Radiation doses in cone-beam breast computed tomography: a Monte Carlo simulation study.

Ying Yi1, Chao-Jen Lai, Tao Han, Yuncheng Zhong, Youtao Shen, Xinming Liu, Shuaiping Ge, Zhicheng You, Tianpeng Wang, Chris C Shaw.   

Abstract

PURPOSE: In this article, we describe a method to estimate the spatial dose variation, average dose and mean glandular dose (MGD) for a real breast using Monte Carlo simulation based on cone beam breast computed tomography (CBBCT) images. We present and discuss the dose estimation results for 19 mastectomy breast specimens, 4 homogeneous breast models, 6 ellipsoidal phantoms, and 6 cylindrical phantoms.
METHODS: To validate the Monte Carlo method for dose estimation in CBBCT, we compared the Monte Carlo dose estimates with the thermoluminescent dosimeter measurements at various radial positions in two polycarbonate cylinders (11- and 15-cm in diameter). Cone-beam computed tomography (CBCT) images of 19 mastectomy breast specimens, obtained with a bench-top experimental scanner, were segmented and used to construct 19 structured breast models. Monte Carlo simulation of CBBCT with these models was performed and used to estimate the point doses, average doses, and mean glandular doses for unit open air exposure at the iso-center. Mass based glandularity values were computed and used to investigate their effects on the average doses as well as the mean glandular doses. Average doses for 4 homogeneous breast models were estimated and compared to those of the corresponding structured breast models to investigate the effect of tissue structures. Average doses for ellipsoidal and cylindrical digital phantoms of identical diameter and height were also estimated for various glandularity values and compared with those for the structured breast models.
RESULTS: The absorbed dose maps for structured breast models show that doses in the glandular tissue were higher than those in the nearby adipose tissue. Estimated average doses for the homogeneous breast models were almost identical to those for the structured breast models (p=1). Normalized average doses estimated for the ellipsoidal phantoms were similar to those for the structured breast models (root mean square (rms) percentage difference = 1.7%; p = 0.01), whereas those for the cylindrical phantoms were significantly lower (rms percentage difference = 7.7%; p < 0.01). Normalized MGDs were found to decrease with increasing glandularity.
CONCLUSIONS: Our results indicate that it is sufficient to use homogeneous breast models derived from CBCT generated structured breast models to estimate the average dose. This investigation also shows that ellipsoidal digital phantoms of similar dimensions (diameter and height) and glandularity to actual breasts may be used to represent a real breast to estimate the average breast dose with Monte Carlo simulation. We have also successfully demonstrated the use of structured breast models to estimate the true MGDs and shown that the normalized MGDs decreased with the glandularity as previously reported by other researchers for CBBCT or mammography.

Entities:  

Mesh:

Year:  2011        PMID: 21452696      PMCID: PMC3030614          DOI: 10.1118/1.3521469

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  25 in total

1.  Dedicated breast CT: radiation dose and image quality evaluation.

Authors:  J M Boone; T R Nelson; K K Lindfors; J A Seibert
Journal:  Radiology       Date:  2001-12       Impact factor: 11.105

2.  Cone-beam CT for breast imaging: Radiation dose, breast coverage, and image quality.

Authors:  Avice O'Connell; David L Conover; Yan Zhang; Posy Seifert; Wende Logan-Young; Chuen-Fu Linda Lin; Lawrence Sahler; Ruola Ning
Journal:  AJR Am J Roentgenol       Date:  2010-08       Impact factor: 3.959

3.  Contrast-enhanced dedicated breast CT: initial clinical experience.

Authors:  Nicolas D Prionas; Karen K Lindfors; Shonket Ray; Shih-Ying Huang; Laurel A Beckett; Wayne L Monsky; John M Boone
Journal:  Radiology       Date:  2010-09       Impact factor: 11.105

