved photon-counting acquisition) or a 2-D space (for contrast agents using energy resolved photon-counting acquisition and all materials using dual-kVp acquisition) as a measure of the degree of separation. Compared to dual-kVp techniques, an energy resolved detector provided a larger separation and the ability to separate different target materials using measurements acquired in different energy window pairs with a single x-ray exposure.
CONCLUSIONS: We concluded that x-ray CT with an energy resolved photon-counting detector with more than two energy windows allows the separation of more than two types of materials, e.g., soft-tissue-like, bone-like, and one or more materials with K-edges in the energy range of interest. Separating material types using energy resolved photon-counting detectors has a number of advantages over dual-kVp CT in terms of the degree of separation and the number of materials that can be separated simultaneously.

Entities:  

Mesh:

Year:  2011        PMID: 21520865      PMCID: PMC3060934          DOI: 10.1118/1.3553401

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  25 in total

1.  Depth-segmented detector for x-ray absorptiometry.

Authors:  G M Stevens; N J Pelc
Journal:  Med Phys       Date:  2000-05       Impact factor: 4.071

2.  An analytical model of the effects of pulse pileup on the energy spectrum recorded by energy resolved photon counting x-ray detectors.

Authors:  Katsuyuki Taguchi; Eric C Frey; Xiaolan Wang; Jan S Iwanczyk; William C Barber
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

3.  Energy-selective reconstructions in X-ray computerized tomography.

Authors:  R E Alvarez; A Macovski
Journal:  Phys Med Biol       Date:  1976-09       Impact factor: 3.609

4.  Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies.

Authors:  W A Kalender; W H Perman; J R Vetter; E Klotz
Journal:  Med Phys       Date:  1986 May-Jun       Impact factor: 4.071

5.  An implementation of dual energy CT scanning.

Authors:  W Marshall; E Hall; A Doost-Hoseini; R Alvarez; A Macovski; D Cassel
Journal:  J Comput Assist Tomogr       Date:  1984-08       Impact factor: 1.826

6.  Dual energy film subtraction technique for detecting calcification in solitary pulmonary nodules.

Authors:  R A Kruger; J D Armstrong; J A Sorenson; L T Niklason
Journal:  Radiology       Date:  1981-07       Impact factor: 11.105

7.  Generalized image combinations in dual KVP digital radiography.

Authors:  L A Lehmann; R E Alvarez; A Macovski; W R Brody; N J Pelc; S J Riederer; A L Hall
Journal:  Med Phys       Date:  1981 Sep-Oct       Impact factor: 4.071

8.  Initial results with prereconstruction dual-energy computed tomography (PREDECT).

Authors:  W H Marshall; R E Alvarez; A Macovski
Journal:  Radiology       Date:  1981-08       Impact factor: 11.105

9.  Experimental dual xenon detectors for quantitative CT and spectral artifact correction.

Authors:  D J Drost; A Fenster
Journal:  Med Phys       Date:  1980 Mar-Apr       Impact factor: 4.071

10.  Split-detector computed tomography: a preliminary report.

Authors:  R A Brooks; G Di Chiro
Journal:  Radiology       Date:  1978-01       Impact factor: 11.105

View more
  34 in total

1.  Energy response calibration of photon-counting detectors using x-ray fluorescence: a feasibility study.

Authors:  H-M Cho; H Ding; B P Ziemer; S Molloi
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

2.  Quantitative percussion diagnostics and bone density analysis of the implant-bone interface in a pre- and postmortem human subject.

Authors:  Cherilyn G Sheets; Dee Dee Hui; Vaibhav Bajaj; James C Earthman
Journal:  Int J Oral Maxillofac Implants       Date:  2013 Nov-Dec       Impact factor: 2.804

3.  Dual-source multienergy CT with triple or quadruple x-ray beams.

Authors:  Lifeng Yu; Liqiang Ren; Zhoubo Li; Shuai Leng; Cynthia H McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2018-07-24

4.  Quantitative material characterization from multi-energy photon counting CT.

Authors:  Adam M Alessio; Lawrence R MacDonald
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

5.  Spectral CT using multiple balanced K-edge filters.

Authors:  Yothin Rakvongthai; William Worstell; Georges El Fakhri; Junguo Bian; Auranuch Lorsakul; Jinsong Ouyang
Journal:  IEEE Trans Med Imaging       Date:  2014-09-19       Impact factor: 10.048

6.  Breast composition measurement with a cadmium-zinc-telluride based spectral computed tomography system.

Authors:  Huanjun Ding; Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

7.  Cascaded systems analysis of photon counting detectors.

Authors:  J Xu; W Zbijewski; G Gang; J W Stayman; K Taguchi; M Lundqvist; E Fredenberg; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

8.  Dual-contrast agent photon-counting computed tomography of the heart: initial experience.

Authors:  Rolf Symons; Tyler E Cork; Manu N Lakshmanan; Robert Evers; Cynthia Davies-Venn; Kelly A Rice; Marvin L Thomas; Chia-Ying Liu; Steffen Kappler; Stefan Ulzheimer; Veit Sandfort; David A Bluemke; Amir Pourmorteza
Journal:  Int J Cardiovasc Imaging       Date:  2017-03-13       Impact factor: 2.357

9.  Volumetric CT with sparse detector arrays (and application to Si-strip photon counters).

Authors:  A Sisniega; W Zbijewski; J W Stayman; J Xu; K Taguchi; E Fredenberg; Mats Lundqvist; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-11-27       Impact factor: 3.609

10.  Image-based Material Decomposition with a General Volume Constraint for Photon-Counting CT.

Authors:  Zhoubo Li; Shuai Leng; Lifeng Yu; Zhicong Yu; Cynthia H McCollough
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.