4.  Dedicated cone-beam breast CT: feasibility study with surgical mastectomy specimens.

Authors:  Wei Tse Yang; Selin Carkaci; Lingyun Chen; Chao-Jen Lai; Aysegul Sahin; Gary J Whitman; Chris C Shaw
Journal:  AJR Am J Roentgenol       Date:  2007-12       Impact factor: 3.959

5.  Visibility of microcalcification in cone beam breast CT: effects of X-ray tube voltage and radiation dose.

Authors:  Chao-Jen Lai; Chris C Shaw; Lingyun Chen; Mustafa C Altunbas; Xinming Liu; Tao Han; Tianpeng Wang; Wei T Yang; Gary J Whitman; Shu-Ju Tu
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

6.  Monte Carlo and phantom study of the radiation dose to the body from dedicated CT of the breast.

Authors:  Ioannis Sechopoulos; Srinivasan Vedantham; Sankararaman Suryanarayanan; Carl J D'Orsi; Andrew Karellas
Journal:  Radiology       Date:  2008-02-21       Impact factor: 11.105

7.  Introducing DeBRa: a detailed breast model for radiological studies.

Authors:  Andy K W Ma; Spencer Gunn; Dimitra G Darambara
Journal:  Phys Med Biol       Date:  2009-06-26       Impact factor: 3.609

8.  Spatial resolution properties in cone beam CT: a simulation study.

Authors:  Lingyun Chen; Chris C Shaw; Mustafa C Altunbas; Chao-Jen Lai; Xinming Liu
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

9.  An X-Ray computed tomography/positron emission tomography system designed specifically for breast imaging.

Authors:  John M Boone; Kai Yang; George W Burkett; Nathan J Packard; Shih-ying Huang; Spencer Bowen; Ramsey D Badawi; Karen K Lindfors
Journal:  Technol Cancer Res Treat       Date:  2010-02

10.  Evaluation of tilted cone-beam CT orbits in the development of a dedicated hybrid mammotomograph.

Authors:  P Madhav; D J Crotty; R L McKinley; M P Tornai
Journal:  Phys Med Biol       Date:  2009-05-28       Impact factor: 3.609

View more
  7 in total

1.  High resolution dual detector volume-of-interest cone beam breast CT--Demonstration with a bench top system.

Authors:  Youtao Shen; Ying Yi; Yuncheng Zhong; Chao-Jen Lai; Xinming Liu; Zhicheng You; Shuaiping Ge; Tianpeng Wang; Chris C Shaw
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Dedicated breast CT: radiation dose for circle-plus-line trajectory.

Authors:  Srinivasan Vedantham; Linxi Shi; Andrew Karellas; Frederic Noo
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

3.  A general method for cupping artifact correction of cone-beam breast computed tomography images.

Authors:  Xiaolei Qu; Chao-Jen Lai; Yuncheng Zhong; Ying Yi; Chris C Shaw
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-10-29       Impact factor: 2.924

4.  Cone beam breast CT with a high pitch (75 μm), thick (500 μm) scintillator CMOS flat panel detector: visibility of simulated microcalcifications.

Authors:  Youtao Shen; Yuncheng Zhong; Chao-Jen Lai; Tianpeng Wang; Chris C Shaw
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

5.  Characterization of the homogeneous tissue mixture approximation in breast imaging dosimetry.

Authors:  Ioannis Sechopoulos; Kristina Bliznakova; Xulei Qin; Baowei Fei; Steve Si Jia Feng
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

6.  Radiation doses in volume-of-interest breast computed tomography--A Monte Carlo simulation study.

Authors:  Chao-Jen Lai; Yuncheng Zhong; Ying Yi; Tianpeng Wang; Chris C Shaw
Journal:  Med Phys       Date:  2015-06       Impact factor: 4.071

7.  Breast tomosynthesis with monochromatic beams: a feasibility study using Monte Carlo simulations.

Authors:  A Malliori; K Bliznakova; I Sechopoulos; Z Kamarianakis; B Fei; N Pallikarakis
Journal:  Phys Med Biol       Date:  2014-08-01       Impact factor: 3.609

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.